To the Editor:-Findlay et al. propose that pulse oximetry can be used to confirm adequate foot perfusion during radical perineal prostatectomy. [1]Specifically, they recommend that a pulse oximeter probe be attached to a toe when the legs are positioned in a fashion that potentially compromises extremity perfusion. Their theory is that “when the foot is perfused, the pulse oximeter displays a normal tracing.” Conversely, “when the blood pressure decreases below a threshold value for that patient, no pulse waveform is displayed.”

That pulse oximetry might be a useful measure of perfusion is intuitive and has been proposed numerous times. [2–6]The difficulty with this approach, however, is that pulse oximeters are “too good.” Specifically, they contain extremely powerful amplifiers that can detect saturation and display a waveform even when flow is critically compromised. A further difficulty is that oximeters actually evaluate arterial pulsation, not flow per se. The result is a substantial potential for false-normal results.

The consensus among vascular and hand surgeons is that prolonged digital systolic arterial blood pressures < 40 mmHg are likely to result in tissue injury. [7–11]My concern about using pulse oximetry as an index of flow is that the technique appears unreliable. In one study, for example, the technique failed to detect critical reductions in distal extremity pressure in two of three cases. [12]Another study demonstrated that the pulse oximeter signal is maintained (without even a “low perfusion” warning) until flow is reduced to [nearly =] 8% of normal. [13]Pulse oximetry obviously detects the most extreme reductions in tissue flow-those bordering on complete ischemia. However, the method proposed by Findlay et al. [1]seems likely to provide false reassurance in a substantial number of cases where flow is actually seriously restricted.

I would thus like to propose an equally easy, but potentially more reliable, measure of foot perfusion: ankle blood pressure. Simply position the cuff of an oscillometric blood pressure monitor around the ankle and confirm that the systolic arterial pressure exceeds 40 mmHg. An alternative is to attach a cuff sized for premature infants to the long toe. Obviously, pressure at these sites should be monitored at relatively infrequent intervals, say [nearly =] 10-min intervals, so the measurements themselves do not excessively restrict flow.

Human extremities are, fortunately, relatively resistant to ischemia. It would nonetheless be risky to maintain critically low foot perfusion for the duration of a radical prostatectomy or similar procedure. Measurements of ankle blood pressure will alert anesthesiologists to inadequate perfusion, allowing interventions that might include increasing systemic blood pressure or altering patient position.

Daniel I. Sessler, M.D.

Department of Anesthesia; University of California, San Francisco; 374 Parnassus Avenue, 3rd Floor; San Francisco, California 94143–0648

sessler@vaxine.ucsf.edu

(Accepted for publication December 9, 1997.)

1.
Findlay JY, Rettke SR, Myers RP: Perfusion monitoring during radical perineal prostatectomy (letter). Anesthesiology 1997; 87:457.
2.
Nowak GS, Moorthy SS, McNiece WL: Use of pulse oximetry for assessment of collateral arterial flow. Anesthesia 1986; 64:527-8.
3.
Matuski A: A modified Allen's test using a pulse oxymeter. Anaesth Intensive Care 1988; 16:126.
4.
Rozenberg B, Rozenberg M, Birkhan J: Allen's test performed by pulse oximeter. Anesthesia 1988; 43:515-6.
5.
Griffiths DM, Ilsley AH, Runciman WB: Pulse meters and pulse oximeters. Anaesth Intensive Care 1988; 16:49.
6.
Raju R: The pulse oximeter and the collateral circulation. Anaesthesia 1986; 41:783-4.
7.
Yao ST, Needham TN, Gourmoos C, Irvine WT: A comparative study of strain-gauge plethysmography and Doppler ultrasound in the assessment of occlusive arterial disease of the lower extremities. Surgery 1972; 71:4-9.
8.
Hirai M: Arterial insufficiency of the hand evaluated by digital blood pressure and arteriographic findings. Circulation 1978; 58:902-8.
9.
Downs AR, Gaskell P, Morrow I, Munson CL: Assessment of arterial obstruction in vessels supplying the fingers by measurement of local blood pressures and the skin temperature response test-Correlation with angiographic evidence. Surgery 1975; 77:530-9.
10.
Bell G, Nielsen PE, Wolfson J, Ulrich J, Engell HC, Lassen NA: Measurements of systolic pressure in the limbs of patients with arterial occlusive disease. Surg Gynecol Obstet 1973; 136:177-81.
11.
Carter SA, Lezak JD: Digital systolic pressures in the lower limb in arterial disease. Circulation 1971; 43:905-14.
12.
Levinsohn DG, Gordon L, Sessler DI: The Allen's test: Analysis of four methods. J Hand Surg [Am] 1991; 16A:279-82.
13.
Lawson D, Norley I, Korbin G: Blood flow limits and pulse oximeter signal detectiony. Anesthesiology 1987; 67:599-603.