Abstract
Compared to pneumatically controlled pressure support (PSP), neurally adjusted ventilatory assist (NAVA) was proved to improve patient–ventilator interactions, while not affecting comfort, diaphragm electrical activity (EAdi), and arterial blood gases (ABGs). This study compares neurally controlled pressure support (PSN) with PSP and NAVA, delivered through two different helmets, in hypoxemic patients receiving noninvasive ventilation for prevention of extubation failure.
Fifteen patients underwent three (PSP, NAVA, and PSN) 30-min trials in random order with both helmets. Positive end-expiratory pressure was always set at 10 cm H2O. In PSP, the inspiratory support was set at 10 cm H2O above positive end-expiratory pressure. NAVA was adjusted to match peak EAdi (EAdipeak) during PSP. In PSN, the NAVA level was set at maximum matching the pressure delivered during PSP by limiting the upper pressure. The authors assessed patient comfort, EAdipeak, rates of pressurization (i.e., airway pressure-time product [PTP] of the first 300 and 500 ms after the initiation of patient effort, indexed to the ideal pressure–time products), and measured ABGs.
PSN significantly increased comfort to (median [25 to 75% interquartile range]) 8 [7 to 8] and 9 [8 to 9] with standard and new helmets, respectively, as opposed to both PSP (5 [5 to 6] and 7 [6 to 7]) and NAVA (6 [5 to 7] and 7 [6 to 8]; P < 0.01 for all comparisons). Regardless of the interface, PSN also decreased EAdipeak (P < 0.01), while increasing PTP of the first 300 ms from the onset of patient effort, indexed to the ideal PTP (P < 0.01) and PTP of the first 500 ms from the onset of patient effort, indexed to the ideal PTP (P < 0.001). ABGs were not different among trials.
When delivering noninvasive ventilation by helmet, compared to PSP and NAVA, PSN improves comfort and patient–ventilator interactions, while not ABGs. (Anesthesiology 2016; 125:1181-9)
Noninvasive ventilation by a facemask or a helmet is used after extubation to reduce the need for reintubation in patients at risk for extubation failure. Due to the large internal volume and the upward displacement (in its standard version), the helmet makes ventilator triggering less efficient and patient–ventilator asynchrony common.
Pressure support (PS) ventilation with a helmet was studied after extubation in 15 patients. Neurally adjusted PS increased comfort and improved patient–ventilator interactions, compared with standard (pneumatic) PS or neurally adjusted ventilatory assist.
BOTH ventilator performance and interface tolerance are important determinants of noninvasive ventilation (NIV) success.1–4 Neurally adjusted ventilator assist (NAVA) is a ventilatory mode where the ventilator is driven by the diaphragm electrical activity (EAdi), rather than by the conventional pneumatic signals, i.e., flow, volume, and airway pressure (Paw). Noninvasive NAVA improves patient–ventilator interaction and reduces asynchronies, compared to pneumatically triggered and cycled-off pressure support (PSP), the most common mode for NIV delivery.5–9 The helmet is a relatively novel interface for NIV that, compared to facemasks, improves patient tolerance to NIV and allows NIV administration for longer periods with fewer interruptions and NIV-related side effects4,10–12 and may reduce intubation rates and 90-day mortality.13 The helmet, however, is characterized by a less efficient rate of pressurization, poor triggering function, and a higher rate of asynchronies.14,15
In hypoxemic patients receiving NIV through a helmet for treatment of postextubation respiratory failure, compared to PSP, NAVA improves patient–ventilator interaction and synchrony, without determining significant variations in EAdi, respiratory rate, and arterial blood gases (ABGs).6 When applying NIV with a standard helmet (SH), secured to the patient by armpit braces, the delivery of pressure to the patient’s airway is altered by the upward displacement of the interface during insufflation.16 A modified new helmet (NH) has been shown to improve, as opposed to a SH, triggering and pressurization performance in PSP on bench,16 in healthy volunteers,17 and in critically ill patients.18
In the current study, we propose a specific NIV setting to generate a neurally controlled pressure support (PSN), consisting of increasing the user-controlled gain factor (NAVA level) at maximum (15 cm H2O/μV), with the upper pressure limit adjusted to achieve a maximum Paw of 20 cm H2O, including positive end-expiratory pressure (PEEP).19 Accordingly, as soon as EAdi rises up, exceeding the trigger threshold (0.5 μV), Paw rapidly increases resulting in a steep pressurization of up to 20 cm H2O.
We hypothesize that PSN would improve comfort and reduce EAdi in intensive care unit (ICU) patients undergoing NIV, by improving the rate of pressurization, compared to both PSP and NAVA, and triggering delays, as opposed to PSP only. Therefore, we compare the effects of PSN, PSP, and NAVA, delivered by both SH and NH, on comfort, respiratory drive, breathing pattern, ABGs, and patient–ventilator interaction and synchrony in patients receiving NIV for prevention of postextubation respiratory failure.
Materials and Methods
The study was conducted in the ICU of the University Hospital “Maggiore della Carità” (Novara, Italy) between July 2012 and January 2013, in accordance with the principles outlined in the Declaration of Helsinki. The institutional ethics committee approved the study (protocol number 484/CE, number of approval 64/12), and patient consent was obtained according to the Italian regulations. At the time the study was conducted, trial registration was not mandatory for this type of investigation. None of the patients in this study had been enrolled in other investigations. Part of the data was previously displayed in abstract form for poster presentations.
Subjects and Study Protocol
We considered eligible any patient aged 18 yr or older on invasive mechanical ventilation with all the following characteristics: (1) previous invasive mechanical ventilation greater than or equal to 48 h; (2) consciousness, as indicated by Glasgow Coma Scale of 11 (i.e., spontaneous eyes opening, obeys command, and no verbal response because of the endotracheal tube in place); (3) no infusion of midazolam and propofol in the previous 24 and 4 h, respectively; (4) readiness for extubation with indication, before extubation, to receive prophylactic NIV to prevent postextubation respiratory failure. Exclusion criteria were as follows: (1) agitation requiring sedation; (2) recent cervical spine injury; (3) obstructive sleep apnea syndrome; (4) pregnancy; (5) contraindications to placement of a nasal-gastric feeding tube; (6) inclusion in other research protocol; and (7) lack of consent.
Patients were considered at risk for extubation failure when meeting at least one of the following: (1) more than one consecutive failure of weaning trial20 ; (2) arterial carbon dioxide tension (Paco2) more than 45 mmHg after extubation; (3) chronic respiratory disorders20 ; and (4) chronic heart failure.20
After enrollment, a nasal-gastric feeding tube designed for EAdi detection (NAVA Catheter; Maquet Critical Care, Sweden) was placed, as previously described.21 All trials were conducted with a Servo-I ventilator (Maquet Critical Care) equipped with a software for air-leak compensation.
We used two separate computer-generated random sequences through sealed, opaque numbered envelops held by physicians and nurses not involved in the study. The first sequence was used for the order of application of the two helmets, while the second for the sequence of administration of the ventilatory modes. Therefore, each patient underwent three consecutive 30-min trials for both helmets with the same random sequence of modes. PEEP was always set at 10 cm H2O and left unmodified throughout the whole study period. The inspired oxygen fraction, set to obtain oxygen saturation measured by pulse oximetry more than 94% and less than 97% before starting the protocol, remained unmodified throughout the study period. The specific settings of the three modes were as follows: (1) PSP, inspiratory support of 10 cm H2O above PEEP, fastest rate of pressurization, expiratory trigger threshold at 50% of the peak inspiratory flow; (2) NAVA, NAVA level set to achieve a comparable peak EAdi (EAdipeak) as during PSP, with a safety Paw upper limit of 30 cm H2O6 ; and (3) PSN, NAVA level set at 15 cm H2O/μV19 with an upper Paw limit of 25 cm H2O, to obtain the same 20 cm H2O of the overall Paw applied in PSP.22 In fact, the maximum pressure applied by the Servo-I ventilator is 5 cm H2O below the preset upper Paw limit.22 Fixed by the manufacturer, the default cycling-off during both NAVA and PSN is 70% of EAdipeak.22
Predefined criteria for protocol interruption were as follows: (1) need for emergency reintubation; (2) oxygen saturation measured by pulse oximetry less than 90%; (3) acute respiratory acidosis, as defined by Paco2 more than 50 mmHg and pH less than 7.30; (3) inability to expectorate secretions; (4) hemodynamic instability (i.e., need for continuous infusion of dopamine or dobutamine greater than 5 µg·kg−1·min−1, norepinephrine greater than 0.1 µg·kg−1·min−1, or vasopressin to maintain mean arterial blood pressure greater than 60 mmHg); (5) life-threatening arrhythmias or electrocardiographic signs of ischemia; or (6) loss of two or more points on the Glasgow Coma Scale.
Data Acquisition and Analysis
Airflow, Paw, and EAdi were acquired from the ventilator through an RS232 interface at a sampling rate of 100 Hz and recorded on a personal computer by means of dedicated software (NAVA Tracker V. 3.0; Maquet Critical Care). The last minute of each trial was manually analyzed off-line using customized software based on Microsoft Excel (Microsoft Corporation, USA), as previously described.21
Mechanical inspiratory time and rate of ventilator cycling (mechanical respiratory rate [RRmec]) were determined from the flow tracing, while the patient’s neural TI (TIneu) and RR (RRneu) were obtained from the EAdi tracing.6 Mechanical and neural inspiratory duty cycles were computed.6 We measured EAdi from baseline to peak (EAdipeak) to assess the neural drive.23 The peak inspiratory and the mean Paw values were also measured.
The pressurization performance was evaluated with the Paw-time product (PTP) of the first 200 ms computed from the onset of ventilator assistance (PTP200) and of the first 300 and 500 ms from the onset of patient effort, indexed to the ideal PTP and expressed in percentage (PTP300-index and PTP500-index, respectively).16,18,24 The ideal PTP was computed considering a perfectly squared rectangle on the Paw-time tracing, having the height of the actual Paw above PEEP and the width of the time window considered (i.e., 0.3 and 0.5 s from the onset of the inspiratory effort, assessed from the EAdi tracing, for PTP300-index and PTP500-index, respectively).16,18,24 The drop in Paw (ΔPtrigger) and the PTP during the triggering phase (PTPt) were determined to evaluate triggering performance.16,18,24
We calculated the inspiratory (DelayTR-insp) and expiratory trigger delays6 and the time of synchrony between diaphragm activity and ventilator assistance, indexed to patient’s own (neural) TI.18,24 The asynchrony index (AI%) was computed as the sum of ineffective efforts, autotrigger, and double trigger, divided by the overall number of triggered and nontriggered breaths.25 We considered an AI% greater than or equal to 10% to indicate a clinically relevant rate of asynchronies.25
At the end of each trial, arterial blood was sampled for ABGs and comfort was assessed using the 11-point numeric rating scale (NRS), as previously reported.18,26,27 Briefly, patients were asked to indicate a number between 0 (worst possible comfort) and 10 (no discomfort at all) on an ICU-adapted, large printed scale including numbers and descriptors.27 Before protocol initiation, all patients received a detailed explanation about the 11-point NRS. The scores obtained were recorded without further indications or comments.18
Statistical Analysis
To calculate the sample size necessary to ascertain a 50% NRS increase with PSN, we have utilized the values from a database of patients previously evaluated with a SH in PSP, with a mean NRS of 5.0 and a SD of 2.7. To detect an increase in comfort of 2.5, with α risk of 0.05 and β risk of 0.20, a sample of 12 patients were deemed necessary. Because this calculation is based on a Student’s paired t test and we performed comparisons among three conditions, we have applied Bonferroni correction, reducing the α risk from 0.05 to 0.017, which increases the sample up to 15 patients.
Data are reported as median (25 to 75% interquartile range) unless otherwise specified. All continuous variables were compared between modes with both helmets. The effects of the three modes were assessed separately for each helmet by the Friedman test and then by the Wilcoxon rank text, with Bonferroni correction for multiple comparisons (P < 0.017). We compared categorical data by the McNemar test, while the Spearman rank test was used to determine the correlation between each individual comfort scores and the corresponding PTP300-index, PTP500-index, PTPt, and DelayTR-insp; for these comparisons, we considered two-sided P < 0.05 significant. Statistical analysis was performed using the Sigmaplot v. 12.0 (Systat Software Inc., USA).
Results
We enrolled 15 consecutive patients, after obtaining written informed consent. Patients’ characteristics at ICU admission are provided in table 1. The main reason for ICU admission was hypoxemic acute respiratory failure (ARF). All patients completed the study protocol without complications. Two patients were reintubated before 48 h after extubation: one because of severe dyspnea and the second due to respiratory acidosis
Comfort
Figure 1 shows individual scores and median values of comfort for all trials. With SH (left panel), PSN (8 [7 to 8]) improved the comfort score, when compared to both PSP (5 [5 to 6]; P < 0.001) and NAVA (6 [5 to 7]; P = 0.001). Similarly, with NH, PSN (9 [8 to 9]) outperformed both PSP (7 [6 to 7]; P < 0.001) and NAVA (7 [6 to 8]; P = 0.003). The comfort scores were directly correlated to PTP300-index (ρ = 0.52; P < 0.001) and PTP500-index (ρ = 0.50; P < 0.001) and inversely correlated to DelayTR-insp (ρ = −0.56; P < 0.001) and PTPt (ρ = −0.53; P < 0.001).
Hollow circles (pneumatically controlled pressure support [PSP]), squares (neurally adjusted ventilatory assist [NAVA]), and triangles (neurally controlled pressure support [PSN]) indicate the individual comfort scores given by the 15 patients on the numeric rating scale with standard helmet (left) and new helmet (right). Solid etched lines depict median values. #P < 0.017 PSNversus PSP; §P < 0.017 PSNversus NAVA.
Hollow circles (pneumatically controlled pressure support [PSP]), squares (neurally adjusted ventilatory assist [NAVA]), and triangles (neurally controlled pressure support [PSN]) indicate the individual comfort scores given by the 15 patients on the numeric rating scale with standard helmet (left) and new helmet (right). Solid etched lines depict median values. #P < 0.017 PSNversus PSP; §P < 0.017 PSNversus NAVA.
Breathing Pattern, Respiratory Drive, and ABGs
As depicted in table 2, RRneu, TIneu, and neural inspiratory duty cycle were unaffected by the ventilatory mode with both helmets. Also shown in table 2, PSN significantly decreased RRmec, as compared to PSP while not to NAVA, when delivering NIV by NH. Conversely, RRmec was no different at all between modes when delivering NIV by SH. Irrespective of the interface, mechanical inspiratory time and mechanical inspiratory duty cycle were greater with NAVA and PSN than with PSP. Figure 2 displays Paw, flow, and EAdi tracings of a representative patient during PSP, NAVA, and PSN delivered by SH. Compared to both PSP and NAVA, PSN reduced EAdipeak. Group median values confirm that EAdipeak is reduced by PSN, as opposed to both PSP and NAVA (table 2). ABGs were not different between trials (table 2).
Respiratory Drive and Timing, Arterial Blood Gases, Pressurization Rate, and Triggering Performance

Examples of tracings from a representative patient breathing during pneumatically controlled pressure support (PSP), neurally adjusted ventilatory assist (NAVA), and neurally controlled pressure support (PSN), delivered through a standard helmet. Airway pressure (Paw), flow, and electrical activity of the diaphragm (EAdi) are displayed from top to bottom. PSN reduces EAdi as opposed to both PSP and NAVA.
Examples of tracings from a representative patient breathing during pneumatically controlled pressure support (PSP), neurally adjusted ventilatory assist (NAVA), and neurally controlled pressure support (PSN), delivered through a standard helmet. Airway pressure (Paw), flow, and electrical activity of the diaphragm (EAdi) are displayed from top to bottom. PSN reduces EAdi as opposed to both PSP and NAVA.
Triggering Performance and Rates of Pressurization
Figure 3 displays Paw profiles of single breaths during PSP (solid line), NAVA (dotted line), and PSN (dashed line), with both interfaces from one patient. The arrow indicates the initiation of diaphragm effort, as assessed by EAdi. Irrespective of the interface, both PSN and NAVA show a shorter DelayTR-insp, as opposed to PSP, while the pressurization rate looks steeper in PSN than in PSP and NAVA. In keeping with these observations, group median data, as displayed in table 2, show that compared to PSP, both PSN and NAVA improved triggering performance irrespective of the interface, while all the indices of pressurization were significantly improved by PSN, as opposed to both PSP and NAVA, with no significant difference between PSP and NAVA. Also shown in table 2, compared to PSP, NAVA improved PTP300-index and PTP500-index with SH, while not with NH.
Airway pressure (Paw) profiles of single breaths during pneumatically controlled pressure support (solid line), neurally adjusted ventilatory assist (dotted line), and neurally controlled pressure support (dashed line) with both standard helmet (left) and new helmet (right), from another patient. In both panels, the arrow indicates the commencement of diaphragm effort, as assessed by diaphragm electrical activity.
Airway pressure (Paw) profiles of single breaths during pneumatically controlled pressure support (solid line), neurally adjusted ventilatory assist (dotted line), and neurally controlled pressure support (dashed line) with both standard helmet (left) and new helmet (right), from another patient. In both panels, the arrow indicates the commencement of diaphragm effort, as assessed by diaphragm electrical activity.
Patient–Ventilator Synchrony
As shown in table 2, irrespective of the interface, PSP and NAVA significantly improved the ratio between time during which respiratory effort and ventilator assistance were synchronous and the TIneu, as opposed to PSP. No patient had an AI% greater than or equal to 10% with PSN and NAVA, while in PSP, the patients with AI% greater than or equal to 10% were 8 with SH (P = 0.02, compared to both PSN and NAVA) and 5 with NH (P = 0.04, compared to PSN and NAVA).
Discussion
We found in patients receiving NIV by helmet that, irrespective of the interface, compared to both PSP and NAVA, PSN improves patient comfort, reduces EAdi, and results in better pressurization and triggering performance, while it does not affect ABGs and respiratory rate. Both PSN and NAVA improve patient–ventilator synchrony, as opposed to PSP, with no significant difference between the two modes.
Although several studies proved that, compared to PSP, NAVA ameliorates patient–ventilator interactions and synchrony during NIV,5–8 none demonstrated a reduction of EAdi, which is the best estimate of the respiratory drive28 and reflects diaphragm effort.29 In fact, in one study, EAdi is slightly, although significantly, higher in NAVA than in PSP,8 while in the others, EAdi is no different between the two modes.5–7 In keeping with these findings, we find similar EAdipeak in NAVA and PSP, while PSN significantly reduces EAdipeak.
With PSN, a towering NAVA level and a safety Paw limit are set on the ventilator. Accordingly, as soon as EAdi rises up exceeding the trigger threshold (0.5 μV), the ventilator immediately applies a boost of pressure to the helmet determining a steep pressurization up to 20 cm H2O. Briefly, the upper pressure limit is achieved when the rise in EAdi reaches 1.5 μV; in fact, 1.5 μV (EAdi) × 15 H2O/μV (NAVA level) > 20 cm H2O. Consequently, as depicted in figure 3, the shape of the airway pressure in PSN is nearly squared, as in an ideal PSP, but EAdi determines on and off cycling instead of flow.
Patient comfort is a major determinant of NIV success30 and depends on both interface tolerance and ventilator performance.31 NIV by helmet reduces the inspiratory effort, compared to spontaneous breathing, although to a lesser extent than by facemask.14,15 Vargas et al.15 showed that a specific ventilator setting for helmet NIV, consisting of a 50% increase in both the expiratory and inspiratory pressures, lessens diaphragm effort down to the values observed during mask NIV; these specific settings, however, did not improve patient comfort. We recently demonstrated NH to be superior to SH in terms of patient comfort and triggering and pressurization performance; the median value of EAdipeak, however, was lower with NH than with SH, without achieving statistical significance.18 Notably, in the current study, EAdi is significantly reduced by PSN with both helmets, compared to both PSP and NAVA.
Moerer et al.32 showed in healthy individuals that, compared to conventional pneumatic triggering during PSP, EAdi triggering significantly improves comfort. In our study, both PSN and NAVA settings improved trigger delays and PTPt, compared to PSP, with no significant difference between NAVA and PSN. While Moerer et al.32 in healthy volunteers set PEEP at 5 cm H2O, we applied PEEP at 10 cm H2O because the main reason for ICU admission was hypoxemic ARF; indeed, as shown in table 2, the median arterial oxygen tension/inspired oxygen fraction ratio never exceeded 250 mmHg throughout the study protocol. In patients undergoing NIV for prevention of postextubation respiratory failure, Vargas et al.15 found that increasing PEEP from 5 to 8 cm H2O reduced the work of breathing and improved patient–ventilator interaction overall, without affecting patient’s comfort.
In hypoxemic patients undergoing invasive PSP, Chiumello et al.33 found that both the lowest and highest pressurization rates were associated with the worst level of comfort. In patients with chronic obstructive pulmonary disease (COPD) who are receiving PSP through a facemask while recovering from an episode of acute hypercapnic respiratory failure, Prinianakis et al.34 also observed that the highest speed of pressurization worsened patient comfort. Different from the study performed in intubated patients,33 Prinianakis et al.34 found the inspiratory muscle effort to be inversely related to the rate of pressurization.34 Importantly, in the study by Prinianakis et al.,34 the highest speed of pressurization caused air leaks that are well-known determinants of patient’s intolerance to NIV. Worth mentioning, both the endotracheal tube and facemask are characterized by very low compliance, which makes ventilator-delivered pressure entirely transmitted to the airway. Quite the opposite, the helmet is inefficient in pressurizing the airway partly because of the soft compliant wall and the increased internal compressible volume14 and primarily consequent to the downward displacement of the soft collar during ventilator insufflation in its standard version.16–18
Our study has some limitations deserving discussion. The number of patients enrolled is relatively small, as in the majority of the physiologic investigations.3,5–8,14,15,18,21,33 We powered our study to detect a 50% improvement in comfort, as assessed by the NRS, with PSN, as opposed to PSP, for at least one of the two interfaces. PSN significantly improves comfort with both interfaces, by 60% and 29% with SH and NH, respectively. Of note, PSN improves comfort with respect to both PSP and NAVA, with small and nonsignificant differences between the latter two modes.
We studied a mixed patient population with mainly hypoxemic ARF, including individuals with diverse underlying diseases, having in common only the indication for prophylactic NIV to prevent extubation failure and reintubation.20 We chose this patient population because, for proper comfort assessment, we consider including awake, nonsedated, and cooperative patients important, which may be problematic in more acutely and severely ill patients.35 In keeping with previous studies, we assess patient comfort by NRS18,26,27 ; this scale, however, is just formally validated for the assessment of pain36–38 and dyspnea.39
We use the 10% threshold to indicate a clinically relevant rate of asynchronies, which may not be correct for NIV. Very recently, Doorduin et al.9 found, in a selected population of COPD patients, that ineffective efforts increase drastically after timing errors between EAdi and airway pressure, reaching 20%. As we do not study COPD patients and do not take into account timing errors in the computation of the AI%, while considering only “major” asynchronies (ineffective efforts, autotriggering, and double triggering), we deem the 10% threshold appropriate, in keeping with several previous studies dealing with asynchronies in NIV.5–8,18,25
Finally, because of their specific characteristics, the helmets are the interfaces with the highest potential for improvement with PSN, compared to both PSP and NAVA. In fact, the improvements observed are generally less prominent with the more performing NH than with the SH.16–18 It is uncertain whether the physiologic benefits we observe may occur when applying NIV by mask.
Conclusions
In patients receiving NIV by helmet for preventing extubation failure and reintubation, compared to both PSP and NAVA, PSN ameliorates patient comfort, while reducing the neural drive and effort and enhancing the pressurization rates. In addition, as opposed to PSP, while not to NAVA, PSN improves triggering performance and patient–ventilator synchrony. Whether these benefits may take place with other interfaces or when applying NIV to other categories of patients remains unclear and needs further evaluation.
Acknowledgments
The authors thank all physicians and nurses of the Intensive Care Unit of “Maggiore della Carità” Hospital of Novara (Novara, Italy) for their helpful and continuous support. The authors are indebted to Intersurgical S.p.A. (Mirandola, Italy) that provided the helmets used for the study and Maquet Critical Care (Solna, Sweden) that provided the neurally adjusted ventilatory assist module and catheters used for the study.
Research Support
Support was provided solely from institutional and/or departmental sources. Dr. Navalesi’s research laboratory has received equipment and/or grants from Maquet Critical Care (Solna, Sweden), Intersurgical S.p.A. (Mirandola, Italy), Draeger Medical GmbH (Corsico, Italy), Biotest (Trezzano sul Naviglio, Italy), and Hillrom (Bussigny, Switzerland).
Competing Interests
Dr. Navalesi contributed to the development of the helmet Next (Castar Next; Intersurgical S.p.A., Mirandola, Italy), whose license for patent belongs to Intersurgical S.p.A., and received royalties for that invention. Dr. Navalesi received honoraria/speaking fees from Maquet Critical Care (Solna, Sweden), Covidien AG (Segrate, Italy), Draeger Medical GmbH (Corsico, Italy), Breas (Saint-Priest, France), Hillrom (Bussigny, Switzerland), and Linde AG (Munich, Germany). The other authors declare no competing interests.