Calibration Curves of Enflurane Using the Beckman LB2® Gas Analyzer with the Halothane Head
Ernesto Goldman, M.D.;
Ernesto Goldman, M.D.
1Instructor, Department of Anesthesiology, University of Colorado Medical Center, Denver, Colorado 80262
*Present address: Department of Anesthesiology, Ohio State University, University Hospital, Columbus, Ohio 43210
Search for other works by this author on:
Duane Sherrill, B.S.;
Duane Sherrill, B.S.
2Research Assistant, Department of Anesthesiology, University of Colorado Medical Center, Denver, Colorado 80262
Search for other works by this author on:
Teruel de Campo, M.D.;
Teruel de Campo, M.D.
3Assistant Professor, Department of Anesthesiology, University of Colorado Medical Center, Denver, Colorado 80262
Search for other works by this author on:
J Antonio Aldrete, M.D.
J Antonio Aldrete, M.D.
4Professor and Chairman, Department of Anesthesiology, University of Colorado Medical Center, Denver, Colorado 80262
Search for other works by this author on:
Anesthesiology July 1980, Vol. 53, 79–81.
Citation
Ernesto Goldman, Duane Sherrill, Teruel de Campo, J Antonio Aldrete; Calibration Curves of Enflurane Using the Beckman LB2® Gas Analyzer with the Halothane Head. Anesthesiology 1980; 53:79–81 doi: https://doi.org/10.1097/00000542-198007000-00020
Download citation file:
Citing articles via
Most Viewed
Related Articles
Technique for Using Video Microscopy and Indicator Dilution for Repeated Measurements of Cardiac Output in Small Animals
Anesthesiology (March 2001)
Validation and Calibration of the Risk Stratification Index
Anesthesiology (April 2017)
The Invention and Development of Blood Gas Analysis Apparatus
Anesthesiology (July 2002)