We read with interest the article by Jeong et al.  1on cerebral oxygen saturation measured by near-infrared spectroscopy (NIRS) and jugular venous bulb oximetry during shoulder surgery in the beach chair position. Their results showed a significant lower jugular venous bulb oxygen saturation during beach chair position in the propofol–remifentanil group compared with the sevoflurane–nitrous oxide group. This difference between the two groups was not found for NIRS-measured cerebral oxygen saturation, which led the authors to conclude that NIRS cannot be recommended for assessing cerebral oxygenation during beach chair position.

The authors explain the lack of agreement between cerebral oxygen saturation measured by NIRS and measured with jugular bulb oximetry by emphasizing on the inherent limitations of the NIRS technology, such as the assumed constant arterial/venous ratio and the contamination by extracerebral blood flow and by other tissues that are not the tissues of interest. We would like to suggest an alternative explanation for the observations. In a recent study by Klein et al .,2it was demonstrated that propofol has a differential effect on the cerebral vessels, and that propofol preserves cerebral oxygen saturation in the frontal cortex, which is the measurement site of NIRS. Therefore, the finding that oxygen saturation measured by NIRS does not differ between the sevoflurane–nitrous oxide and the propofol–remifentanil group, and it does with jugular venous bulb oximetry, could be attributable to the fact that NIRS measures oxygen saturation in the frontal cortex, whereas jugular venous bulb oximetry measures global cerebral oxygen saturation.

Another concern relates to the fact that the authors used the original Bland-Altman method to assess whether cerebral oxygen saturations measured by jugular venous bulb oximetry or by NIRS were interchangeable. However, this Bland-Altman technique accounts for independent data, not for repeated measurements over a period of time, such as in this study. For both the linear regression and Bland-Altman analysis performed in this study, a technique that accounts for the within-patient–dependent data is mandatory. Consequently, the agreement estimation might have been obscured by the use of repeated-measures data.3 

*Ghent University Hospital, Ghent, Belgium. annelies.moerman@ugent.be

1.
Jeong H, Jeong S, Lim HJ, Lee J, Yoo KY. Cerebral oxygen saturation measured by near-infrared spectroscopy and jugular venous bulb oxygen saturation during arthroscopic shoulder surgery in beach chair position under sevoflurane-nitrous oxide or propofol-remifentanil anesthesia. ANESTHESIOLOGY. 2012;117:1047–56
2.
Klein KU, Fukui K, Schramm P, Stadie A, Fischer G, Werner C, Oertel J, Engelhard K. Human cerebral microcirculation and oxygen saturation during propofol-induced reduction of bispectral index. Br J Anaesth. 2011;107:735–41
3.
Myles PS, Cui J. Using the Bland-Altman method to measure agreement with repeated measures. Br J Anaesth. 2007;99:309–11