To the Editor:--Proper positioning of the endotracheal tube (ETT) commonly is confirmed by auscultation of bilateral breath sounds. However, Brunel et al. reported that 60% of mainstem bronchial intubations occurred despite the presence of equal breath sounds from the bilateral chest. Schwartz et al. noted that endobronchial intubation was not detected by chest auscultation. We used a fiberoptic bronchoscope (FOB) to evaluate the reliability of auscultatory confirmation of ETT position.

After nasotracheal intubation, the FOB (Olympus LF-2) was inserted through the ETT. Then, the FOB and the ETT with the cuff inflated were advanced. When it was confirmed with the FOB that the ETT tip reached the carina, a mark was made on the ETT at the nares. When unilateral breath sounds became first more bronchial in quality and then disappeared, second and third marks were made on the ETT, respectively. In this fashion, the distance from the carina to the nares and the distances from the ETT tip to the nares at change and at disappearance of breath sounds were measured.

The carina was 28.7 plus/minus 1.2 cm from the nares in females who averaged 156 plus/minus 7 cm in height. The ETT tip advanced beyond the carina invariably entered the right mainstem bronchus at 30.1 plus/minus 1.3 and 31.9 plus/minus 1.0 cm from the nares at change and disappearance of breath sounds, respectively. The ETT used in this study, Portex Blue Line Tracheal Tube (ID 7.5 mm), had a cuff 2.5 cm from the tip, a 1.0-cm bevel facing to the left, and no Murphy's eye. Taking into consideration these structural features, it was presumed that breath sounds changed when the proximal end of the bevel was advanced about 0.5 cm into the right mainstem bronchus, perhaps because of gas flow through the narrow space between the ETT and the bronchus. Breath sounds disappeared when the cuff was advanced about 0.7 cm into the intubated bronchus, possibly because of interruption of gas supply to the opposite bronchus. Therefore, the ETT tip should be withdrawn at least 1.5 or 3.2 cm if breath sounds from the left side of the chest change or disappear, respectively, after intubation and during anesthesia. Our results suggest that the fiberoptic bronchoscopy is more reliable than chest auscultation in confirming ETT position.

Kazuna Sugiyama, D.D.S., Ph.D., Professor.

Kozo Yokoyama, D.D.S., Ph.D., Research Associate, Department of Anesthesia, Kagoshima University Dental Hospital, 8–35–1 Sakuragaoka, Kagoshima 890, Japan.

1.
Schwartz N: Monitoring bilateral breath sounds. Anesth Analg 73:828, 1991.
2.
Owen RL, Cheney FW: Endobronchial intubation: A preventable complication. ANESTHESIOLOGY 67:255-257, 1987.
3.
Brunel W, Coleman DL, Schwartz DE, Peper E, Cohen NH: Assessment of routine chest roentgenograms and the physical examination to confirm endotracheal tube position. Chest 96:1043-1045, 1989.
4.
Schwartz DE, Lieberman JA, Cohen NH: Women are at greater risk than men for malpositioning of the endotracheal tube after emergent intubation. Crit Care Med 22:1127-1131, 1994.
5.
Kubota Y, Toyoda Y, Kubota H, Yamasaki Y: Confirmation of endotracheal tube positioning. Chest 98:1306, 1990.