Respiratory viruses are transmitted via respiratory particles that are emitted when people breath, speak, cough, or sneeze. These particles span the size spectrum from visible droplets to airborne particles of hundreds of nanometers. Barrier face coverings (“cloth masks”) and surgical masks are loose-fitting and provide limited protection from airborne particles since air passes around the edges of the mask as well as through the filtering material. Respirators, which fit tightly to the face, provide more effective respiratory protection. Although healthcare workers have relied primarily on disposable filtering facepiece respirators (such as N95) during the COVID-19 pandemic, reusable elastomeric respirators have significant potential advantages for the COVID-19 and future respiratory virus pandemics. However, currently available elastomeric respirators were not designed primarily for healthcare or pandemic use and require further development to improve their suitability for this application. The authors believe that the development, implementation, and stockpiling of improved elastomeric respirators should be an international public health priority.

Elastomeric respirators are reusable respirators made of plastic or non–latex-containing rubber materials with exchangeable filter cartridges or canisters. Elastomeric respirators have been widely used in industry, but not in health care. We believe there is an urgent need to develop elastomeric respirators specifically for respiratory protection from pandemic respiratory viruses. While such elastomeric respirator development may not take place in time for widespread use during the current COVID-19 pandemic, it is vitally important to be prepared for the next respiratory pandemic, which could occur at any time.1  There is also the possibility that immune escape of SARS-CoV-2 variants could result in a protracted COVID-19 pandemic, heightening the importance of effective respiratory protection.2  In this article, we explain why respiratory protection is important, describe the major types of respiratory protection, and elaborate on the potential role for elastomeric respirators.

Respiratory viruses may spread by fomite (touching a contaminated surface and then transferring the virus to the respiratory tract), droplet, or airborne routes (table 1). Determining the relative importance of these routes for an individual virus is difficult and may be controversial; for SARS-CoV-2, the fomite route of transmission is thought to be less important.3,4  Many experts now believe that there is a continuous spectrum of different sized respiratory particles produced by persons infected with respiratory viruses, and that the traditional distinction between droplet and airborne transmission is arbitrary and obsolete.5–7  During the current COVID-19 pandemic, there has been controversy, as in previous respiratory virus pandemics, concerning whether respiratory protection should protect against only droplet or droplet and airborne particles.8  However, there is now substantial evidence that airborne transmission of SARS-CoV-2 transmission is likely5,9–12 ; inhalation of respiratory particles may be the predominant route of transmission. Difficulty in culturing putative airborne pathogens from air samples because of various technical obstacles does not rule out airborne transmission.5,12  Therefore, it seems prudent to protect healthcare workers (and the public) accordingly, by using respiratory protection whenever possible. The two major categories of respirators used for respiratory protection in health care include negative and positive pressure mode respirators.

Table 1.

Respiratory Protection Terminology

Respiratory Protection Terminology
Respiratory Protection Terminology

The wearer of a negative pressure mode respirator draws air through a filter and into a facepiece during inhalation. There are two main types of negative pressure respirators: disposable filtering facepiece respirators13  (such as the N95 filtering facepiece respirators, familiar to many healthcare workers), and elastomeric respirators,14  which are reusable (table 1; fig. 1). The effectiveness of respiratory protection depends upon fit, filtering efficiency, and resistance to breathing (table 1).

Fig. 1.

Examples of filtering facepiece respirators, elastomeric respirators, and a hybrid respirator. Elastomeric respirators with P100 filters enclosed in plastic cartridges are shown in A and C. Both elastomeric respirators have exhalation valves at the bottom of the front of the mask. The respirator in A also has a “speaking diaphragm” in the center front of the respirator. A closeup view of the speaking diaphragm is shown in B. The filtering facepiece respirator in D is a typical N95. The filtering facepiece respirator in E is a unique strapless N95 that is attached to the face by an adhesive strip around the edge of the mask. An example of a hybrid respirator design is shown in F, combining a plastic mask body and sealing surface with a melt-blown fabric filtering area. The example shown is a “quarter face” respirator that does not cover the chin; the other respirators shown are “half face” respirators; quarter face respirators have an Assigned Protection Factor of 5, while half face respirators have an Assigned Protection Factor of 10 (see table 1).

Fig. 1.

Examples of filtering facepiece respirators, elastomeric respirators, and a hybrid respirator. Elastomeric respirators with P100 filters enclosed in plastic cartridges are shown in A and C. Both elastomeric respirators have exhalation valves at the bottom of the front of the mask. The respirator in A also has a “speaking diaphragm” in the center front of the respirator. A closeup view of the speaking diaphragm is shown in B. The filtering facepiece respirator in D is a typical N95. The filtering facepiece respirator in E is a unique strapless N95 that is attached to the face by an adhesive strip around the edge of the mask. An example of a hybrid respirator design is shown in F, combining a plastic mask body and sealing surface with a melt-blown fabric filtering area. The example shown is a “quarter face” respirator that does not cover the chin; the other respirators shown are “half face” respirators; quarter face respirators have an Assigned Protection Factor of 5, while half face respirators have an Assigned Protection Factor of 10 (see table 1).

Close modal

Fit defines the extent to which there is an occlusive seal of the respirator with the face, preventing leakage and forcing all inhaled and exhaled air to pass through the filtering material (or in some cases through an exhalation valve). In contrast, barrier face coverings15  (“cloth masks”) and surgical masks16  are not designed to make a tight seal to the face (table 1). Only when there is an occlusive seal with the face, as with a respirator, does almost all of the air entering the facepiece pass through the filtering material, providing the optimal opportunity to remove particles from the air.17,18  In the United States, the Occupational Safety and Health Administration (Washington, D.C.) Respiratory Protection Standard19  requires that employees must be medically cleared to wear a respirator and that employers must measure the fit of respirators for employees upon first selection and annually thereafter using standardized tests that may be qualitative or quantitative.19  Numerous studies have shown that the fit of a filtering facepiece respirator varies considerably depending upon the particular design and materials, and the facial characteristics of an individual user; not all respirators fit the full range of human faces.20  The face must be clean-shaven in order to obtain an effective fit. A recent American Society for Testing and Materials (West Conshohocken, Pennsylvania) standard for “respirator fit capability” may be used to characterize how well a particular respirator will fit a variety of facial characteristics.21  Moreover, filtering facepiece respirators are disposable, and the quality of the fit may deteriorate with repeated use.22 

Filtering facepiece respirators typically utilize a layer of melt-blown polypropylene (or other synthetic material) fabric as the filter. The randomly oriented fibers of the melt-blown fabric filter particles by a combination of impaction, diffusion, and interception.23  In addition, an electrostatic charge is often added to enhance filtration, causing the material to function as a permanent magnet, or “electret.”24 

Elastomeric respirators are constructed of various solid plastic or rubber materials that form the mask body and the portion of the mask that seals to the face. Filtering is accomplished by one or more exchangeable filters, constructed of material similar to that of filtering facepiece respirators, that attach to the mask body. The filtering material may be protected within a canister or cartridge (figs. 1 and 2). In the United States, the National Institute for Occupational Safety and Health (Washington, D.C.), a branch of the Centers for Disease Control and Prevention (Atlanta, Georgia), certifies the filtering efficiency of respirators using standardized tests25  (table 1).

Fig. 2.

Examples of elastomeric respirator filters. The outer facing side of the filters is shown in the upper photos, and the side of the filters that attaches to the respirator is shown in the lower photos. Both filters are P100 (equivalent to high-efficiency particulate air). The filtering material of the left filter is exposed, while the filtering material of the right filter is completely enclosed in plastic. The right filters are also shown mounted on an elastomeric respirator in fig. 1C.

Fig. 2.

Examples of elastomeric respirator filters. The outer facing side of the filters is shown in the upper photos, and the side of the filters that attaches to the respirator is shown in the lower photos. Both filters are P100 (equivalent to high-efficiency particulate air). The filtering material of the left filter is exposed, while the filtering material of the right filter is completely enclosed in plastic. The right filters are also shown mounted on an elastomeric respirator in fig. 1C.

Close modal

In addition to fit and filtration, resistance to breathing is an important property. Increased resistance to breathing results in increased negative and positive pressures inside the respirator during inhalation and exhalation, respectively, which could increase leakage around the respirator facepiece. Resistance to breathing also affects comfort and tolerability.26  In the United States, the National Institute for Occupational Safety and Health requires that respirator filtering material be tested for the pressure gradient across the filtering material at 85 l/min constant airflow; the “inspiratory” pressure gradient must be less than 35 mm H2O (343 Pa), and “expiratory” pressure gradient must be less than 25 mm H2O (245 Pa). Methods for testing the pressure gradient across filtering material typically specify the cross-sectional area of the material to be tested, since the cross-sectional area affects the pressure gradient under conditions of constant volumetric flow rate. The National Institute for Occupational Safety and Health method for certifying filtering facepiece respirators specifies that the entire respirator is sealed onto a plate for testing; approved filtering facepiece respirators typically have inspiratory and expiratory pressure gradients in the approximate range of 5 to 15 mm H2O at 85 l/min airflow.27  Currently, there is not a standardized test for measuring the pressure gradient across a respirator during actual use. A study of a single model of a filtering facepiece respirator found peak pressure gradients of less than 20 mm H2O during exercise.28 

Powered air purifying respirators employ a fan or pump that brings air through the filter and into the facepiece. Filtration is important, but fit is less so; a powered air purifying respirator with a loose-fitting hood offers greater protection to the wearer than an N95 filtering facepiece respirator because it provides more than enough filtered air to overcome any inward airflow between the hood and the wearer’s head.29  Powered air purifying respirators with a loose-fitting hood generally have an Assigned Protection Factor (table 1) of 25, or 2.5 times the Assigned Protection Factor of 10 for a filtering facepiece or elastomeric respirator.19  Loose-fitting powered air purifying respirators have the advantage of not requiring fit testing, and contrary to negative pressure mode respirators, do not increase resistance to breathing. Their performance is not affected by the presence of facial hair as they do not make a tight seal to the face. It is important to understand that the exhaled air passes around the edges of a loose-fitting powered air purifying respirator hood without filtration; only the inhaled air is filtered.

Disadvantages of powered air purifying respirators include their greater complexity of use (i.e., fan motors and batteries), higher cost, and the possibility of impaired communication caused by fan noise. A potentially serious shortcoming of loose-fitting powered air purifying respirators is that the exhaled air is exhausted into the ambient air without filtration. This could hypothetically result in contamination of a sterile field or transmission of a respiratory pathogen by the wearer.30  The unfiltered exhaust from a powered air purifying respirator is analogous to an unfiltered exhalation valve on a negative pressure mode respirator. We suggest that powered air purifying respirators should be available for healthcare workers who, because of facial features or facial hair, are unable to be fitted for a filtering facepiece or elastomeric respirator, or when an increased Assigned Protection Factor is desired. However, we also suggest that during a pandemic, negative pressure mode respirators (i.e., filtering facepiece respirators or elastomeric respirators) are likely to be a better solution for providing respiratory protection for large numbers of healthcare workers.

Tight-fitting half facepiece powered air purifying respirators are also available, although they have been used much more widely in industry than in health care (fig. 3). Tight-fitting powered air purifying respirators require fit testing and would not be suitable for use with facial hair that interferes with the seal of the mask body to the face. An Assigned Protection Factor of 50 is possible, five times that of a negative pressure elastomeric respirator and twice that of a typical loose-fitting powered air purifying respirator,19  making this an attractive option when higher levels of protection are desired.

Fig. 3.

An example of a tight-fitting, half facepiece powered air purifying respirator is shown in A (CleanSpace Technology; Australia); the unfiltered exhalation valve is in the center of the facepiece. In B, another view of the battery and fan unit is shown. In C, a view of the accessory N95 filter attachment for the exhalation valve is shown; when used, filtration of the exhalation valve provides “source control,” preventing the wearer of the respirator from exhaling respiratory pathogens into the surrounding air. This respirator is intended to make a highly effective seal against the face, and should be fit tested before use, just as with a negative pressure elastomeric respirator or filtering facepiece respirator. Air is drawn through a high-efficiency particulate air filter (shown in D) and pushed into the respirator under positive pressure by the battery-powered fan located behind the wearer’s head. Exhalation occurs through an unfiltered valve on the front of the respirator, unless the accessory N95 filter is attached to the exhalation valve (the replaceable N95 filter is shown inside of the filter holder in E). This respirator has at least three advantages compared to a loose-fitting powered air purifying respirator. First, because this respirator has both a tight seal against the face and positive pressure produced by the battery powered fan, the Assigned Protection Factor is 50, twice that of a typical loose-fitting powered air purifying respirator, or five times that of a negative pressure elastomeric respirator. Second, it is possible to filter the exhaled air, which is not possible with a loose-fitting powered air purifying respirator. Third, if the power source or fan fails (for example, if the battery is exhausted), the respirator will function as an ordinary negative pressure elastomeric respirator, with the wearer pulling inhaled air through the high-efficiency particulate air filter; if a loose-fitting powered air purifying respirator power source or fan fails, there is no respiratory protection.

Fig. 3.

An example of a tight-fitting, half facepiece powered air purifying respirator is shown in A (CleanSpace Technology; Australia); the unfiltered exhalation valve is in the center of the facepiece. In B, another view of the battery and fan unit is shown. In C, a view of the accessory N95 filter attachment for the exhalation valve is shown; when used, filtration of the exhalation valve provides “source control,” preventing the wearer of the respirator from exhaling respiratory pathogens into the surrounding air. This respirator is intended to make a highly effective seal against the face, and should be fit tested before use, just as with a negative pressure elastomeric respirator or filtering facepiece respirator. Air is drawn through a high-efficiency particulate air filter (shown in D) and pushed into the respirator under positive pressure by the battery-powered fan located behind the wearer’s head. Exhalation occurs through an unfiltered valve on the front of the respirator, unless the accessory N95 filter is attached to the exhalation valve (the replaceable N95 filter is shown inside of the filter holder in E). This respirator has at least three advantages compared to a loose-fitting powered air purifying respirator. First, because this respirator has both a tight seal against the face and positive pressure produced by the battery powered fan, the Assigned Protection Factor is 50, twice that of a typical loose-fitting powered air purifying respirator, or five times that of a negative pressure elastomeric respirator. Second, it is possible to filter the exhaled air, which is not possible with a loose-fitting powered air purifying respirator. Third, if the power source or fan fails (for example, if the battery is exhausted), the respirator will function as an ordinary negative pressure elastomeric respirator, with the wearer pulling inhaled air through the high-efficiency particulate air filter; if a loose-fitting powered air purifying respirator power source or fan fails, there is no respiratory protection.

Close modal

Full face elastomeric respirators (sometimes referred to as “gas masks”) that cover the eyes, nose, and mouth within a single mask body are another option to consider. Such respirators are used routinely by the military, police, and fire fighters and for industrial applications in which infectious agents, chemical toxins, or radioactive particles can irritate or damage the eyes or gain entry to the body through the eyes, as well as the respiratory system. While eye protection in the form of glasses or goggles has been advised for protection from SARS-CoV-2,31  the evidence for respiratory viruses causing infection from contact with the conjunctiva is not strong.32–37  In general, full face masks are heavier, are less comfortable in warm environments, and may have greater interference with the field of vision in comparison to half masks (which do not cover the eyes). Nevertheless, a study of acceptability to healthcare workers of military-style elastomeric full face respirators in comparison to filtering facepiece respirators (used with eyeglasses or goggles) found a high level of acceptance for the full face respirators when worn for brief periods (up to 40 min).38 

It is important to note that while surgical masks are typically constructed from melt-blown fabric, and bear some superficial physical resemblance to filtering facepiece respirators, surgical masks are not respirators. Some surgical masks are cleared for marketing (in the United States) by the Food and Drug Administration (Silver Spring, Maryland) and meet American Society for Testing and Materials F2100 standards16  (table 1). Surgical masks are intended to block large droplets, splashes, sprays, or splatter from reaching the mouth or nose and to prevent the user’s droplets from exposing others. Surgical masks may capture some but not all of the particles contained in a wearer’s cough or sneeze.39  Since surgical masks are not designed to make an occlusive seal with the face, the protection from inhalation or exhalation of infectious particles will be limited in comparison to a properly fitted respirator.40–46  The American Conference of Governmental Industrial Hygienists (Cincinnati, Ohio) has published an informative graphic illustrating the comparative effectiveness of cloth face coverings, surgical masks, and respirators.47 

Filtering facepiece respirators such as N95 or the European Union equivalent FFP2 have been avidly sought during the COVID-19 pandemic; public health authorities initially advised that these respirators should only be used by healthcare workers, but recently some governments have recommended or even required their use by the public.48  While filtering facepiece respirators provide effective respirator protection when properly fitted, they have significant shortcomings (table 2). A major shortcoming is the overwhelming number of respirators required during a respiratory pandemic, inevitably resulting in acute shortages, even when respirators are stockpiled in advance. During the COVID-19 pandemic, due to the shortage of respirators, there were numerous attempts to construct homemade respirators using three-dimensional printed materials, respiratory therapy equipment, anesthesia circuit filters, and other available components.49–51  This situation revealed the critical shortage of commercially available respiratory protection. To the contrary, elastomeric respirators can be used for prolonged periods of time, and thus provide a durable solution during a pandemic.

Table 2.

Elastomeric Respirator Compared to Filtering Facepiece Respirator

Elastomeric Respirator Compared to Filtering Facepiece Respirator
Elastomeric Respirator Compared to Filtering Facepiece Respirator

Elastomeric respirators are widely used outside of health care for protection from aerosols and harmful vapors and gases. Their use in health care has been limited, but successful implementation has been reported.52–54  In 2019, before the COVID-19 pandemic, a consensus report from the National Academy of Sciences, Engineering, and Medicine (Washington, D.C.)concluded that “reusable elastomeric respirators could be a viable option for use in surge situations” and that surge use would be enhanced if “reusable elastomeric respirators were a part of healthcare facilities’ day-to-day respiratory protection program.”55  However, the report also acknowledged the shortcomings of existing elastomeric respirator products, and recommended “design of innovative reusable respirators and the implementation of robust respiratory protection programs…taking into account the distinctive characteristics of the healthcare workplace…” In light of the experience gained with respirators during the COVID-19 pandemic, this document now seems prescient.

There are several potential advantages of elastomeric respirators in comparison to filtering facepiece respirators (table 2).56  Elastomeric respirators are ideally suited to pandemic surge use because they can be reused indefinitely, and filter cartridges or canisters can last for extended periods of time (see the Durability of Elastomeric Respirators section). Elastomeric respirators are capable of robust user seal checks because occlusion of the filter ports completely prevents breathing if the mask makes an effective seal to the face. In addition, elastomeric respirators tend to have better fit57  than filtering facepiece respirators.

As already mentioned, the advantage of elastomeric respirators is that a set of filters can last for an extended period of time58 ; however, the precise longevity is uncertain because the concentration and nature of particles to be filtered will have a significant effect on the lifetime of the filter. High-efficiency filters such as 95 (FFP2), 99 (FFP3), 100 series (N, R, or P), or high-efficiency particulate air filters can be used until they load with particles to the point that the resistance to breathing becomes excessive. Particles that are filtered become tightly bound to the filtering material by Van der Waals forces and do not come out of the filter in significant quantitites.59  When filters are used in relatively dust-free environments such as hospitals, they would be expected to be highly durable, possibly allowing users to work through a pandemic with one or two sets of filters. Filters can be encased in plastic enclosures that can be handled without touching the filter material itself, allowing for external cleaning by, for example, disinfecting wipes (figs. 1 and 2). An ideal elastomeric respirator for a pandemic should be designed with filter ports able to accommodate filters from multiple manufacturers (to prevent supply chain disruptions). Since there is not a universal filter connection design, the use of filters from multiple sources would require the use of adapters specifically made for this purpose.38  Sources of filters during a surge might include anesthesia and ventilator breathing circuit filters.60  In the United States, the National Institute for Occupational Safety and Health would need to be involved in the design and certification of novel elastomeric respirators designed for use with a variety of filters.

Existing elastomeric respirators should undergo further development to improve their functionality. The properties of ideal elastomeric respirators are shown in table 3, and an artist’s conception of a future respirator is shown in fig. 4 (filters removed to show the respirator facepiece more clearly). Elastomeric respirators can make verbal communication difficult, in most cases more so than filtering facepiece respirators.61,62  A few elastomeric respirators are equipped with a “speaking diaphragm” (fig. 1), which transmits sounds by a thin disc located in a circumscribed area of the mask body. For elastomeric respirators to succeed in health care, sound transmission must be improved substantially by the use of “speaking diaphragms” or other measures. Transparent mask materials may also assist with effective communication by making facial expressions and lip movement discernable.63  Management of moisture is also a potential challenge since condensation of exhaled water vapor inside the mask can result in accumulation of moisture.

Table 3.

Properties of an Ideal Pandemic Elastomeric Respirator

Properties of an Ideal Pandemic Elastomeric Respirator
Properties of an Ideal Pandemic Elastomeric Respirator
Fig. 4.

Artist’s conception of improved elastomeric respirator, with filter ports on either side of the respirator facepiece; the filters are deliberately not shown. The respirator body is transparent, and there is a speaking diaphragm to facilitate communication. The respirator would accommodate a wide variety of filters through the use of adapters, which could make the respirator more suitable to surge situations in which the supply chain for filters could be interrupted. The sealing surface is highlighted. There is no exhalation port; inhalation and exhalation take place through the same filters. Straps that hold the respirator onto the face are deliberately not shown, but are an important feature for fit and comfort. In the United States, respirators are certified for use by the National Institute for Occupational Safety and Health (Washington, D.C.); thus, the National Institute for Occupational Safety and Health would need to be involved in improved elastomeric respirator design.

Fig. 4.

Artist’s conception of improved elastomeric respirator, with filter ports on either side of the respirator facepiece; the filters are deliberately not shown. The respirator body is transparent, and there is a speaking diaphragm to facilitate communication. The respirator would accommodate a wide variety of filters through the use of adapters, which could make the respirator more suitable to surge situations in which the supply chain for filters could be interrupted. The sealing surface is highlighted. There is no exhalation port; inhalation and exhalation take place through the same filters. Straps that hold the respirator onto the face are deliberately not shown, but are an important feature for fit and comfort. In the United States, respirators are certified for use by the National Institute for Occupational Safety and Health (Washington, D.C.); thus, the National Institute for Occupational Safety and Health would need to be involved in improved elastomeric respirator design.

Close modal

All but a few available elastomeric respirators have an unfiltered, valved exhalation port, which is undesirable in healthcare settings. Unfiltered exhaled air may contaminate a sterile field during a procedure, or result in transmission of a respiratory pathogen if the person wearing the respirator is infected. While these exhalation ports can be overcome by user modifications, such as by covering the exhalation port and removing the inhalation port valves, a respirator designed either without an exhalation port or with an exhalation port that can be switched off would be preferable. In some nonpandemic situations, unfiltered exhaled air may be acceptable—for example, if the person wearing the respirator is extremely unlikely to be infected with a specific transmissible respiratory pathogen (e.g., Mycobacterium tuberculosis in low prevalence areas). Some users have worn surgical masks over an elastomeric respirator with the idea of providing filtration for the exhaled air; however, this solution is unreliable due to the shortcomings of surgical masks described in the Surgical Masks section.

One of the potential challenges for elastomeric respirators is cleaning. Reusable respirators require periodic cleaning and decontamination. Elastomeric respirators have typically been designed for low-level disinfection such as cleaning with soap and water, disinfectant wipes, dishwasher disinfection at low temperature, or immersion in relatively weak chemicals such as diluted bleach.64  While the SARS-CoV-2 virus is inactivated by low-level disinfection, many other pathogens are resistant to this method. The elastomeric respirator is in close proximity to the user’s mouth, nose, and airway, has contact with respiratory secretions, and has potential contact with a wide variety of pathogens commonly found in the healthcare environment, including multidrug-resistant organisms. Therefore, an argument could be made that elastomeric respirators are semicritical devices that should be subject to high-level disinfection.65  Semicritical devices are those that come into contact with intact mucous membranes. High-level disinfection destroys all microorganisms except some bacterial spores. High-level disinfection could be particularly important if elastomeric respirators are processed by a hospital central processing unit and reissued to multiple users. High-level disinfection would require that elastomeric respirators be submersible (with the filter cartridges removed), and that the materials be resistant to heat or chemicals required for high-level disinfection. During surge conditions when other disinfection processes may be in short supply, heat-resistant respirator materials could be disinfected with boiling water.66  The hard surfaces of reusable filter cartridges or canisters (fig. 2) could be cleaned with disinfectant wipes (low-level disinfection) between uses.

There are several examples of respirators that combine a plastic or rubber mask frame and sealing surface with melt-blown fabric filtering, rather than filter cartridges or canisters. Thus, these respirators have features of both traditional filtering facepiece respirators and elastomeric respirators. An example of such a hybrid respirator is shown in fig. 1.

The environmental impact of elastomeric respirators in comparison to disposable filtering facepiece respirators has not been well-studied. The environmental impact of cleaning reusable respirators should be considered. Respiratory protection must be effective and available for all people who need it. The experience of the COVID-19 pandemic has clearly shown that control of viral transmission must occur everywhere in the world in order to prevent the selection of more fit and dangerous variants of the virus. Respirators must be designed to effectively fit the wide variety of facial shapes and sizes found around the globe. Respirators must be produced in large enough numbers and at a low enough cost to ensure availability to low- and middle-income countries.

Conclusions

Elastomeric respirators have significant advantages during respiratory pandemics due to their durability and capacity for repeated use. Unlike filtering facepiece respirators, elastomeric respirators allow for robust user seal checks. However, to optimize their utility in healthcare settings, elastomeric respirators should be further enhanced with respect to ease of communication, moisture control, suitability for high-level disinfection, capacity to filter exhaled air, and adaptability to filters from multiple manufacturers. The authors believe that the development, implementation, and stockpiling of improved elastomeric respirators around the world should be an international public health priority.

Research Support

Support was provided solely from institutional and/or departmental sources.

Competing Interests

The authors declare no competing interests.

1.
Editorial: How to be ready for the next influenza pandemic.
Lancet Infect Dis
.
2018
;
18
:
697
2.
Greaney
AJ
,
Starr
TN
,
Gilchuk
P
,
Zost
SJ
,
Binshtein
E
,
Loes
AN
,
Hilton
SK
,
Huddleston
J
,
Eguia
R
,
Crawford
KHD
,
Dingens
AS
,
Nargi
RS
,
Sutton
RE
,
Suryadevara
N
,
Rothlauf
PW
,
Liu
Z
,
Whelan
SPJ
,
Carnahan
RH
,
Crowe
JE
, Jr
,
Bloom
JD
:
Complete mapping of mutations to the SARS-CoV-2 spike receptor-binding domain that escape antibody recognition.
Cell Host Microbe
.
2021
;
29
:
44
57.e9
3.
Science brief: SARS-CoV-2 and surface (fomite) transmission for indoor community environments.
.
4.
Kanamori
H
,
Weber
DJ
,
Rutala
WA
:
Role of the healthcare surface environment in severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) transmission and potential control measures.
Clin Infect Dis
.
2021
;
72
:
2052
61
5.
Fennelly
KP
:
Particle sizes of infectious aerosols: Implications for infection control.
Lancet Respir Med
.
2020
;
8
:
914
24
6.
Gralton
J
,
Tovey
E
,
McLaws
ML
,
Rawlinson
WD
:
The role of particle size in aerosolised pathogen transmission: A review.
J Infect
.
2011
;
62
:
1
13
7.
Tellier
R
,
Li
Y
,
Cowling
BJ
,
Tang
JW
:
Recognition of aerosol transmission of infectious agents: A commentary.
BMC Infect Dis
.
2019
;
19
:
101
8.
Feng
S
,
Shen
C
,
Xia
N
,
Song
W
,
Fan
M
,
Cowling
BJ
:
Rational use of face masks in the COVID-19 pandemic.
Lancet Respir Med
.
2020
;
8
:
434
6
9.
Morawska
L
,
Milton
DK
:
It is time to address airborne transmission of coronavirus disease 2019 (COVID-19).
Clin Infect Dis
.
2020
;
71
:
2311
3
10.
Prather
KA
,
Wang
CC
,
Schooley
RT
:
Reducing transmission of SARS-CoV-2.
Science
.
2020
;
368
:
1422
4
11.
Center for Infectious Disease Research and Policy: Commentary: Covid-19 transmission messages should hinge on science.
.
12.
Greenhalgh
T
,
Jimenez
JL
,
Prather
KA
,
Tufekci
Z
,
Fisman
D
,
Schooley
R
:
Ten scientific reasons in support of airborne transmission of SARS-CoV-2.
Lancet
.
2021
;
397
:
1603
5
13.
NIOSH approved particulate filtering facepiece respirators.
.
14.
Centers for Disease Control
:
Elastomeric respirators.
.
15.
ASTM barrier face coverings.
Available at: https://www.astm.org/DATABASE.CART/WORKITEMS/WK73471.htm. Accessed February 6, 2020
.
16.
ASTM F2100 standard specification for performance of materials used in medical face masks.
Available at: https://www.astm.org/Standards/F2100.htm. Accessed August 30, 2021
.
17.
Cho
KJ
,
Reponen
T
,
McKay
R
,
Shukla
R
,
Haruta
H
,
Sekar
P
,
Grinshpun
SA
:
Large particle penetration through N95 respirator filters and facepiece leaks with cyclic flow.
Ann Occup Hyg
.
2010
;
54
:
68
77
18.
Grinshpun
SA
,
Haruta
H
,
Eninger
RM
,
Reponen
T
,
McKay
RT
,
Lee
SA
:
Performance of an N95 filtering facepiece particulate respirator and a surgical mask during human breathing: Two pathways for particle penetration.
J Occup Environ Hyg
.
2009
;
6
:
593
603
19.
Occupational Safety and Health Agency
:
Assigned protection factors for the revised respiratory protection standard.
Available at: https://www.osha.gov/Publications/3352-APF-respirators.html. Accessed January 3, 2021
.
20.
Coffey
CC
,
Lawrence
RB
,
Campbell
DL
,
Zhuang
Z
,
Calvert
CA
,
Jensen
PA
:
Fitting characteristics of eighteen N95 filtering-facepiece respirators.
J Occup Environ Hyg
.
2004
;
1
:
262
71
21.
ASTM respirator fit capability standard.
Available at: https://www.astm.org/Standards/F3407.htm. Accessed March 13, 2021
.
22.
Bergman
MS
,
Viscusi
DJ
,
Zhuang
Z
,
Palmiero
AJ
,
Powell
JB
,
Shaffer
RE
:
Impact of multiple consecutive donnings on filtering facepiece respirator fit.
Am J Infect Control
.
2012
;
40
:
375
80
23.
NASA
:
HEPA filtration.
.
24.
Van Turnhout
J
,
Adamse
J
,
Hoeneveld
W
: Electret filters for high-efficiency air cleaning.
J Electrostatics
.
1980
;
8
:
369
79
25.
National Institute for Occupational Safety and Health (NIOSH): NIOSH N95 mask testing.
.
26.
Ruskin
KJ
,
Ruskin
AC
,
Musselman
BT
,
Harvey
JR
,
Nesthus
TE
,
O’Connor
M
:
COVID-19, personal protective equipment, and human performance.
Anesthesiology
.
2021
;
134
:
518
25
27.
National Institute for Occupational Safety and Health (NIOSH): Inhalation and exhalation resistance and filtration performance of stockpiled air-purifying respirators: Overall performance of nearly 4,000 respirators sampled from ten stockpile facilities.
.
28.
Jones
JG
:
The physiological cost of wearing a disposable respirator.
Am Ind Hyg Assoc J
.
1991
;
52
:
219
25
29.
Institute of Medicine: The use and effectiveness of powered air purifying respirators in health care: Workshop summary.
.
30.
Howard
RA
,
Lathrop
GW
,
Powell
N
:
Sterile field contamination from powered air-purifying respirators (PAPRs) versus contamination from surgical masks.
Am J Infect Control
.
2020
;
48
:
153
6
31.
Bowdle
A
,
Jelacic
S
,
Shishido
S
,
Munoz-Price
LS
:
Infection prevention precautions for routine anesthesia care during the SARS-CoV-2 pandemic.
Anesth Analg
.
2020
;
131
:
1342
54
32.
Belser
JA
,
Gustin
KM
,
Katz
JM
,
Maines
TR
,
Tumpey
TM
:
Influenza virus infectivity and virulence following ocular-only aerosol inoculation of ferrets.
J Virol
.
2014
;
88
:
9647
54
33.
Belser
JA
,
Lash
RR
,
Garg
S
,
Tumpey
TM
,
Maines
TR
:
The eyes have it: Influenza virus infection beyond the respiratory tract.
Lancet Infect Dis
.
2018
;
18
:
e220
7
34.
Belser
JA
,
Rota
PA
,
Tumpey
TM
:
Ocular tropism of respiratory viruses.
Microbiol Mol Biol Rev
.
2013
;
77
:
144
56
35.
Chen
L
,
Liu
M
,
Zhang
Z
,
Qiao
K
,
Huang
T
,
Chen
M
,
Xin
N
,
Huang
Z
,
Liu
L
,
Zhang
G
,
Wang
J
:
Ocular manifestations of a hospitalised patient with confirmed 2019 novel coronavirus disease.
Br J Ophthalmol
.
2020
;
104
:
748
51
36.
Chu
DK
,
Akl
EA
,
Duda
S
,
Solo
K
,
Yaacoub
S
,
Schünemann
HJ
;
COVID-19 Systematic Urgent Review Group Effort (SURGE) Study Authors
:
Physical distancing, face masks, and eye protection to prevent person-to-person transmission of SARS-CoV-2 and COVID-19: A systematic review and meta-analysis.
Lancet
.
2020
;
395
:
1973
87
37.
Lu
CW
,
Liu
XF
,
Jia
ZF
:
2019-nCoV transmission through the ocular surface must not be ignored.
Lancet
.
2020
;
395
:
e39
38.
Adamson
S
,
Carpenter
H
,
Pang
G
,
Pincus
JM
,
Gregory
BE
,
Reade
MC
:
Staff perceptions of military chemical-biological-radiological-nuclear (CBRN) air-purifying masks during a simulated clinical task in the context of SARS-CoV-2.
Anaesth Intensive Care
.
2021
;
49
:
190
7
39.
Akhtar
J
,
Garcia
AL
,
Saenz
L
,
Kuravi
S
,
Shu
F
,
Kota
K
:
Can face masks offer protection from airborne sneeze and cough droplets in close-up, face-to-face human interactions?-A quantitative study.
Phys Fluids (1994)
.
2020
;
32
:
127112
40.
Center for Infectious Disease Research and Policy: Commentary: Masks-for-all for COVID-19 not based on sound data.
.
41.
Cappa
CD
,
Asadi
S
,
Barreda
S
,
Wexler
AS
,
Bouvier
NM
,
Ristenpart
WD
:
Expiratory aerosol particle escape from surgical masks due to imperfect sealing.
Sci Rep
.
2021
;
11
:
12110
42.
Drewnick
F
,
Pikmann
J
,
Fachinger
F
,
Moormann
L
,
Sprang
F
,
Borrmann
S
:
Aerosol filtration efficiency of household materials for homemade face masks: Influence of material properties, particle size, particle electrical charge, face velocity, and leaks.
Aerosol Sci Tech
.
2021
;
55
:
63
79
43.
Gandhi
M
,
Marr
LC
:
Uniting infectious disease and physical science principles on the importance of face masks for COVID-19.
Med (N Y)
.
2021
;
2
:
29
32
44.
Lindsley
WG
,
Blachere
FM
,
Law
BF
,
Beezhold
DH
,
Noti
JD
:
Efficacy of face masks, neck gaiters and face shields for reducing the expulsion of simulated cough-generated aerosols.
Aerosol Sci Tech
.
2021
;
55
:
449
57
45.
Rengasamy
S
,
Miller
A
,
Eimer
BC
,
Shaffer
RE
:
Filtration performance of FDA-cleared surgical masks.
J Int Soc Respir Prot
.
2009
;
26
:
54
70
46.
Ueki
H
,
Furusawa
Y
,
Iwatsuki-Horimoto
K
,
Imai
M
,
Kabata
H
,
Nishimura
H
,
Kawaoka
Y
:
Effectiveness of face masks in preventing airborne transmission of SARS-CoV-2.
mSphere
.
2020
;
5
:
e00637
20
47.
ACGIH: COVID-19
:
Workers need respirators.
.
48.
CNN: European countries mandate medical-grade masks over homemade cloth face coverings.
.
49.
O’Shaughnessy
PT
,
Strzelecki
B
,
Ortiz-Hernandez
M
,
Aubin
P
,
Jing
X
,
Chang
Q
,
Xiang
J
,
Thorne
PS
,
Stapleton
JT
:
Characterization of performance and disinfection resilience of nonwoven filter materials for use in 3D-printed N95 respirators.
J Occup Environ Hyg
.
2021
;
18
:
265
75
50.
University of Florida mask.
.
51.
52.
Chalikonda
S
,
Waltenbaugh
H
,
Angelilli
S
,
Dumont
T
,
Kvasager
C
,
Sauber
T
,
Servello
N
,
Singh
A
,
Diaz-Garcia
R
:
Implementation of an elastomeric mask program as a strategy to eliminate disposable N95 mask use and resterilization: Results from a large academic medical center.
J Am Coll Surg
.
2020
;
231
:
333
8
53.
Hines
SE
,
Brown
C
,
Oliver
M
,
Gucer
P
,
Frisch
M
,
Hogan
R
,
Roth
T
,
Chang
J
,
McDiarmid
M
:
User acceptance of reusable respirators in health care.
Am J Infect Control
.
2019
;
47
:
648
55
54.
University of Maryland elastomeric respirator implementation guide.
.
55.
National Academies of Sciences Engineering and Medicine
:
Reusable Elastomeric Respirators in Health Care: Considerations for Routine and Surge Use
.
Washington, D.C.
,
National Academies Press
,
2019
56.
Brosseau
LM
,
Jones
RM
,
Harrison
R
:
Elastomeric respirators for all healthcare workers.
Am J Infect Control
.
2021
;
49
:
405
6
57.
Duling
MG
,
Lawrence
RB
,
Slaven
JE
,
Coffey
CC
:
Simulated workplace protection factors for half-facepiece respiratory protective devices.
J Occup Environ Hyg
.
2007
;
4
:
420
31
58.
First
M
:
HEPA filters.
Appl Biosaf
.
1998
;
3
:
33
42
59.
Birkner
JS
,
Kovalchik
S
,
Fung
D
,
Hinds
WC
,
Kennedy
NJ
:
Particle release from respirators, part II: Determination of the effect of tension applied in simulation of removal.
J Occup Environ Hyg
.
2011
;
8
:
10
2
60.
Heuer
JF
,
Crozier
TA
,
Howard
G
,
Quintel
M
:
Can breathing circuit filters help prevent the spread of influenza A (H1N1) virus from intubated patients?
GMS Hyg Infect Control
.
2013
;
8
:
Doc09
61.
Palmiero
AJ
,
Symons
D
,
Morgan
JW
, III
,
Shaffer
RE
:
Speech intelligibility assessment of protective facemasks and air-purifying respirators.
J Occup Environ Hyg
.
2016
;
13
:
960
8
62.
Radonovich
LJ
, Jr
,
Yanke
R
,
Cheng
J
,
Bender
B
:
Diminished speech intelligibility associated with certain types of respirators worn by healthcare workers.
J Occup Environ Hyg
.
2010
;
7
:
63
70
63.
Wentworth
AJ
,
Byrne
JD
,
Orguc
S
,
Sands
J
,
Maji
S
,
Tov
C
,
Babaee
S
,
Huang
HW
,
Boyce
H
,
Chai
PR
,
Min
S
,
Li
C
,
Chu
JN
,
Som
A
,
Becker
SL
,
Gala
M
,
Chandrakasan
A
,
Traverso
G
:
Prospective evaluation of the Transparent, Elastomeric, Adaptable, Long-Lasting (TEAL) Respirator.
ACS Pharmacol Transl Sci
.
2020
;
3
:
1076
82
64.
Bessesen
MT
,
Adams
JC
,
Radonovich
L
,
Anderson
J
:
Disinfection of reusable elastomeric respirators by health care workers: A feasibility study and development of standard operating procedure.
Am J Infect Control
.
2015
;
43
:
1376
65.
Rutala
WA
,
Weber
DJ
:
Disinfection, sterilization, and antisepsis: An overview.
Am J Infect Control
.
2019
;
47S
:
A3
9
66.
WHO boil water.
.