Background

It is currently not known whether vaporized perfluorohexane is superior to partial liquid ventilation (PLV) for therapy of acute lung injury. In this study, the authors compared the effects of both therapies in oleic acid-induced lung injury.

Methods

Lung injury was induced in 30 anesthetized and mechanically ventilated pigs by means of central venous infusion of oleic acid. Animals were assigned to one of the following groups: (1) control or gas ventilation (GV), (2) 2.5% perfluorohexane vapor, (3) 5% perfluorohexane vapor, (4) 10% perfluorohexane vapor, or (5) PLV with perfluorooctane (30 ml/kg). Two hours after randomization, lungs were recruited and positive end-expiratory pressure was adjusted to obtain minimal elastance. Ventilation was continued during 4 additional hours, when animals were killed for lung histologic examination.

Results

Gas exchange and elastance were comparable among vaporized perfluorohexane, PLV, and GV before the open lung approach was used and improved in a similar fashion in all groups after positive end-expiratory pressure was adjusted to optimal elastance (P < 0.05). A similar behavior was observed in functional residual capacity (FRC) in animals treated with vaporized perfluorohexane and GV. Lung resistance improved after recruitment (P < 0.05), but values were higher in the 10% perfluorohexane and PLV groups as compared with GV (P < 0.05). Interestingly, positive end-expiratory pressure values required to obtain minimal elastance were lower with 5% perfluorohexane than with PLV and GV (P < 0.05). In addition, diffuse alveolar damage was significantly lower in the 5% and 10% perfluorohexane vapor groups as compared with PLV and GV (P < 0.05).

Conclusions

Although the use of 5% vaporized perfluorohexane permitted the authors to reduce pressures needed to stabilize the lungs and was associated with better histologic findings than were PLV and GV, none of these perfluorocarbon therapies improved gas exchange or lung mechanics as compared with GV.

THE administration of perfluorocarbons in liquid form to the lungs has been associated with improved gas exchange, respiratory mechanics, and lung structure in different models of acute lung injury.1–3Perfluorocarbons have a comparatively high solubility to oxygen and carbon dioxide,4which permits pulmonary gas exchange even when the lungs are filled with those substances, as, for example, during total or partial liquid ventilation (PLV). Because of the characteristically high density of this chemical group, liquid perfluorocarbon columns may also exert a mechanical effect in dependent lung zones, contributing to the opening of atelectatic areas and maintaining them open at end-expiration,5as well as diverting pulmonary blood flow to nondependent lung zones.6In addition, perfluorocarbons are able to attenuate the proinflammatory response during lung injury, which may improve pulmonary function and histologic findings.7 

Although PLV represents a considerable advance in the ease of perfluorocarbon administration as compared with total liquid ventilation, filling the lungs with those substances may lead to transient hypoxia, impairment of hemodynamics, barotrauma, and liquothoraces.8,9Moreover, a clinical trial failed to demonstrate the ability of PLV to decrease mortality in acute respiratory failure.8Because of those facts, investigators have developed alternative forms of administration of perfluorocarbons, including nebulization and vaporization, with promising results in experimental studies.10,11In the latter approach, a perfluorocarbon with high vapor pressure, as, for example, perfluorohexane, can be administered by means of conventional anesthesia vaporizers. Because the temperature of the vaporizer is lower than body temperature, perfluorohexane is kept in the vapor state, and no liquid phase is established in the lungs, making this modality attractive for the clinical setting. To date, however, only a few studies have addressed the vaporization of perfluorohexane during lung injury, and none of them have compared this new approach to PLV.11–13Moreover, to our knowledge, only one study has investigated dose-dependent effects of vaporized perfluorohexane, suggesting that the optimal dosage is situated below 18%,14,i.e. , less than the concentration used in the first reports on perfluorohexane.11,13However, the study showing a dose-dependent effect14was performed in the isolated, perfused rabbit lung, and extrapolation to the whole animal model of lung injury is uncertain.

In view of those facts, we decided to compare the effects of 2.5%, 5%, and 10% perfluorohexane vapor with the effects of PLV and gas ventilation on gas exchange, respiratory system mechanics, lung volume, and histologic correlates in an oleic acid (OA)–induced acute lung injury. To improve the relevance of the findings, control and treatment animals were compared at relatively low positive end-expiratory pressures (PEEP), as well as after recruiting and stabilizing the lungs at higher PEEP levels (open lung approach).

All animal procedures were approved by the Institutional Animal Care Committee and the Government of the State of Saxony, Germany, and conformed to the Guide for the Care and Use of Laboratory Animals .15 

Anesthesia

Thirty female pigs weighing 31.1 ± 3.5 kg (range, 25–37 kg) were premedicated with ketamine (2–3 mg/kg intramuscular) and brought to the experimental operation room, where an ear vein was punctured. After that, the trachea was intubated with a cuffed endotracheal tube (7.5 ID), and an esophageal catheter (Erich Jaeger GmbH, Höchberg, Germany) was placed and advanced into the mid chest until optimization of the position was performed (see Instrumentation and Measurement Devices section). Thereafter, anesthesia was deepened and maintained with midazolam (0.5- to 1-mg/kg bolus plus 1–3 mg · kg−1·h−1intravenous) and ketamine (3- to 4-mg/kg bolus plus 5–10 mg · kg−1·h−1intravenous). Paralysis was achieved with 2 mg pancuronium bromide administered intravenously every hour. Volume status was maintained by means of the administration of a crystalloid solution (E153; Serumwerk Bernburg AG, Bernburg, Germany) at a rate of 10–15 mg · kg−1·h−1, with a colloid solution (10% hydroxyethyl starch; Fresenius Kabi Deutschland GmbH, Bad Homburg, Germany) being given as necessary to keep the hemoglobin concentration in blood approximately constant (normovolemia).

Mechanical Ventilation

Volume-controlled ventilation was initiated using an anesthesia machine with a semiclosed circle system (Cato®; Drägerwerk AG, Lübeck, Germany). Ventilator settings were as follows: fraction of inspired oxygen (Fio2) of 1.0 (fresh gas flow of 4 l/min), PEEP of 5 cm H2O, and inspiratory:expiratory ratio of 1:1.7, which corresponds approximately to the physiologic value. We also used tidal volumes (VT) of 9 ml/kg, which corresponds to an intermediary value between protective (6 ml/kg) and conventional ventilation (12 ml/kg). The reason for choosing this value was that in our experience, carbon dioxide elimination is easier to achieve while keeping inspiratory plateau pressures lower than 35 cm H2O. Respiratory rates were adjusted to an arterial carbon dioxide tension (Paco2) of 35–45 mmHg. Except for PEEP, which was readjusted later (see Protocol for Measurements section), all settings of the mechanical ventilator were kept constant throughout the experiments.

Instrumentation and Measurement Devices

The right anterior surface of the neck was infiltrated with 5 ml lidocaine, 1%, and a paramedian skin incision was performed. After carefully preparing the external carotid artery, an indwelling catheter was inserted, through which the sensor of a continuous blood gas analysis system (Paratrend 7+®; Diametrics, St. Paul, MN) was introduced. Mean arterial pressure was continuously monitored with a CMS Monitor (Agilent Technologies, Böblingen, Germany). Also, a pulmonary artery catheter (Opticath®; Abbott Laboratories, Abbott Park, IL) was advanced through an introducer set placed in the external jugular vein until a typical pulmonary artery curve could be visualized. Mean pulmonary artery pressures were then continuously measured with the CMS Monitor. Urine was collected by means of a bladder catheter, which was inserted during a mini-laparotomy.

A heated Fleisch pneumotachograph No. 2 (Fleisch, Lausanne, Switzerland) connected to a differential pressure transducer (PXL12X5DN; Sensortechnics, Troy, NY) was placed between the Y-piece of the ventilator tubing and the endotracheal tube. Airway pressure (Paw) was monitored by a second pressure transducer (SCX01DNC; SenSym ICT, Milpitas, CA), whose tip was placed at the proximal ending of the pneumotachograph (near to animal). The tip of the capillary of a respiratory mass spectrometer (LR6000; Logan Research, Rochester, United Kingdom) was placed at the distal ending of the pneumotachograph, and the respiratory gas mixture was sampled at 60 ml/min. Finally, the esophageal catheter balloon was inflated with 1–2.5 ml room air and connected to a third pressure transducer (SCX01DNC) to measure the esophageal pressure (Pes). The position of the esophageal catheter was optimized to obtain a correlation coefficient of approximately 1 between ΔPawand ΔPesduring thoracoabdominal compression maneuvers with a closed respiratory system, i.e. , without a phase lag in the plot the two signals, according to a method described elsewhere.16 

Gas Exchange

Arterial and mixed venous blood samples were analyzed for oxygen and carbon dioxide tension, pH, base excess, and [HCO3] using an ABL 505 (Radiometer, Copenhagen, Denmark), and for arterial oxygen saturation and hemoglobin concentration using an OSM 3 Hemoximeter (Radiometer) calibrated for swine blood. Gas tension measurements were performed at 37°C and corrected for pulmonary artery temperature at sampling. The oxygen content in arterial, end-capillary, and mixed venous blood, as well as intrapulmonary right-to-left shunt (Q̇s/Q̇t), were calculated using standard formulae, as described elsewhere.11 

Data Acquisition System

Gas concentrations measured by the mass spectrometer, airway flow measured by the pneumotachograph V̇, Paw, and Peswere digitized at 150 Hz and acquired by the Data Acquisition System, which is a main platform written for the LabVIEW environment (National Instruments, Austin, TX). The Data Acquisition System supports the integration of other subroutines which were developed for calibration, calculation of parameters, and control of processes (see Calculations of Respiratory Mechanics section).

Calibration of the Mass Spectrometer

The following gases and vapor were measured at 32 Hz by the mass spectrometer: N2, O2, CO2, Ar, and vaporized perfluorohexane. Calibration was performed with a two-point linear procedure internal to the device, which requires two gas mixtures with at least one common gas. The first mixture contained 78% N2, 17% O2, and 5% CO2and was produced and certified by Messer Griesheim GmbH (Ludwigshafen, Germany). The second mixture contained 71.29% N2, 19.19% O2, 0.86% Ar, and 8.57% perfluorohexane and was obtained by vigorously shaking a chamber of variable volume containing room air and a liquid phase of perfluorohexane at 0°C. The fraction of perfluorohexane in the second mixture was estimated according to equation 1, which was derived by the National Institute of Standards and Technology (NIST) based on the data from Crowder et al.  17:

where FPFHis the fraction of perfluorohexane vapor and TKis the temperature in °K.

Calibration of the Pneumotachograph

For the calibration of the pneumotachograph, animals were ventilated with an intensive care mechanical ventilator (Evita 2; Drägerwerk AG, Lübeck, Germany) whose respiratory settings were the same as those of the anesthesia machine. Before the change of the ventilators, VTof the Evita 2 was verified with a side stream spirometry tester (Datex, Helsinki, Finland). The flow signal was recorded by the Data Acquisition System during 2 min, and the inspiratory flow was increased by 50% in the last 30 s of acquisition without changing VT. Increase of flow was necessary to obtain a wider flow range where the calibration procedure should be performed. Signals were processed off-line by a subroutine written for Matlab® (MathWorks, Natick, MA) to calculate the coefficients of a third-order polynomial that best fitted to the flow data according to a procedure described elsewhere.18Calibration was accepted if the error of inspiratory and expiratory VTin each respiratory cycle was less than 3%. Correction of flow for the variation of the viscosity in the presence of perfluorohexane was performed using empiric factors obtained in our laboratory. Such procedures were necessary because we noticed that the calibration of the Fleisch pneumotachograph outside of the respiratory circuit with a 3-l syringe, as described by other investigators for spontaneous breathing,18was not appropriate for use during mechanical ventilation. After calibration, animals were disconnected from the Evita 2 and reattached to the anesthesia machine. Change of ventilators was always preceded by clamping of the endotracheal tube during inspiration to avoid loss of lung volume.

Compensation for Delay between Flow and Gas Concentration

For the calculation of time delay between flow and gas concentration, a special syringe containing two gas mixtures separated by the piston was used, as described in detail in the work by Brunner et al.  19Briefly, the wall of the syringe has a leak where the tip of the capillary of the mass spectrometer is attached. When the piston is moved, a step change in gas concentration is applied to the tip of the capillary. If the volumes on both sides of the piston are known, the exact moment the piston passes at the tip of the capillary can be determined, and the delay in measuring the gas concentration step in the mass spectrometer can be assessed. In this work, the step change in gas concentration was achieved with a first gas mixture corresponding to room air and a second gas mixture consisting of 78% N2, 17–19% O2, and 3–5% CO2. Time delay (TD) between flow and gas concentration measured in this way can be divided into a viscosity-dependent part and a viscosity-independent part. For the purpose of simplification, it was assumed in this work that TDconsisted only of a viscosity-dependent part, which was calculated according equation 2:

where j is the sample, TDois the time elapsed from the application of a carbon dioxide step change at the tip of the capillary of the mass spectrometer until 50% of the final carbon dioxide value, and η0is the viscosity of the gas mixture used. The variable η̄ represents the average viscosity of the time sample j traveling in the capillary, whose calculation has been shown in the detail in the work by Brunner et al.  19Viscosities of isolated gases and mixtures were calculated according to Turner et al.  20 

Calibration of Esophageal and Airway Pressures

Pawand Peswere calibrated using a water column as reference.

Calculations of Respiratory Mechanics

Elastance and resistance variables were determined on-line according to a simplified version of the equation of motion, as described in equation 3:

where P is the motion pressure, V is volume, t is time, Prestcorresponds to the pressure at rest volume, and E and R represent elastance and resistance, respectively. Partitioning of E and R into lung and chest wall components (EL, ECW, RL, and RCW, respectively) was achieved by applying the appropriate motion pressures, as described in equations 4 and 5:

where RS represents the respiratory system values, P0is the residual airway pressure at the end-expiratory lung volume (approximately PEEP), and P1is the residual pressure in the pleural space at end-expiration. Curve fitting to a one-compartment model was performed with data acquired during periods of 2 min, and parameter extraction was obtained using a recursive minimal least squares method according to a method described elsewhere.21After sudden changes in lung mechanics, values usually stabilized within 2 min. The chest wall components of elastance and resistance (ECWand RCW, respectively) were calculated as the difference between the respective respiratory system and lung components.

Calculation of Functional Residual Capacity

Functional residual capacity measurements were obtained with the multiple breath nitrogen washout method. The Fio2was reduced to 0.4, while leaving other settings of the anesthesia machine unchanged, and enough time was allowed for the respiratory system to stabilize (up to 15 min). After stabilization, the endotracheal tube was clamped at end-expiration until the animal was connected to the Evita 2 mechanical ventilator. The settings of the Evita 2 were the same as those from the anesthesia machine, except for the Fio2, which was fixed at 0.8. Flow and gas concentration data were acquired by the Data Acquisition System for off-line processing. The washout maneuver was interrupted when differences between inspiratory and expiratory N2concentrations (FIN2and FEN2respectively) were negligible, i.e. , FIN2− FEN2< 1%, or remained unchanged longer than 1 min. After that, the endotracheal tube was clamped, and the animal was connected back to the anesthesia machine, whose settings were the same as before the washout maneuver.

Calculation of FRC followed the mass balance principle, according to equation 6:

where tBis the begin of the washout maneuver and tEis the end of the calculation period. Time tEwas defined as the time when FRC calculation achieves a plateau.

Cardiac Output Measurements

Cardiac output was measured by means of the thermodilution method. Four measurements with 10 ml iced saline 0.9% solution equally spread over the respiratory cycle were obtained and averaged.

Oleic Acid–induced Lung Injury

Acute lung injury was induced by means of the central venous infusion of OA. Initially, 0.05 ml/kg OA was diluted in arterial blood to obtain 10 ml total volume. The mixture was gently shaken, and aliquots of 0.25–1 ml were administered carefully through the proximal lumen of the pulmonary catheter over 1 min, while increasing the parallel administration of crystalloid to a rate of 800–900 ml/h to minimize the impact of boluses of OA and improve volume status during injury. The injection of the mixture was stopped if the mean arterial pressure decreased to under 50 mmHg, and administration of adrenalin boluses was permitted if spontaneous recovery was not observed within 30 s. This procedure was repeated during 1–2 h, until the arterial oxygen tension (Pao2) decreased to below 200 mmHg and did not show a trend toward improvement within 10 min. If necessary, a second dose of OA was prepared in the same way and administered within this period, the total OA amount being situated within 0.03–0.125 ml/kg. During the injury period, the Pao2was assessed continuously with the Paratrend 7+ (Diametrics) to allow better control of injury, but decisions regarding continuation or interruption of the procedure were taken by means of arterial blood sampling and analysis with the ABL 505.

Protocol for Measurements

Figure 1shows the time course of events. After instrumentation, a period of 30 min was allowed for animals to stabilize. Baseline measurements were then obtained and injury was performed in 30 animals, as previously described. After injury (time 0), animals were assigned to one of the five therapy groups after opening sealed envelopes (n = 6/group): (1) control—no therapy with perfluorocarbon or gas ventilation (GV), (2) therapy with 2.5% perfluorohexane, (3) therapy with 5% perfluorohexane, (4) therapy with 10% perfluorohexane, and (5) PLV with perfluorooctane (30 ml/kg) (see Partial Liquid Ventilation section). Two hours thereafter, measurements were repeated (time 2), and the lungs were recruited by means of a continuous pressure of 40 cm H2O during 30 s. To avoid derecruitment and stabilize the lungs thereafter, PEEP was set at 20 cm H2O while other respiratory parameters remained unchanged. After that, PEEP was reduced in steps of 5 cm H2O up to the start value, with periods of 3 min being allowed at each level (20, 15, 10, and 5 cm H2O). Measurements of total respiratory elastance (ERS= EL+ ECW) and Pao2were taken at each step. After this procedure, recruitment was repeated in the same manner, and PEEP was set at the level corresponding to the lowest ERSlevel. If ERSvalues were comparable between two steps (difference < 5 cm H2O/l), the higher Pao2level was used to set PEEP. Measurements were repeated 4 and 6 h after injury (times 4 and 6, respectively). Except for PEEP, mechanical ventilator settings were the same as before lung injury. Finally, animals were killed by means of the intravenous injection of 2 g thiopental and 50 ml KCl, 1 m.

Fig. 1. Time course of events. PEEP = positive end-expiratory pressure .

Fig. 1. Time course of events. PEEP = positive end-expiratory pressure .

Close modal

Administration of Vaporized Perfluorohexane

Perfluorohexane (C6F14; ABCR, Karlsruhe, Germany) with a purity of 95% was used in this work. This compound was chosen because of its relatively low molecular weight (338 amu) and unique physicochemical properties among the perfluorocarbons, particularly the relatively high vapor pressure (273 mmHg at 30°C), which permits its administration in vapor form.22Accordingly, administration of perfluorohexane vapor was accomplished by means of two standard bypass vaporizers, type 19 n (Drägerwerk AG, Lübeck, Germany), which were modified by a scaled opening of the dosage control cone. Vaporizers were connected in series in the inspiratory limb of the anesthesia machine and opened to permit mixing of perfluorohexane with the fresh gas flow. Inspiratory and expiratory concentrations of perfluorohexane were measured continuously by the mass spectrometer, ensuring a controlled application.

Partial Liquid Ventilation

Perfluorooctane (C8F18; ABCR) with a purity of 98% was used during PLV in this work. This substance was chosen because it has a relatively low vapor pressure at 37°C (60.9 mmHg), which permits its use without important losses by means of evaporation, and other physicochemical properties that are comparable to those of perfluorooctyl bromide, namely specific weight (1.74 mg/ml), solubility for oxygen (52.1 ml/100 ml), viscosity (0.85 cs °C), and surface tension (13.9 dyn/cm),22which has been used for PLV in different studies. Before administration, oxygen was bubbled in 30 ml/kg perfluorooctane at a rate of 4 l/min during 2 min. After that, animals were disconnected from the anesthesia machine, and aliquots of 10%, 20%, 30%, and 40% of the total amount were dropped into the endotracheal tube with 1- to 2-min intervals between consecutive applications. Compensation for evaporative losses was performed by means of administration of perfluorooctane every 2 h through a special side port constructed for the endotracheal tube, to avoid disconnection from the ventilator. The compensation volume was extrapolated from the work by Reickert et al.  23According to these authors, the maximum compensation rate for PLV with perfluorooctylbromide in adults should be 9.4 ± 3 ml/h. Because the vapor pressure of perfluorooctane is approximately fivefold that of perfluorooctylbromide and the volume of perfluorooctane used in our animals was approximately one half of that used in humans,2350 ml perfluorooctane was refilled every 2 h in our study.

Histologic Analysis

Postmortem, a median sternotomy was performed to excise the heart and lungs en bloc . We opted not to use continuous positive airway pressure during this procedure because we were aware of the possibility that lungs treated with perfluorohexane or PLV might not collapse as easily as nontreated lungs at atmospheric pressure. Because one of the histologic aspects to be investigated was development of microatelectasis, such an effect would be more easily detected if no continuous positive airway pressure was used. The heart was dissected from the lungs, and eight samples from defined sectors were obtained. Blocks of lung tissue of approximately 8 cm3were taken from the outermost apex and the central region of both the upper and lower lobe of the right lung. The lung tissue blocks were immediately immersed in a 4% neutral-buffered formaldehyde solution and fixed for several days. After embedding the blocks in paraffin, slices of 5 μm were cut and stained with hematoxylin and eosin, using routine histologic techniques. The sections of the samples that were used for staining were randomly selected by assisting technicians unfamiliar with the experiment. To quantify the extension and severity of histologic lung injury, hematoxylin and eosin–stained sections were observed by an expert in lung pathology (M. K.) who was blinded to the groups. The pathologic features of the lung in adult respiratory distress syndrome can be characterized under the term diffuse alveolar damage  (DAD), which was first proposed by Katzenstein and Askin.24DAD corresponds to a stereotypic pathologic sequence of events after severe acute lung injury regardless of the initial cause. The scoring system used in this work was adapted and modified to include a weighting system similar to that proposed by Broccard et al.  25This modification was performed to improve the quantification of the pathologic features according to their extension. We restricted the numerous features described in the literature to seven basic ones, which, in our opinion, are more representative for our model, namely alveolar edema, interstitial edema, hemorrhage, inflammatory infiltration, epithelial damage, microatelectasis, and overdistension. These features were evaluated in four separated nonoverlapping fields of view of hematoxylin and eosin–stained section by light microscopy (Leica DM RB; Wetzlar, Germany) using magnifications of ×25, ×100, ×250, and ×400. In our scoring system, values from 0 to 3 represented the severity of the feature, as follows: 0 = normal appearance, 1 = slight effect, 2 = intermediate effect, and 3 = severe effect. A further part of our system was used to describe the extent of involvement in each field of view, as follows: 0 = lack of involvement of the feature, 1 = up to 25%, 2 = 25–50%, 3 = 50–75%, 4 = 75% to almost total involvement, and 5 = total involvement of the field. For each feature evaluated, severity was multiplied by the extent, leading to values in the range of 0–15. Values of all fields per lung (n = 16) were averaged. The sum of features corresponded to the DAD score.

Statistical Analysis

Values are given as mean ± SD. Paired Student t  tests were used to assess the effects of lung injury. Comparability among groups at baseline and time 0, total amounts of OA per group, PEEP levels after lung recruitment, and histologic findings were tested with one-way analysis of variance. Repeated-measures two-way analysis of variance was used to determine the effects of therapies on time course of variables (time, group, and time * group effects). When significance was achieved, it was followed by post hoc  analysis (Student-Newman-Keuls test). Statistical analysis was performed using SPSS, version 11.5 (SPSS, Chicago, IL), and significance was accepted at P < 0.05.

General Aspects

Values of all variables investigated were comparable at baseline and immediately after lung injury (time 0). Also, the amount of OA needed to induce lung injury did not differ significantly among groups (P = 0.64). After the recruitment maneuvers, PEEP had to be increased in every animal to optimize the lung elastance (table 1). In animals treated with PLV, minimal elastance was obtained at PEEP values higher than in animals treated with perfluorohexane (P < 0.05) but comparable to GV. Animals treated with 5% perfluorohexane needed the lowest PEEP levels, which were significantly different from GV and PLV groups (P < 0.05).

Table 1. Positive End-expiratory Pressure before and after Lung Recruitment 

Table 1. Positive End-expiratory Pressure before and after Lung Recruitment 
Table 1. Positive End-expiratory Pressure before and after Lung Recruitment 

Gas Exchange

Administration of OA resulted in a significant decrease of Pao2/Fio2in all groups at time 0 as compared with baseline (P < 0.05; fig. 2). At time 2, Pao2/Fio2values decreased further (P < 0.05) but were comparable among groups. Two hours after lung recruitment and optimization of PEEP (time 4), Pao2/Fio2values were higher than at time 0 and time 2 (P < 0.05) but were not significantly different among groups. By end of the observation period, improvement in oxygenation was still significant with respect to time 2 (P < 0.05), but Pao2/Fio2values of PLV and perfluorohexane groups were not significantly different from GV. After lung injury, Paco2increased and arterial blood pH decreased significantly (P < 0.05; table 2). A further increase in Paco2could be observed at times 2, 4, and 6 as compared with time 0 (P < 0.05). However, Paco2values at time 4 and 6 were comparable to those at time 2. No statistical significant differences among therapy groups and GV could be detected at any time point.

Fig. 2. Arterial oxygen tension (Pao2)/fraction of inspired oxygen (Fio2) ratio. Values are given as mean ± SD for the control (gas ventilation), vaporized perfluorohexane (PFH), and partial liquid ventilation (PLV) groups. Time 0 corresponds to injury. After time 2, a recruitment maneuver was performed, and the positive end-expiratory pressure increased to optimal elastance while other respiratory settings were kept constant (see text). There were no significant differences between any therapy group and control. NS = not significant. 

Fig. 2. Arterial oxygen tension (Pao2)/fraction of inspired oxygen (Fio2) ratio. Values are given as mean ± SD for the control (gas ventilation), vaporized perfluorohexane (PFH), and partial liquid ventilation (PLV) groups. Time 0 corresponds to injury. After time 2, a recruitment maneuver was performed, and the positive end-expiratory pressure increased to optimal elastance while other respiratory settings were kept constant (see text). There were no significant differences between any therapy group and control. NS = not significant. 

Close modal

Table 2. Gas Exchange 

Table 2. Gas Exchange 
Table 2. Gas Exchange 

Animals showed a significant increase in Q̇s/Q̇tand decrease in Sv̄o2at time 0 (P < 0.05) as a result of the administration of OA (table 2). A further deterioration in Q̇s/Q̇t, but not Sv̄o2, was observed at time 2 (P < 0.05). After the recruitment maneuver and optimization of the ventilation according to the pulmonary mechanics, both Q̇s/Q̇tand Sv̄o2improved significantly as compared with time 2 (P < 0.05), but in a similar fashion in PLV and perfluorohexane groups when compared with GV.

Respiratory System Mechanics

Administration of OA led to a significant increase in EL(P < 0.05; table 3), and a slight deterioration was observed 2 h thereafter (P < 0.05). After recruitment of the lungs and increase of PEEP, ELimproved significantly (P < 0.05) and remained at lower levels up to time 6. In contrast, ECWshowed no statistical significant changes throughout the experiments (table 3). Elastance parameters of PLV and perfluorohexane groups did not differ significantly from GV at any time point.

Table 3. Respiratory Mechanics and Lung Volumes 

Table 3. Respiratory Mechanics and Lung Volumes 
Table 3. Respiratory Mechanics and Lung Volumes 

After recruitment maneuvers (time 4 and 6), RLshowed a significant improvement (P < 0.05), but values remained higher than immediately after establishment of lung injury (time 0; P < 0.05). Compared with GV, PLV led to a consistent increase of RLafter time 2 toward the end of the observation period (P < 0.05). Also, treatment with 10% perfluorohexane was associated with higher RLvalues than GV at time 6 (P < 0.05). RCWdecreased slightly after administration of OA (P < 0.05), but values did not change significantly thereafter. In addition, PLV and perfluorohexane groups did not differ significantly from GV at any time point.

Functional Residual Capacity

Functional residual capacity decreased significantly after lung injury (P < 0.05), and 2 h thereafter, a further deterioration could be observed (P < 0.05; table 3). After recruitment of the lungs, FRC values increased significantly as compared with times 0 and 2 (P < 0.05) and remained at higher levels toward the end of the observation period. Although different PEEP levels among groups were required to stabilize the lungs (table 1), FRC values of perfluorohexane groups did not differ significantly from GV.

Hemodynamics

Mean arterial pressure and mean pulmonary arterial pressures values increased significantly after induction of lung injury by OA and were accompanied by an impairment of cardiac output (P < 0.05; table 4). Mean arterial pressure values decreased slightly but significantly at time 2 (P < 0.05), whereas mean pulmonary arterial pressure and cardiac output did not change as compared with time 0. After PEEP was set to optimize the mechanics of the respiratory system, mean arterial pressure and mean pulmonary arterial pressure values decreased (P < 0.05), but no change could be detected in cardiac output. However, PLV and perfluorohexane groups did not differ significantly from GV at any time point.

Table 4. Hemodynamics 

Table 4. Hemodynamics 
Table 4. Hemodynamics 

Histologic Analysis

Results of histologic analysis are summarized in table 5. Perfluorohexane at 5% and 10%, but not 2.5% perfluorohexane or PLV, significantly reduced DAD scores as compared with GV (P < 0.05). Animals treated with 5% perfluorohexane achieved the lowest score compared with all other concentrations tested, PLV, and GV. Accordingly, analysis revealed that most features were improved with 5% perfluorohexane as compared with GV, with the exception of microatelectasis and overdistension. Also, 10% perfluorohexane importantly reduced intraalveolar edema and epithelial damage as compared with control animals. PLV led to less intraalveolar edema, hemorrhage, and inflammatory infiltration than in controls (P < 0.05). However, overdistension was more evident with PLV than in other groups, and microatelectasis was more pronounced with the liquid perfluorocarbon than with 5% perfluorohexane and GV.

Table 5. Histologic Examination 

Table 5. Histologic Examination 
Table 5. Histologic Examination 

The most important finding of this study was that the use of 5% perfluorohexane vapor permitted stabilization of the lungs at lower PEEP levels during the open lung approach and was associated with improved histologic scores as compared with PLV and GV. Despite such beneficial effects, neither the vapor nor the liquid perfluorocarbon led to further improvements in gas exchange and pulmonary mechanics during conventional ventilation or the open lung approach. These issues will be discussed systematically.

Gas Exchange

Our finding that vaporized perfluorohexane and PLV did not improve gas exchange during conventional ventilation is in disagreement with previous reports in the literature.11,12,26–29A possible explanation for this discrepancy lies in the protocol for induction of lung injury. We administered OA in relatively small aliquots during a period of 1–2 h, which may lead to a more severe injury. We chose this regimen to avoid spontaneous recovery and because the slow onset of injury should mimic the clinical picture more closely. It could also be argued that the lack of effect of perfluorohexane vapor on gas exchange lies in the concentrations used. The first works showing improvement in oxygenation with this modality used concentrations as high as 18%,11,13whereas we worked with 10%. We decided to test a lower range because a more recent investigation on dose-dependent effects performed by one of the authors (J.-U. B.) suggested that concentrations far below 18%, namely 4.5% and 9%, are associated with better protection against the development of lung injury.14Although it cannot be completely excluded that higher concentrations might have been beneficial in the current study, the observation that none of the parameters investigated showed a linear relation with the concentration used makes it unlikely that 18% perfluorohexane would have been beneficial. In fact, Kemming et al.  30could not detect a positive effect of 18% perfluorohexane on gas exchange in lung injury induced by intravenous endotoxin.

Another possible explanation for the lack of positive effects lies in the mechanical ventilation protocol used in our work. After a short period of 2 h, lungs were recruited according to the open lung approach, which per se  is able to improve gas exchange dramatically and may have masked possible beneficial effects of perfluorocarbons. In fact, an increase in Pao2could be observed in some of the animals of the PLV group immediately after the beginning of that therapy with the Paratrend 7+® device (data not shown) but lasted only a few minutes. It is noteworthy that most previous experimental studies dealing with vaporized perfluorohexane and PLV have not used respiratory strategies in their control groups, which are common during ventilation of patients with acute lung injury. Fujino et al.  31compared the effects of PLV versus  GV with PEEP set 1 cm H2O above the lower inflection point in the static pressure–versus –volume curve and observed that oxygenation increased in a similar fashion in both groups. In our study, we also observed a dramatic increase in Pao2/Fio2after recruitment was performed and PEEP was set to minimize elastance, which is consistent with previous clinical and experimental studies.32,33Nevertheless, oxygenation was comparable between PLV and the control group in our study too. In contrast to Fujino et al. ,31however, we did not observe a decrease in Paco2values in the PLV group as compared with GV. This discrepancy is explained by differences in the ventilatory strategies used. Fujino et al.  31worked with pressure-controlled ventilation, and lower PEEP values were necessary to stabilize the lungs in their PLV group. This resulted in higher driving pressures and VTvalues in the PLV than in the GV group.31In our study, animals were ventilated in a volume-controlled mode and, therefore, VTvalues remained constant.

Respiratory System Mechanics

In contrast to other reports in the literature,29,34PLV combined with relatively low PEEP values was not able to improve ELin our animals. Accordingly, filling the lungs with 30 ml/kg perfluorooctane did not result in significant changes of EL. A similar value was observed in the groups treated with vaporized perfluorohexane.

The improvement of ELvalues after recruitment of the lungs is consistent with previous clinical and experimental studies.32,35,36Such effect is probably mediated by accommodation of edema in fluid-filled alveoli of lung-dependent regions.36Unfortunately, however, the use of vaporized perfluorohexane or PLV did not add any benefit to elastance parameters.

The finding that minimal elastance in animals treated with PLV was not different from GV suggests that the dynamic pressure–versus –volume curve was not shifted to the left during PLV, as reported by Fujino et al.  31This discrepancy is most probably explained by different methods used to optimize the respiratory system mechanics. Fujino et al.  31analyzed the static pressure–versus –volume curves, whereas we decreased PEEP from 20 cm H2O to 5 cm H2O in steps of 5 cm H2O and measured the elastance at each PEEP level. The definition of lower inflection point is somewhat arbitrary, but it has been usually related to the point, or zone, in the pressure-versus -volume curve above which elastance is lowest. Accordingly, Hickling37suggested that a decremental PEEP trial, as used in our work, is useful to determine the PEEP needed to maintain the lungs open. It is worth noting that the group treated with 5% perfluorohexane required the lowest PEEP levels for achieving minimal elastance, suggesting that this concentration was able to reduce the pressure needed to stabilize the lungs.

In contrast to elastance, resistance was influenced by administration of the perfluorocarbons. In a rigid tube, resistance is proportional to the viscosity of the streaming medium,38which is much higher for perfluorocarbons in liquid and vapor form than for air.39,40Therefore, it is not surprising that 2 h after instillation of perfluorooctane into the lungs, RLbut not RCWvalues were higher in PLV than in GV animals. The marked overall decrease in RLvalues which could be observed after lung recruitment most probably resulted from an increase in the cross-sectional area of the airways that occurred at higher lung volumes. However, such effect seems to have been outweighed by the higher viscosity of the streaming medium in animals treated with PLV and 10% perfluorohexane, as suggested by increased RLvalues in those groups.

Functional Residual Capacity

The dynamic in the variation of FRC in our study is consistent with previous reports in the literature, which showed that the open lung concept is able to restore the lung volume after injury with OA.41,42Although vaporized perfluorohexane did not contribute to an increase FRC, PEEP values used in the 5% perfluorohexane group were lower than in control animals, suggesting that this therapy may lead to an increase in aerated lung volume.

Lung Histology

Although PLV was associated with improved intraalveolar edema, hemorrhage, and inflammatory infiltration, these beneficial effects were outweighed by microatelectasis and overdistension, leading to relatively high DAD scores. However, the finding that both microatelectasis and overdistension occurred simultaneously in animals treated with PLV seems paradoxical. Because of the relatively high density of perfluorooctane, this substance spreads preferentially in dependent lung zones. Therefore, if the lung volume becomes higher than the instilled volume of perfluorocarbon, it is theoretically possible that PLV shifts from ventral to dorsal lung zones, following the gravity gradient. We could observe that the meniscus of perfluorooctane in the endotracheal tube disappeared after the recruitment of the lungs, despite compensation for evaporation loss. Therefore, the open lung concept probably resulted in uneven distribution of liquid and, consequently, of histologic findings in PLV animals. It could be argued that use of lower liquid perfluorocarbon dosages, e.g. , 20–22 ml/kg, as used by Fujino et al. ,31instead of 30 ml/kg, could lead to a better performance of PLV, minimizing the findings of overdistension. However, lower perfluorooctane volumes could also have led to development of more microatelectasis. Although the dosage of PLV was relatively high in our study, the meniscus in the tube disappeared after recruiting the lungs. This observation suggests that perhaps dosages higher than 30 ml/kg are necessary to completely fill the lungs with liquid perfluorocarbons during the open lung approach.

In contrast to PLV, vaporized perfluorohexane led to a reduction of most features investigated, resulting in a consistent reduction of the DAD score. Interestingly, a nonlinear, dose-dependent effect of vaporized perfluorohexane was observed, with an optimum at 5%. In fact, treatment with 5% perfluorohexane proved superior to 10% perfluorohexane and PLV with regard to the potential of reducing lung injury. Curiously, this optimum represents approximately the same concentration that was associated with best protection against development of lung injury after a challenge with Ca-ionophore A23187 in the isolated and perfused rabbit lung.14 

It was beyond the scope of this study to investigate how the perfluorocarbons used could have contributed to lung protection in our animals. Possible mechanisms that could explain the effects of PLV include reduction of the inflammatory response43and stabilization of alveoli at end-expiration, which may result in better distribution of mechanical stress across lung zones and minimization of further lung injury of mechanical ventilation.44Perfluorohexane also pursues antiinflammatory and anticoagulatory properties that could explain, at least in part, our results.45In addition, vaporized perfluorohexane seems to have the potential to attenuate ventilator-induced lung injury by unknown mechanisms, as demonstrated recently by our group.46 

The use of PLV has been practically abandoned after clinical trials failed to demonstrate any beneficial effect of administration of perfluorocarbons as liquid during acute lung injury. Our results demonstrated that although the therapy with another administration form, namely vaporization, was superior to PLV to improve lung structure, such effect was not accompanied by improved functional parameters, i.e. , gas exchange and respiratory system mechanics, except to a reduction of the pressures needed to stabilize the lungs in the OA model of lung injury. Because the observation period was limited to 6 h, we do not know whether the improved histologic findings would prove beneficial in the long term. This issue remains speculative and requires that further studies with vaporized perfluorohexane be performed.

1.
Curtis SE, Peek JT, Kelly DR: Partial liquid breathing with perflubron improves arterial oxygenation in acute canine lung injury. J Appl Physiol 1993; 75:2696–702
2.
Fitzpatrick JC, Jordan BS, Salman N, Williams J, Cioffi Jr, WG Pruitt BA Jr: The use of perfluorocarbon-associated gas exchange to improve ventilation and decrease mortality after inhalation injury in a neonatal swine model. J Pediatr Surg 1997; 32:192–6
3.
Hernan LJ, Fuhrman BP, Kaiser RE, Penfil S, Foley C, Papo MC, Leach CL: Perfluorocarbon-associated gas exchange in normal and acid-injured large sheep. Crit Care Med 1996; 24:475–81
4.
Kaisers U, Kelly KP, Busch T: Liquid ventilation. Br J Anaesth 2003; 91:143–51
5.
Tooley R, Hirschl RB, Parent A, Bartlett RH: Total liquid ventilation with perfluorocarbons increases pulmonary end-expiratory volume and compliance in the setting of lung atelectasis. Crit Care Med 1996; 24:268–73
6.
Harris RS, Willey-Courand DB, Head A, Galletti GG, Call DM, Venegas JG: Regional VA, Q, and VA/Q during PLV: effects of nitroprusside and inhaled nitric oxide. Am J Respir Crit Care Med 2002; 92:297–312
7.
van Eeden SF, Klut ME, Leal MA, Alexander J, Zonis Z, Skippen P: Partial liquid ventilation with perfluorocarbon in acute lung injury: Light and transmission electron microscopy studies. Am J Respir Cell Mol Biol 2000; 22:441–50
8.
Hirschl RB, Croce M, Gore D, Wiedemann H, Davis K, Bartlett RH: Prospective, randomized, controlled pilot study of partial liquid ventilation in adult acute respiratory distress syndrome. Am J Respir Crit Care Med 2002; 165:781–7
9.
Cox PN, Frndova A, Tan PS, Nakamura T, Miyasaka K, Sakurai Y, Middleton W, Mazer D, Bryan AC: Concealed air leak associated with large tidal volumes in partial liquid ventilation. Am J Respir Crit Care Med 1997; 156:992–7
10.
Kandler MA, von der Hardt K, Schoof E, Dotsch J, Rascher W: Persistent improvement of gas exchange and lung mechanics by aerosolized perfluorocarbon. Am J Respir Crit Care Med 2001; 164:31–5
11.
Bleyl JU, Ragaller M, Tscho U, Regner M, Kanzow M, Hübler M, Rasche S, Albrecht DM: Vaporized perfluorocarbon improves oxygenation and pulmonary function in an ovine model of acute respiratory distress syndrome. Anesthesiology 1999; 91:461–9
12.
Bleyl JU, Ragaller M, Tscho U, Regner M, Hübler M, Kanzow M, Albrecht DM: Changes in pulmonary function and oxygenation during application of perfluorocarbon vapor in healthy and oleic acid-injured animals. Crit Care Med 2002; 30:1340–7
13.
Hübler M, Souders JE, Shade ED, Polissar NL, Schimmel C, Hlastala MP: Effects of vaporized perfluorocarbon on pulmonary blood flow and ventilation/perfusion distribution in a model of acute respiratory distress syndrome. Anesthesiology 2001; 95:1414–21
14.
Bleyl JU, Heller AR, Heintz M, Gama de Abreu M, Albrecht DM, Koch T: Dose dependent effects of perfluorohexane vapor on inflammatory lung injury in the isolated, perfused rabbit lung. Am J Respir Crit Care Med 2002; 162:A46
15.
Guide for the Care and Use of Laboratory Animals. Washington, D.C., Institute of Laboratory Animal Resources, Commission on Life Sciences, National Research Council National Academy Press, 1996
Washington, D.C
,
Institute of Laboratory Animal Resources, Commission on Life Sciences, National Research Council National Academy Press
16.
Lanteri CJ, Kano S, Sly PD: Validation of esophageal pressure occlusion test after paralysis. Pediatr Pulmonol 1994; 17:56–62
17.
Crowder GA, Taylor ZL, Reed TM, Young JA: Vapor pressures and triple point temperatures for several pure fluorocarbons. J Chem Eng Data 1967; 12:481–5
18.
Yeh MP, Gardner RM, Adams TD, Yanowitz FG: Computerized determination of pneumotachometer characteristics using a calibrated syringe. J Appl Physiol 1982; 53:280–5
19.
Brunner JX, Wolff G, Cumming G, Langenstein H: Accurate measurement of N2 volumes during N2 washout requires dynamic adjustment of delay time. J Appl Physiol 1985; 59:1008–12
20.
Turner MJ, MacLeod IM, Rothberg AD: Effects of temperature and composition on the viscosity of respiratory gases. J Appl Physiol 1989; 67:472–7
21.
Avanzolini G, Barbini P, Cappello A, Cevenini G, Chiari L: A new approach for tracking respiratory mechanical parameters in real-time. Ann Biomed Eng 1997; 25:154–63
22.
Ragaller M, Bleyl JU, Koch T, Albrecht DM: From isoflurane to perfluorohexane? Perfluorocarbons-therapeutic strategies in acute lung failure. Anaesthesist 2000; 49:291–301
23.
Reickert C, Pranikoff T, Overbeck M, Kazerooni E, Massey K, Bartlett R, Hirschl R: The pulmonary and systemic distribution and elimination of perflubron from adult patients treated with partial liquid ventilation. Chest 2001; 119:515–22
24.
Katzenstein AL, Askin FB: Interpretation and significance of pathologic findings in transbronchial lung biopsy. Am J Surg Pathol 1980; 4:223–34
25.
Broccard AF, Shapiro RS, Schmitz LL, Ravenscraft SA, Marini JJ: Influence of prone position on the extent and distribution of lung injury in a high tidal volume oleic acid model of acute respiratory distress syndrome. Crit Care Med 1997; 25:16–27
26.
Gauger PG, Overbeck MC, Chambers SD, Cailipan CI, Hirschl RB: Partial liquid ventilation improves gas exchange and increases EELV in acute lung injury. J Appl Physiol 1998; 84:1566–72
27.
Tutuncu AS, Faithfull NS, Lachmann B: Comparison of ventilatory support with intratracheal perfluorocarbon administration and conventional mechanical ventilation in animals with acute respiratory failure. Am Rev Respir Dis 1993; 148:785–92
28.
Hirschl RB, Tooley R, Parent AC, Johnson K, Bartlett RH: Improvement of gas exchange, pulmonary function, and lung injury with partial liquid ventilation: A study model in a setting of severe respiratory failure. Chest 1995; 108:500–8
29.
Vasquez de Anda GF, Lachmann RA, Verbrugge SJC, Gommers D, Haitsma JJ, Lachmann B: Partial liquid ventilation improves lung function in ventilation-induced lung injury. Eur Respir J 2001; 18:93–9
30.
Kemming GI, Flondor M, Hanser A, Pallivathukal S, Holtmannspötter M, Kneisel FF, Reuter DA, Kisch-Wedel H, Zwissler B: Effects of perfluorohexan vapor on gas exchange, respiratory mechanics, and lung histology in pigs with lung injury after endotoxin infusion. Anesthesiology 2005; 103:585–94
31.
Fujino Y, Goddon S, Chiche JD, Hromi J, Kacmarek RM: Partial liquid ventilation ventilates better than gas ventilation. Am J Respir Crit Care Med 2000; 162:650–7
32.
Amato MBP, Barbas CSV, Medeiros DM, Schettino GPP, Lorenzi Filho G, Kairalla RA, Deheinzelin D, Morais C, Fernandes EO, Takagaki TY, de Carvalho CRR: Beneficial effects of the “open lung approach” with low distending pressures in acute respiratory distress syndrome. Am J Respir Crit Care Med 1995; 152:1835–46
33.
Vasquez de Anda GF, Hartog A, Verbrugge SJC, Gommers D, Lachmann B: The open lung concept: Pressure-controlled ventilation is as effective as high-frequency oscillatory ventilation in improving gas exchange and lung mechanics in surfactant-deficient animals. Intensive Care Med 1999; 25:990–6
34.
Papo MC, Paczan PR, Fuhrman BP, Steinhorn DM, Hernan LJ, Leach CL, Holm BA, Fisher JE, Kahn BA: Perfluorocarbon-associated gas exchange improves oxygenation, lung mechanics, and survival in a model of adult respiratory distress syndrome. Crit Care Med 2004; 24:466–74
35.
van Kaam AH, de Jaegere A, Haitsma JJ, Van Aalderen WM, Kok JH, Lachmann B: Positive pressure ventilation with the open lung concept optimizes gas exchange and reduces ventilator-induced lung injury in newborn piglets. Pediatr Res 2003; 53:245–53
36.
Martynowicz MA, Walters BJ, Hubmayr RD: Mechanisms of recruitment in oleic acid-injured lungs. J Appl Physiol 2000; 90:1744–53
37.
Hickling KG: Best compliance during a decremental, but not incremental, positive end-expiratory pressure trial is related to open-lung positive end-expiratory pressure: A mathematical model of acute respiratory distress syndrome lungs. Am J Respir Crit Care Med 2001; 163:69–78
38.
Schmalisch G, Schmidt M, Proquitté H, Foitzik B, Rüdiger M, Wauer RR: Measurement of changes in respiratory mechanics during partial liquid ventilation using jet pulses. Crit Care Med 2004; 31:1435–41
39.
Davies MW, Dunster KR: Effect of perfluorocarbon (perfluorooctyl bromide) vapor on tidal volume measurement during partial liquid ventilation. Crit Care Med 2002; 30:1123–5
40.
Greenspan JS, Wofson MR, Shaffer TH: Liquid ventilation. Semin Perinatol 2000; 24:396–405
41.
East TD, in’t Veen JC, Pace NL, McJames S,: Functional residual capacity as a noninvasive indicator of optimal positive end-expiratory pressure. J Clin Monit 1988; 4:91–8
42.
Kloot TE, Blanch L, Youngblood MA, Weinert C, Adams AB, Marini JJ, Shapiro RS, Nahum A: Recruitment maneuvers in three experimental models of acute lung injury: Effect on lung volume and gas exchange. Am J Respir Crit Care Med 2000; 161:1485–94
43.
Woods CM, Neslund G, Kornbrust E, Flaim SF: Perflubron attenuates neutrophil adhesion to activated endothelial cells in vitro. Am J Physiol Lung Cell Mol Physiol 2000; 278:L1008–17
44.
Dreyfuss D, Martin-Lefevre L, Saumon G: Hyperinflation-induced lung injury during alveolar flooding in rats: Effect of perfluorocarbon instillation. Am J Respir Crit Care Med 1999; 159:1752–7
45.
Koch T, Ragaller M, Haufe D, Hofer A, Grosser M, Albrecht DM, Kotzsch M, Luther T: Perfluorohexane attenuates proinflammatory and procoagulatory response of activated monocytes and alveolar macrophages. Anesthesiology 2001; 94:101–9
46.
Gama de Abreu M, Wilmink B, Hübler M, Koch T: Vaporized perfluorohexane attenuates ventilator-induced lung injury in isolated, perfused rabbit lungs. Anesthesiology 2005; 102:597–605