Background:

Intellivent is a new full closed-loop controlled ventilation that automatically adjusts both ventilation and oxygenation parameters. The authors compared gas exchange and breathing pattern variability of Intellivent and pressure support ventilation (PSV).

Methods:

In a prospective, randomized, single-blind design crossover study, 14 patients were ventilated during the weaning phase, with Intellivent or PSV, for two periods of 24 h in a randomized order. Arterial blood gases were obtained after 1, 8, 16, and 24 h with each mode. Ventilatory parameters were recorded continuously in a breath-by-breath basis during the two study periods. The primary endpoint was oxygenation, estimated by the calculation of the difference between the Pao2/Fio2 ratio obtained after 24 h of ventilation and the Pao2/Fio2 ratio obtained at baseline in each mode. The variability in the ventilatory parameters was also evaluated by the coefficient of variation (SD to mean ratio).

Results:

There were no adverse events or safety issues requiring premature interruption of both modes. The Pao2/Fio2 (mean ± SD) ratio improved significantly from 245 ± 75 at baseline to 294 ± 123 (P = 0.03) after 24 h of Intellivent. The coefficient of variation of inspiratory pressure and positive end-expiratory pressure (median [interquartile range]) were significantly higher with Intellivent, 16 [11–21] and 15 [7–23]%, compared with 6 [5–7] and 7 [5–10]% in PSV. Inspiratory pressure, positive end-expiratory pressure, and Fio2 changes were adjusted significantly more often with Intellivent compared with PSV.

Conclusions:

Compared with PSV, Intellivent during a 24-h period improved the Pao2/Fio2 ratio in parallel with more variability in the ventilatory support and more changes in ventilation settings.

What We Already Know about This Topic
  • The best method for weaning from mechanical ventilation has not been established

What This Article Tells Us That Is New
  • Full closed-loop controlled ventilation, in comparison with pressure support ventilation, improved oxygenation, ventilatory variability, and time spent in an adequate ventilation zone in critically ill patients.

PRESSURE support ventilation (PSV) is the most widely used assisted mode of ventilation during the weaning process, in both medical and surgical critically ill patients.1  However, PSV provides a fixed inspiratory pressure (PINSP), regardless of the patient’s ventilatory demand or gas exchange, which limits breathing pattern variability.2–5  Given the high variability in disease processes and states, the application of predefined, uniform values for ventilator parameters, such as a fixed PINSP or tidal volume (VT), is unlikely to provide optimal assist at all times.6  In contrast, variability in the breathing pattern may be useful in improving gas exchange as suggested by previous reports in animals7,8  or recently in humans.2 

New ventilatory modes can offer ventilation automatically adjusted to the patient’s ventilatory demand.2,5,9,10  Studies that evaluated these modes have shown benefits on the optimization of ventilation,2,11  burden of care,12,13  and duration of weaning.12,14  No automatic management of oxygenation (i.e., both fraction of inspired oxygen [Fio2] and of the positive end-expiratory pressure [PEEP]) was available to date.15,16 

Intellivent is a new full closed-loop solution for passive and active breathing patients receiving invasive mechanical ventilation and includes automatic adjustment of minute ventilation (MV), Fio2, and PEEP. Intellivent has been studied in sedated, passively ventilated critically ill adult patients with acute respiratory failure, but only for short duration (2–4 h).17  To our knowledge, no physiological study has been performed to evaluate Intellivent for a longer ventilation period in nonsedated actively breathing critically ill patients and during the weaning period.

The aim of this prospective, randomized, crossover study was to compare ventilatory parameters and gas exchange between Intellivent and PSV given more than 24 h, in critically ill patients, during the weaning phase.

Compared with PSV, which provides a fixed level of assistance (PINSP) regardless of the patient’s ventilatory demand, we hypothesized that Intellivent would improve oxygenation by offering more variable ventilation.18,19 

This single-site study was carried out in the 16-bed medical–surgical intensive care unit of the St Eloi Hospital, a 660-bed teaching and referral facility of the University of Montpellier in France. The experimental protocol was approved by the Ethics Committee of the Saint-Eloi Teaching Hospital (Comité de Protection des Personnes Sud Méditerranée IV, Montpellier, France), and written informed consent was provided by patient or next of kin. This study followed the CONSORT recommendations concerning randomized trial reporting.20 

Patients

From March 2011 to May 2011 (2.5 months), 50 consecutive patients were screened and 16 enrolled. The patients were included if they were in spontaneous mode (active patient, i.e., able to trigger a breath) with an expected duration of invasive mechanical ventilation longer than 48 h. Patients were not included in case of clinical instability, whatever the reason, and when a decision to withhold life-sustaining treatment was made. Pregnant women and children younger than 18 yr were also not included.

General Ventilator Settings

The two ventilation modes (PSV and Intellivent) were given by the same ventilator (Hamilton S1; Hamilton Medical, Rhäzuns, Switzerland). The rise time (50 ms), inspiratory flow trigger (2 l/min), and expiratory trigger sensitivity (30% of the peak flow) were identical in the two modes.

PSV

In PSV, the Fio2 was set to achieve a pulse oxygen saturation (Spo2) greater than 92% and the PEEP level was set between 5 and 10 cm H2O. The level of PINSP was adjusted to obtain a VT between 6 and 8 ml/kg of predicted body weight (PBW)21  (as calculated with the following formula for men: PBW (kg) = 50 + 2.3 ([height (cm)/2.54] − 60) and for women: PBW (kg) = 45.5 + 2.3 ([height (cm)/2.54] − 60) and a respiratory rate (RR) between 20 and 30 breaths/min.

Intellivent

In Intellivent, initial MV22  is automatically determined by the ventilator according to the PBW set by the clinician. The MV is automatically adjusted to maintain end-tidal partial pressure of carbon dioxide (Petco2) within expert-based acceptable ranges23  when the patient is not triggering the breath or to maintain the patient’s RR within acceptable ranges, as defined by the Otis least work of breathing concept,24  when the patient is triggering the breath. To adjust MV in order to maintain an acceptable range of Petco2 or RR, the ventilator adjusts both the VT and the RR as it is in adaptive support ventilation.4,25  In brief, based on the breath-by-breath expiratory time constant (RCEXP) estimation, optimal VT and RR are derived. When MV needs to be adjusted to keep the patient’s Petco2 or RR within the defined range, the ventilator adjusts PINSP and the mandatory breath to target optimal VT and RR to the patient (appendix 1).

To avoid extreme and potentially dangerous values of VT and RR, Intellivent uses, on a breath-by-breath basis, a safety window for the given VT and RR values. The minimal target VT is defined as twice the anatomical dead space estimated from the PBW. The maximal target VT is defined as the maximal pressure (set by the clinician) times the dynamic compliance of the total respiratory system. The minimal value for the target RR is 5 breaths/min. The maximal value for the target RR is defined as the ratio 20/RCEXP.26 

PEEP and Fio2 are automatically adjusted based on the ARDSnetwork PEEP–Fio2 tables27  to maintain an Spo2 within expert-based acceptable ranges (appendices 1 and 2). The tables are user adjustable by selecting the maximal PEEP delivered. The PEEP/Fio2 tables from the ARDSnetwork are used only as starting values in the algorithm; but adjustable by the user according to the local policies by setting the maximal PEEP value (appendix 2). Setting low maximal PEEP makes the algorithm adjusting the Fio2 more than the PEEP. In case of moderate decrease in Spo2, Fio2 increases by 10% of actual value every 30 s and PEEP increases by 1 cm H2O every 6 min. If Spo2 is above the target range, Fio2 decreases by 5% of the actual value every minute and PEEP decreases by 1 cm H2O every 6 min. If Petco2 and Spo2 informations are of poor quality or lost, the controllers automatically pause and an alarm is generated. Automatic control is resolved when signal of good quality is measured again. In addition, Fio2 is automatically increased to 100% if Spo2 is below 85%, and 100% Fio2 manual bypass is still available. In addition to the adaptive support ventilation safety windows, minimal and maximal MV, Fio2, and PEEP settings are set by the users before starting Intellivent. By adjusting, on a breath-by-breath basis, the level of PINSP, RR, PEEP, and Fio2, Intellivent may generate more variability than conventional ventilation such as PSV.

Protocol

We applied a prospective, randomized, single-blind crossover study design very similar to that previously reported.2,4,28  Determination of the first used ventilatory mode (PSV or Intellivent) was randomized. Randomization was carried out using a random-number table. Each patient was consecutively ventilated for 24 h with the PSV mode and with the Intellivent mode in a random order. Throughout the protocol, suctioning via the endotracheal tube was performed on a per need basis and routine care, such as physiotherapy and nursing was performed as usual in the unit.

Measurements

Standard three-lead monitoring electrodes continuously monitored heart rate and rhythm. Spo2 was continuously monitored using pulse oximetry. Systolic and diastolic arterial blood pressures were continuously monitored through a 20-gauge catheter inserted in a radial or femoral artery. Blood samples were obtained at baseline (in the first hour after mechanical ventilation for each mode), after 8, 16, and 24 h of mechanical ventilation for arterial blood gas analysis (GEM Premier 3000 analyzer; Instrumentation Laboratory, Lexington, MA) through the arterial catheter.

The following variables such as airway pressures (PINSP as inspiratory pressure level above PEEP delivered by the ventilator, PMEAN as mean airway pressure, and PMAX as maximal airway pressure), VT, RR, MV, RCEXP, PEEP, Fio2, and Petco2 were collected continuously breath-by-breath, by a dedicated software (Study recorder software; Hamilton Medical) via a RS32 cable, exported through a Universal Serial Bus support and analyzed using a customized software based on Microsoft Excel®(Redmond, WA).

Every 8 h, according to the unit protocol, the nurse in charge of the patient evaluated the pain and comfort using the Behavioral Pain Scale and the sedation and agitation level using the Richmond Agitation Sedation Scale.29,30  Setting changes made by the attending physician were also recorded.

Sample Size and Statistical Analysis

The primary endpoint was oxygenation, estimated by the calculation of the difference between the Pao2/Fio2 ratio obtained after 24 h of ventilation and the Pao2/Fio2 ratio obtained at baseline in each mode. To calculate the number of patients needed, we used data previously published2,31  showing in postoperative patients a mean Pao2/Fio2 ratio of 202 ± 48 mmHg in PSV. Assuming an α risk of 0.05 and a β risk of 0.20, we calculated that at least 14 patients would be required to identify, after 24 h of mechanical ventilation, a difference of 20% between the variation of the Pao2/Fio2 ratio obtained in Intellivent in comparison to PSV. Therefore, we decided to include 16 patients. The secondary endpoints were the variability in the ventilation variables and the time spent with acceptable ventilation. Ventilation variables were collected continuously breath-by-breath during 24 h (see above). The variability in the ventilation parameters was evaluated by the coefficients of variation for PINSP, RR, VT, MV, PEEP, Fio2, RCEXP, and Petco2 calculated as the ratio of the SD to the mean multiplied by 100 as previously described.2,32  Acceptable ventilation was defined with very permissive ranges, and calculated as the number of breath with RR between 12 and 35 breaths/min, a VT between 5 and 12 ml/kg of PBW, and Petco2 less than 55 mmHg, over the total number of breath collected.2,27,28  Values are expressed as median [interquartile range, IQR] or mean ± SD according to the type of variable distribution, from data collected breath-by-breath for 24 h. Normality of the distribution was assessed with Kolmogorov–Smirnov test. Comparisons were performed using Wilcoxon and Mann–Whitney tests according, and two-tailed P values less than 0.05 were considered significant. Statistical analysis was performed using SAS/STAT software version 8.1 (SAS Institute, Cary, NC) by an independent statistician.

Among the 16 enrolled patients, 2 did not complete the study because of early extubation, and 14 patients were finally analyzed (fig. 1). There were no safety issues requiring premature interruption of Intellivent for the studied patients. Diagnosis at the time of admission in intensive care unit and clinical characteristics of the patients are shown in table 1. Ventilation settings and main monitored parameters obtained during the first hour after inclusion were shown in table 2.

Fig. 1.

Trial profile. PSV = pressure support ventilation.

Fig. 1.

Trial profile. PSV = pressure support ventilation.

Close modal
Table 1.

Characteristics of the 14 Patients Studied

Characteristics of the 14 Patients Studied
Characteristics of the 14 Patients Studied
Table 2.

Ventilation Settings and Main Monitored Parameters Obtained the First Hour after Inclusion in PSV and Intellivent

Ventilation Settings and Main Monitored Parameters Obtained the First Hour after Inclusion in PSV and Intellivent
Ventilation Settings and Main Monitored Parameters Obtained the First Hour after Inclusion in PSV and Intellivent

Arterial blood gases were shown in the table 3. The Pao2/Fio2 ratio improved significantly from 245 ± 75 at baseline to 294 ± 123 mmHg (P = 0.035) after 24 h of Intellivent, whereas no significant change was observed with PSV (fig. 2). The Pao2/Fio2 ratio variation after 24 h of mechanical ventilation was significantly higher in Intellivent than in PSV mode (+18 ± 32% vs. −3 ± 20%; P = 0.026).

Fig. 2.

Individual variations in Pao2/Fio2 ratio for the 14 patients after mechanical ventilation with pressure support ventilation (PSV; A) and with Intellivent (B). The horizontal bars represent the mean values. P value refers to the test of the first versus the last time point. Fio2 = inspired oxygen fraction; NS = not significant; Pao2 = partial pressure of arterial oxygen.

Fig. 2.

Individual variations in Pao2/Fio2 ratio for the 14 patients after mechanical ventilation with pressure support ventilation (PSV; A) and with Intellivent (B). The horizontal bars represent the mean values. P value refers to the test of the first versus the last time point. Fio2 = inspired oxygen fraction; NS = not significant; Pao2 = partial pressure of arterial oxygen.

Close modal
Table 3.

Arterial Blood Gases between PSV and Intellivent

Arterial Blood Gases between PSV and Intellivent
Arterial Blood Gases between PSV and Intellivent

The 24-h average values of the ventilation parameters are reported in table 4. VT was 7.6 ml/kg [IQR, 6.6–9.0] during PSV period compared with 8.4 ml/kg [IQR, 7.9–8.6] PBW during Intellivent period (P = 0.04). The RR was 22 breaths/min [IQR, 19–27] during PSV period compared with 19 breaths/min [IQR, 15–22] during Intellivent period (P = 0.01). Typical tracings obtained during 24 h of both PSV and Intellivent are shown in figure 3. There is obviously more variability in PINSP, PEEP, and Fio2 with Intellivent as compared with PSV. The coefficient of variation of PINSP and PEEP was significantly higher with Intellivent as compared with PSV (table 4).

Fig. 3.

Experimental records that help in illustrating the effects of the two ventilatory modes during 24 h of mechanical ventilation with pressure support ventilation (PSV; A) and with Intellivent (B) in a representative patient. From top to bottom, inspiratory pressure above end-expiratory pressure (PINSP), mean airway pressure (Pmean), positive end-expiratory pressure (PEEP), fraction of inspired oxygen (Fio2), pulsatile oxygen saturation (Spo2), and tidal volume (VT). Note that PINSP, Pmean, PEEP, and Fio2 are more variable in Intellivent than in PSV.

Fig. 3.

Experimental records that help in illustrating the effects of the two ventilatory modes during 24 h of mechanical ventilation with pressure support ventilation (PSV; A) and with Intellivent (B) in a representative patient. From top to bottom, inspiratory pressure above end-expiratory pressure (PINSP), mean airway pressure (Pmean), positive end-expiratory pressure (PEEP), fraction of inspired oxygen (Fio2), pulsatile oxygen saturation (Spo2), and tidal volume (VT). Note that PINSP, Pmean, PEEP, and Fio2 are more variable in Intellivent than in PSV.

Close modal
Table 4.

Ventilation Parameters with PSV and with Intellivent

Ventilation Parameters with PSV and with Intellivent
Ventilation Parameters with PSV and with Intellivent

Times spent in different ranges of VT, RR, and Petco2 in both modalities are shown in figure 4. The number of changes in PINSP, PEEP, and Fio2 was significantly higher with Intellivent compared with PSV (table 5). No significant differences were observed between the two modes for the Behavioral Pain Scale and Richmond Agitation Sedation Scale scores over the study period.

Fig. 4.

Contributions to inadequate ventilation of low tidal volume (VT) of <5 ml/kg of predicted body weight (PBW), high VT of >12 ml/kg PBW, low respiratory rate (RR) of <12 breaths/min, high RR of >35 breaths/min, and high end-tidal partial pressure of carbon dioxide (Petco2) of >55 mmHg during the 24 h of pressure support ventilation (PSV, A) and 24 h of Intellivent (B) in the 14 studied patients. A patient can be in adequate ventilation for low RR and high VT, at the same time that may explain why patient number 11 spent more than 100% of the time in inadequate ventilation with Intellivent. With PSV, inadequate ventilation represented 17.1% [interquartile range, 4.3–39.6] of the total ventilation duration in this mode; with Intellivent, inadequate ventilation represented 3.7% [interquartile range, 1.8–10.1] of the total ventilation duration in this mode.

Fig. 4.

Contributions to inadequate ventilation of low tidal volume (VT) of <5 ml/kg of predicted body weight (PBW), high VT of >12 ml/kg PBW, low respiratory rate (RR) of <12 breaths/min, high RR of >35 breaths/min, and high end-tidal partial pressure of carbon dioxide (Petco2) of >55 mmHg during the 24 h of pressure support ventilation (PSV, A) and 24 h of Intellivent (B) in the 14 studied patients. A patient can be in adequate ventilation for low RR and high VT, at the same time that may explain why patient number 11 spent more than 100% of the time in inadequate ventilation with Intellivent. With PSV, inadequate ventilation represented 17.1% [interquartile range, 4.3–39.6] of the total ventilation duration in this mode; with Intellivent, inadequate ventilation represented 3.7% [interquartile range, 1.8–10.1] of the total ventilation duration in this mode.

Close modal
Table 5.

Number of Changes in PINSP, PEEP, and Fio2 during PSV and Intellivent

Number of Changes in PINSP, PEEP, and Fio2 during PSV and Intellivent
Number of Changes in PINSP, PEEP, and Fio2 during PSV and Intellivent

The current study demonstrates that over a 24-h study period, ventilation with Intellivent was associated with a higher Pao2/Fio2 ratio and a greater variability in PINSP and PEEP compared with PSV. In addition, the study suggests that the use of Intellivent for 24 h of mechanical ventilation is feasible and safe for critically ill patients during the weaning period.

The current study is the first to report long-term (i.e., 24-h period) safe use of the Intellivent mode and to associate its use with improvement in oxygenation. The improvement in oxygenation observed with Intellivent is probably not related to a single mechanism, but we could speculate that it is because of more complex association of different features of Intellivent. Some features related to Intellivent can be proposed such as a slight, but significantly, higher airway pressures (PINSP and PMAX, not PMEAN) leading a higher VT (table 4), an increase in variability of inspiratory pressures and PEEP, which may assimilated to more physiological sigh, and be considered as repeated alveolar auto-recruitment. Although, median VT was significantly higher in Intellivent than in PSV (8.4 [IQR, 7.9–8.6] vs. 7.6 [IQR, 6.6–9.0] ml/kg PBW; P = 0.04; table 4), VT remained lower than 10 ml/kg PBW more than 90% of the time spent in each modes with no significant difference between both Intellivent and PSV (fig. 4). Indeed, risk factors to develop acute lung injury or acute respiratory distress syndrom in passive ventilated patients are VT above 10 ml/kg PBW and an end-inspiratory pressure above 30 cm H2O. In the current study, no patient was passively ventilated, and the association of a VT above 10 ml/kg PBW and an end-inspiratory pressure above 30 cm H2O occurred together exceptionally either in Intellivent or PSV. Low VT with subsequently high transpulmonary pressure may basically be more dangerous than higher VT with lower transpulmonary pressure. In combination with more variability in airway pressures (PINSP, PMAX, and PMEAN) and in PEEP with Intellivent, it suggests causality between Intellivent-induced variability and improvement in oxygenation. Several publications have already reported better oxygenation when pressure or volume applied to the respiratory system is variable, whatever the mode of ventilation being used.2,7,8,33  The current study can only speculate on the mechanism responsible for such improvement. The Jensen inequality, i.e., the local nonlinear pressure–volume relationship, has been suggested as a mechanistic explanation,34,35  heterogeneity in local time constant,17,36  improved surfactant production,37  and improved ventilation–perfusion matching38  may also play a role in better oxygenation with more variable ventilation.

More ventilator adjustments were made during Intellivent ventilation compared with PSV. Although expected with closed-loop systems, which are by design able to continuously adjust the ventilatory parameters according to the changes in patient’s condition, it is worthwhile discussing the possible impact of adjusting the ventilator more often. First, as already discussed earlier, more adjustment makes ventilation more variable, and this may have clinical impacts (such as oxygenation in the current study). Second, it may help to maintain the patient within predefined acceptable ventilation ranges. In the current study, time spent by the patient with acceptable ventilation during Intellivent was not very different compared with PSV (fig. 4). However, during Intellivent, 11 of 14 patients (76%) spent more than 90% of ventilation time in acceptable ventilation (as defined in the methodology section) as compared with only 5 patients with PSV (P = 0.04). In other studies,39,40  Intellivent was able to keep the patient more often with acceptable ventilation. In these studies,39,40  most of the patients were not triggering the breath and were therefore fully controlled by the ventilator. In the current study, all patients were able to trigger the breath and to control their breathing pattern at least partially. The definition of “acceptable ventilation”, which was relatively permissive in the current study (because of the selected population), is indeed also important in analyzing the results and can be extensively discussed. Finally, although not analyzed in the current study, the number of ventilator adjustments with Intellivent may reflect how often the ventilator should be adjusted, whereas the number of changes in PSV may reflect basically how often the ventilator can be adjusted manually considering human resources and knowledge available at the bedside day and night. The current study is definitely not able to draw any conclusions on possible clinical impact of continuous and more frequent adjustments of the ventilator but give enough confidence to design large randomized controlled trials to address such a question. In addition, the current study is the first to report the use of Intellivent for more than couple of hours, in adult patients with variable conditions and during the weaning phase.

All patients were able to complete the study, and no patients were removed from Intellivent for major safety issues, suggesting that 24-h use of Intellivent may be safe in critically ill patients during the weaning period.

This study has some limitations. The study was not designed to evaluate the safeness and effectiveness of Intellivent as a routine mode of ventilation in patients in intensive care unit and therefore is underpowered for that. Interestingly, in three patients, Fio2 had to be adjusted manually because of a discrepancy between Spo2 and Sao2 obtained from arterial blood sampling. However, because of the population selection (few hypoxemic patients), the Fio2 and PEEP algorithms and the robustness of Spo2 information (filtering, artifact, and motion rejections, and so on) for running the loops were not really challenged in the current study. Moreover, although we evaluated the agitation and sedation–analgesia levels every 4 h (using Richmond Agitation Sedation Scale and Behavioral Pain Scale scores), there was no specific auto-evaluation of the ventilatory comfort.

The current prospective study is the first to report 24-h use of Intellivent in spontaneously breathing patients during the weaning process. As compared with PSV, the Pao2/Fio2 ratio at 24 h was improved in Intellivent with more variability in airway pressures, PEEP, and Fio2. Adjustment of the ventilator was much more frequent with Intellivent as compared with PSV, which may explain the variability and ultimately the better oxygenation observed with Intellivent. The current study definitely warrants further prospective controlled studies to estimate the potential clinical impact of Intellivent as compared with conventional modes of ventilation.

Appendix 1. Ventilation and Oxygenation Controllers

Fig. 5.

Ventilation controller: First of all, the system detected whether the patient is spontaneously breathing (based on consecutive number of breaths triggered by the patient). If the patient is not spontaneously breathing, the regulation is based on end-tidal partial pressure of carbon dioxide (Petco2) and if the patient’s Petco2 is outside the target ranges, the controller adjusts minute ventilation (MV): increasing if Petco2 is above the target range and decreasing if Petco2 is below the target range. If the patient is spontaneously breathing, the regulation is based on the respiratory rate (RR) and if the patient’s RR (RRspont) is outside the target range, MV is adjusted: increasing MV if the patient RR is above the target range and decreasing if the patient RR is below the target range. In both situations, changes in MV are going through the adaptive support ventilation (ASV) controller, which is deciding whether the mandatory RR (RRmand) or the level of inspiratory pressure (PINSP) should be adjusted. There is an adjustment of the Petco2 target range based on the level of PINSP: the higher the PINSP, the more permissive the target ranges. There is also an adjustment of the RR target range based on MV: the higher the MV, the wider that the RR target range. Oxygenation controller: The patient’s pulsatile oxygen saturation (Spo2) value is compared with Spo2 target range, and if the patient’s Spo2 is outside the target, the controller is adjusting the fraction of inspired oxygen (Fio2) or the positive end-expiratory pressure (PEEP), depending on the selector (S). The choice between PEEP and Fio2 is based on a predefined PEEP–Fio2 table which is adjustable depending on the user-set maximal PEEP. There is an adjustment of the target Spo2 depending on the PEEP value: the higher the PEEP, the more permissive is the target. In case of moderate decrease in Spo2, Fio2 increases by 10% of actual value every 30 s and PEEP increases by 1 cm H2O every 6 min. If Spo2 is above the target range, Fio2 decreases by 5% of the actual value every minute and PEEP decreases by 1 cm H2O every 6 min. A minimal PEEP level can be set by the user which may limit in some cases a rapid decrease in PEEP which may result in derecruitment over time. If Petco2 and Spo2 informations are of poor quality or lost, the controllers automatically pause and an alarm is generated. Automatic control is resolved when signal of good quality is measured again. In addition, Fio2 is automatically increased to 100% if Spo2 is below 85%, and 100% Fio2 manual bypass is still available.

Fig. 5.

Ventilation controller: First of all, the system detected whether the patient is spontaneously breathing (based on consecutive number of breaths triggered by the patient). If the patient is not spontaneously breathing, the regulation is based on end-tidal partial pressure of carbon dioxide (Petco2) and if the patient’s Petco2 is outside the target ranges, the controller adjusts minute ventilation (MV): increasing if Petco2 is above the target range and decreasing if Petco2 is below the target range. If the patient is spontaneously breathing, the regulation is based on the respiratory rate (RR) and if the patient’s RR (RRspont) is outside the target range, MV is adjusted: increasing MV if the patient RR is above the target range and decreasing if the patient RR is below the target range. In both situations, changes in MV are going through the adaptive support ventilation (ASV) controller, which is deciding whether the mandatory RR (RRmand) or the level of inspiratory pressure (PINSP) should be adjusted. There is an adjustment of the Petco2 target range based on the level of PINSP: the higher the PINSP, the more permissive the target ranges. There is also an adjustment of the RR target range based on MV: the higher the MV, the wider that the RR target range. Oxygenation controller: The patient’s pulsatile oxygen saturation (Spo2) value is compared with Spo2 target range, and if the patient’s Spo2 is outside the target, the controller is adjusting the fraction of inspired oxygen (Fio2) or the positive end-expiratory pressure (PEEP), depending on the selector (S). The choice between PEEP and Fio2 is based on a predefined PEEP–Fio2 table which is adjustable depending on the user-set maximal PEEP. There is an adjustment of the target Spo2 depending on the PEEP value: the higher the PEEP, the more permissive is the target. In case of moderate decrease in Spo2, Fio2 increases by 10% of actual value every 30 s and PEEP increases by 1 cm H2O every 6 min. If Spo2 is above the target range, Fio2 decreases by 5% of the actual value every minute and PEEP decreases by 1 cm H2O every 6 min. A minimal PEEP level can be set by the user which may limit in some cases a rapid decrease in PEEP which may result in derecruitment over time. If Petco2 and Spo2 informations are of poor quality or lost, the controllers automatically pause and an alarm is generated. Automatic control is resolved when signal of good quality is measured again. In addition, Fio2 is automatically increased to 100% if Spo2 is below 85%, and 100% Fio2 manual bypass is still available.

Close modal

Appendix 2. PEEP–Fio2 Table Algorithm

Fig. 6.

Positive end-expiratory pressure (PEEP) and fraction of inspired oxygen (Fio2) are automatically adjusted based on the ARDSnetwork PEEP–Fio2 tables27,41  to maintain pulsatile oxygen saturation (Spo2) within expert-based acceptable ranges. The tables are user adjustable by selecting the maximal PEEP delivered. The PEEP/Fio2 tables from the ARDSnetwork are used only as starting values in the algorithm; but adjustable by the user according to the local policies by setting the maximal PEEP value.

Fig. 6.

Positive end-expiratory pressure (PEEP) and fraction of inspired oxygen (Fio2) are automatically adjusted based on the ARDSnetwork PEEP–Fio2 tables27,41  to maintain pulsatile oxygen saturation (Spo2) within expert-based acceptable ranges. The tables are user adjustable by selecting the maximal PEEP delivered. The PEEP/Fio2 tables from the ARDSnetwork are used only as starting values in the algorithm; but adjustable by the user according to the local policies by setting the maximal PEEP value.

Close modal
1.
Peñuelas
O
,
Frutos-Vivar
F
,
Fernández
C
,
Anzueto
A
,
Epstein
SK
,
Apezteguía
C
,
González
M
,
Nin
N
,
Raymondos
K
,
Tomicic
V
,
Desmery
P
,
Arabi
Y
,
Pelosi
P
,
Kuiper
M
,
Jibaja
M
,
Matamis
D
,
Ferguson
ND
,
Esteban
A
;
Ventila Group
:
Characteristics and outcomes of ventilated patients according to time to liberation from mechanical ventilation.
Am J Respir Crit Care Med
2011
;
184
:
430
7
2.
Coisel
Y
,
Chanques
G
,
Jung
B
,
Constantin
JM
,
Capdevila
X
,
Matecki
S
,
Grasso
S
,
Jaber
S
:
Neurally adjusted ventilatory assist in critically ill postoperative patients: A crossover randomized study.
Anesthesiology
2010
;
113
:
925
35
3.
Jaber
S
,
Delay
JM
,
Matecki
S
,
Sebbane
M
,
Eledjam
JJ
,
Brochard
L
:
Volume-guaranteed pressure-support ventilation facing acute changes in ventilatory demand.
Intensive Care Med
2005
;
31
:
1181
8
4.
Jaber
S
,
Sebbane
M
,
Verzilli
D
,
Matecki
S
,
Wysocki
M
,
Eledjam
JJ
,
Brochard
L
:
Adaptive support and pressure support ventilation behavior in response to increased ventilatory demand.
Anesthesiology
2009
;
110
:
620
7
5.
Ranieri
VM
,
Grasso
S
,
Mascia
L
,
Martino
S
,
Fiore
T
,
Brienza
A
,
Giuliani
R
:
Effects of proportional assist ventilation on inspiratory muscle effort in patients with chronic obstructive pulmonary disease and acute respiratory failure.
Anesthesiology
1997
;
86
:
79
91
6.
Thille
AW
,
Cabello
B
,
Galia
F
,
Lyazidi
A
,
Brochard
L
:
Reduction of patient-ventilator asynchrony by reducing tidal volume during pressure-support ventilation.
Intensive Care Med
2008
;
34
:
1477
86
7.
Gama de Abreu
M
,
Spieth
PM
,
Pelosi
P
,
Carvalho
AR
,
Walter
C
,
Schreiber-Ferstl
A
,
Aikele
P
,
Neykova
B
,
Hubler
M
,
Koch
T
:
Noisy pressure support ventilation: A pilot study on a new assisted ventilation mode in experimental lung injury.
Crit Care Med
2008
;
36
:
818
27
8.
Spieth
PM
,
Carvalho
AR
,
Güldner
A
,
Pelosi
P
,
Kirichuk
O
,
Koch
T
,
de Abreu
MG
:
Effects of different levels of pressure support variability in experimental lung injury.
Anesthesiology
2009
;
110
:
342
50
9.
Burns
KE
,
Lellouche
F
,
Lessard
MR
:
Automating the weaning process with advanced closed-loop systems.
Intensive Care Med
2008
;
34
:
1757
65
10.
Lellouche
F
,
Brochard
L
:
Advanced closed loops during mechanical ventilation (PAV, NAVA, ASV, SmartCare).
Best Pract Res Clin Anaesthesiol
2009
;
23
:
81
93
11.
Arnal
JM
,
Wysocki
M
,
Nafati
C
,
Donati
S
,
Granier
I
,
Corno
G
,
Durand-Gasselin
J
:
Automatic selection of breathing pattern using adaptive support ventilation.
Intensive Care Med
2008
;
34
:
75
81
12.
Gruber
PC
,
Gomersall
CD
,
Leung
P
,
Joynt
GM
,
Ng
SK
,
Ho
KM
,
Underwood
MJ
:
Randomized controlled trial comparing adaptive-support ventilation with pressure-regulated volume-controlled ventilation with automode in weaning patients after cardiac surgery.
Anesthesiology
2008
;
109
:
81
7
13.
Sulzer
CF
,
Chioléro
R
,
Chassot
PG
,
Mueller
XM
,
Revelly
JP
:
Adaptive support ventilation for fast tracheal extubation after cardiac surgery: A randomized controlled study.
Anesthesiology
2001
;
95
:
1339
45
14.
Lellouche
F
,
Mancebo
J
,
Jolliet
P
,
Roeseler
J
,
Schortgen
F
,
Dojat
M
,
Cabello
B
,
Bouadma
L
,
Rodriguez
P
,
Maggiore
S
,
Reynaert
M
,
Mersmann
S
,
Brochard
L
:
A multicenter randomized trial of computer-driven protocolized weaning from mechanical ventilation.
Am J Respir Crit Care Med
2006
;
174
:
894
900
15.
Raemer
DB
,
Ji
XB
,
Topulos
GP
:
FIx controller: An instrument to automatically adjust inspired oxygen fraction using feedback control from a pulse oximeter.
J Clin Monit
1997
;
13
:
91
101
16.
Urschitz
MS
,
Horn
W
,
Seyfang
A
,
Hallenberger
A
,
Herberts
T
,
Miksch
S
,
Popow
C
,
Müller-Hansen
I
,
Poets
CF
:
Automatic control of the inspired oxygen fraction in preterm infants: A randomized crossover trial.
Am J Respir Crit Care Med
2004
;
170
:
1095
100
17.
Arnal
JM
,
Paquet
J
,
Wysocki
M
,
Demory
D
,
Donati
S
,
Granier
I
,
Corno
G
,
Durand-Gasselin
J
:
Optimal duration of a sustained inflation recruitment maneuver in ARDS patients.
Intensive Care Med
2011
;
37
:
1588
94
18.
Mutch
WA
,
Eschun
GM
,
Kowalski
SE
,
Graham
MR
,
Girling
LG
,
Lefevre
GR
:
Biologically variable ventilation prevents deterioration of gas exchange during prolonged anaesthesia.
Br J Anaesth
2000
;
84
:
197
203
19.
Tobin
MJ
,
Mador
MJ
,
Guenther
SM
,
Lodato
RF
,
Sackner
MA
:
Variability of resting respiratory drive and timing in healthy subjects.
J Appl Physiol
1988
;
65
:
309
17
20.
Boutron
I
,
Moher
D
,
Altman
DG
,
Schulz
KF
,
Ravaud
P
;
CONSORT Group
:
Extending the CONSORT statement to randomized trials of nonpharmacologic treatment: Explanation and elaboration.
Ann Intern Med
2008
;
148
:
295
309
21.
Bigatello
LM
,
Pesenti
A
:
Ventilator-induced lung injury: Less ventilation, less injury.
Anesthesiology
2009
;
111
:
699
700
22.
Jolliet
P
,
Watremez
C
,
Roeseler
J
,
Ngengiyumva
JC
,
de Kock
M
,
Clerbaux
T
,
Tassaux
D
,
Reynaert
M
,
Detry
B
,
Liistro
G
:
Comparative effects of helium-oxygen and external positive end-expiratory pressure on respiratory mechanics, gas exchange, and ventilation-perfusion relationships in mechanically ventilated patients with chronic obstructive pulmonary disease.
Intensive Care Med
2003
;
29
:
1442
50
23.
Wysocki
M
,
Arnal
J
,
Feissel
M
,
l’Her
E
,
Roeseler
J
,
Lellouche
F
:
What ventilation and blood gases are “desirable” in adult mechanically ventilated patients.
Intensive care medicine
2010
, pp
Abstract 136
24.
Laubscher
TP
,
Heinrichs
W
,
Weiler
N
,
Hartmann
G
,
Brunner
JX
:
An adaptive lung ventilation controller.
IEEE Trans Biomed Eng
1994
;
41
:
51
9
25.
Jung
B
,
Constantin
JM
,
Rossel
N
,
Le Goff
C
,
Sebbane
M
,
Coisel
Y
,
Chanques
G
,
Futier
E
,
Hugon
G
,
Capdevila
X
,
Petrof
B
,
Matecki
S
,
Jaber
S
:
Adaptive support ventilation prevents ventilator-induced diaphragmatic dysfunction in piglet: An in vivo and in vitro study.
Anesthesiology
2010
;
112
:
1435
43
26.
Tassaux
D
,
Dalmas
E
,
Gratadour
P
,
Jolliet
P
:
Patient-ventilator interactions during partial ventilatory support: A preliminary study comparing the effects of adaptive support ventilation with synchronized intermittent mandatory ventilation plus inspiratory pressure support.
Crit Care Med
2002
;
30
:
801
7
27.
ARDS-Network
:
Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. The ARDS network.
N Engl J Med
2000
;
342
:
1301
8
28.
Dojat
M
,
Harf
A
,
Touchard
D
,
Lemaire
F
,
BrochardV
L
:
Clinical evaluation of a computer-controlled pressure support mode.
Am J Respir Crit Care Med
2000
;
161
(
4 Pt 1
):
1161
6
29.
Chanques
G
,
Payen
JF
,
Mercier
G
,
de Lattre
S
,
Viel
E
,
Jung
B
,
Cissé
M
,
Lefrant
JY
,
Jaber
S
:
Assessing pain in non-intubated critically ill patients unable to self report: An adaptation of the Behavioral Pain Scale.
Intensive Care Med
2009
;
35
:
2060
7
30.
Chanques
G
,
Sebbane
M
,
Barbotte
E
,
Viel
E
,
Eledjam
JJ
,
Jaber
S
:
A prospective study of pain at rest: Incidence and characteristics of an unrecognized symptom in surgical and trauma versus medical intensive care unit patients.
Anesthesiology
2007
;
107
:
858
60
31.
Jaber
S
,
Sebbane
M
,
Koechlin
C
,
Hayot
M
,
Capdevila
X
,
Eledjam
JJ
,
Prefaut
C
,
Ramonatxo
M
,
Matecki
S
:
Effects of short vs. prolonged mechanical ventilation on antioxidant systems in piglet diaphragm.
Intensive Care Med
2005
;
31
:
1427
33
32.
Wysocki
M
,
Cracco
C
,
Teixeira
A
,
Mercat
A
,
Diehl
JL
,
Lefort
Y
,
Derenne
JP
,
Similowski
T
:
Reduced breathing variability as a predictor of unsuccessful patient separation from mechanical ventilation.
Crit Care Med
2006
;
34
:
2076
83
33.
Sydow
M
,
Burchardi
H
,
Ephraim
E
,
Zielmann
S
,
Crozier
TA
:
Long-term effects of two different ventilatory modes on oxygenation in acute lung injury. Comparison of airway pressure release ventilation and volume-controlled inverse ratio ventilation.
Am J Respir Crit Care Med
1994
;
149
:
1550
6
34.
Brewster
JF
,
Graham
MR
,
Mutch
WA
:
Convexity, Jensen’s inequality and benefits of noisy mechanical ventilation.
J R Soc Interface
2005
;
2
:
393
6
35.
Suki
B
,
Alencar
AM
,
Sujeer
MK
,
Lutchen
KR
,
Collins
JJ
,
Andrade
JS
Jr
,
Ingenito
EP
,
Zapperi
S
,
Stanley
HE
:
Life-support system benefits from noise.
Nature
1998
;
393
:
127
8
36.
Ma
B
,
Suki
B
,
Bates
JH
:
Effects of recruitment/derecruitment dynamics on the efficacy of variable ventilation.
J Appl Physiol
2011
;
110
:
1319
26
37.
Arold
SP
,
Suki
B
,
Alencar
AM
,
Lutchen
KR
,
Ingenito
EP
:
Variable ventilation induces endogenous surfactant release in normal guinea pigs.
Am J Physiol Lung Cell Mol Physiol
2003
;
285
:
L370
5
38.
Carvalho
AR
,
Spieth
PM
,
Pelosi
P
,
Beda
A
,
Lopes
AJ
,
Neykova
B
,
Heller
AR
,
Koch
T
,
Gama de Abreu
M
:
Pressure support ventilation and biphasic positive airway pressure improve oxygenation by redistribution of pulmonary blood flow.
Anesth Analg
2009
;
109
:
856
65
39.
Arnal
JM
,
Wysocki
M
,
Novotni
D
,
Demory
D
,
Lopez
R
,
Donati
S
,
Granier
I
,
Corno
G
,
Durand-Gasselin
J
:
Safety and efficacy of a fully closed-loop control ventilation (IntelliVent-ASV) in ICU patients: A prospective randomized cross-over study.
Intensive Care Med
2012
;
38
:
781
7
40.
Lellouche
F
,
Bouchard
P
,
Laubscher
T
,
Blackburn
S
,
L’Her
E
,
Wysocki
M
:
Prospective randomized controlled study comparing conventional ventilation versus a fully closed-loop ventilation (IntelliVent®) in post cardiac surgery ICU patients.
Intensive care medicine
2010
, pp
Abstract 36
41.
Brower
RG
,
Lanken
PN
,
MacIntyre
N
,
Matthay
MA
,
Morris
A
,
Ancukiewicz
M
,
Schoenfeld
D
,
Thompson
BT
;
National Heart, Lung, and Blood Institute ARDS Clinical Trials Network
:
Higher versus lower positive end-expiratory pressures in patients with the acute respiratory distress syndrome.
N Engl J Med
2004
;
351
:
327
36