To the Editor:

In a randomized clinical trial, Rosenstock et al.1  showed no significant difference in time to awake intubation by experienced investigators using the McGrath video laryngoscope (MVL) compared with the fiberoptic bronchoscope (FOB) in difficult airway patients. Accordingly, the authors conclude that awake MVL intubation seems to be a potential alternative to awake fiberoptic intubation. However, an important issue ignored by them is that awake intubation actually includes two parts: airway topical anesthesia and subsequent intubation.2  Moreover, effective airway topical anesthesia is a prerequisite to successfully perform awake intubation.3  When adequate airway topical anesthesia is obtained, subsequent intubation is usually easy. To obtain a uniform airway topical anesthesia in the two groups, transtracheal injection of lidocaine was used in this study. This method is invasive and carries more potential risk than other topical anesthesia methods do. More importantly, it can be difficult or even impossible to perform if the patient’s neck anatomy is troublesome to locate.4  In this study, a total of seven patients were excluded because transtracheal injection was impossible.

In our view, a limitation of this study design is lack of assessment on the performance of airway topical anesthesia provided by the two devices. As a “gold standard” tool in managing difficult airway, FOB is not only a common choice for awake intubation, but can also provide flexibility in selectively anesthetizing the airway by a “spray as you go” technique.5  That is, two parts of the awake intubation can be completed with an FOB. In the Discussion section, the authors claim, “Awake MVL intubation may not prove as easy in using the ‘spray as you go’ technique, because insertion of the MVL blade causes pressure on the tongue and on the laryngeal structures, thereby probably creating a greater degree of patient discomfort compared with introducing the FOB.” It would be interesting to know whether there is any evidence to support the above comments. Had the authors performed airway topical anesthesia with the MVL?

The MVL has an anatomically shaped blade with an extra curve, and oropharyngeal tissues do not need to be retracted and compressed to achieve a straight line of sight during laryngoscopy with the MVL.6  Thus, there is usually no need for significant lifting force to visualize the glottis. It has been shown that the use of Glidescope video laryngoscope with an anatomically shaped blade creates less pressure on the tongue when compared with the Macintosh blade.7,8  After topical anesthesia of the tongue and pharynx with lidocaine spray, patients can well tolerate the MVL with minimal discomfort.9  In our experience, once the oropharyngeal mucosa is anesthetized by the method described in this study, the MVL can be advanced easily to a position in the hypopharynx where the epiglottis and larynx can be clearly visualized. At this point, aliquots of lidocaine can be sprayed using a MADgic® atomizer (Wolfe Tory Medical Inc., Salt Lake City, UT). The MADgic® atomizer is then advanced through the glottis into the larynx and trachea to spray further aliquots of lidocaine in the remaining airway. This modified spray-as-you-go technique with the video laryngoscope can provide excellent airway topical anesthesia and is less affected by secretions or blood compared with fibreoptic technique. It has been used successfully in difficult airway patients who undergo awake intubation with Glidescope video laryngoscope.10  All of these suggest that performing airway topical anesthesia under superior vision of the airway with a video laryngoscope on awake subjects is feasible. Unfortunately, there has been no randomized clinical study comparing video laryngoscopic and fiberoptic techniques of airway topical anesthesia. Before we have enough evidence to make a conclusion that the video laryngoscope is a useful alternative to the FOB for awake intubation, therefore, further studies are needed to evaluate and compare performances of both airway topical anesthesia and awake intubation in difficult airway patients. In such a study, other than the intubation time and success rate, the observed variables should also include the patient’s comfort during airway topical anesthesia and awake intubation, time required for airway topical anesthesia, awake intubating condition, possible difficulties and so forth.2,5 

References

1.
Rosenstock
CV
,
Thøgersen
B
,
Afshari
A
,
Christensen
AL
,
Eriksen
C
,
Gätke
MR
:
Awake fiberoptic or awake video laryngoscopic tracheal intubation in patients with anticipated difficult airway management: A randomized clinical trial.
Anesthesiology
2012
;
116
:
1210
6
2.
Xue
FS
,
He
N
,
Liao
X
,
Xu
XZ
,
Xu
YC
,
Yang
QY
,
Luo
MP
,
Zhang
YM
:
Clinical assessment of awake endotracheal intubation using the lightwand technique alone in patients with difficult airways.
Chin Med J
2009
;
122
:
408
15
3.
Kundra
P
,
Kutralam
S
,
Ravishankar
M
:
Local anaesthesia for awake fibreoptic nasotracheal intubation.
Acta Anaesthesiol Scand
2000
;
44
:
511
6
4.
Simmons
ST
,
Schleich
AR
:
Airway regional anesthesia for awake fiberoptic intubation.
Reg Anesth Pain Med
2002
;
27
:
180
92
5.
Xue
FS
,
Liu
HP
,
He
N
,
Xu
YC
,
Yang
QY
,
Liao
X
,
Xu
XZ
,
Guo
XL
,
Zhang
YM
:
Spray-as-you-go airway topical anesthesia in patients with a difficult airway: A randomized, double-blind comparison of 2% and 4% lidocaine.
Anesth Analg
2009
;
108
:
536
43
6.
Frerk
CM
,
Lee
G
:
Laryngoscopy: Time to change our view.
Anaesthesia
2009
;
64
:
351
4
7.
Russell
T
,
Lee
C
,
Firat
M
,
Cooper
RM
:
A comparison of the forces applied to a manikin during laryngoscopy with the GlideScope and Macintosh laryngoscopes.
Anaesth Intensive Care
2011
;
39
:
1098
102
8.
Carassiti
M
,
Zanzonico
R
,
Cecchini
S
,
Silvestri
S
,
Cataldo
R
,
Agrò
FE
:
Force and pressure distribution using Macintosh and GlideScope laryngoscopes in normal and difficult airways: A manikin study.
Br J Anaesth
2012
;
108
:
146
51
9.
Thong
SY
,
Shridhar
IU
,
Beevee
S
:
Evaluation of the airway in awake subjects with the McGrath videolaryngoscope.
Anaesth Intensive Care
2009
;
37
:
497
8
10.
Doyle
DJ
:
Awake intubation using the GlideScope video laryngoscope: Initial experience in four cases.
Can J Anaesth
2004
;
51
:
520
1