In Reply:-In our study, [1]we connected the negative (cathodal) electrode of the nerve stimulator to the insulated needle. We reported that we considered our “stimulating needle to be close to the nerve when the stimulating current that elicited a motor response was <1 mA.” This was the same current reported by Pither et al., [2]Singelyn et al., [3]and Mansour. [4]We actually used currents less than 1 mA. I reexamined our data and found that the currents used ranged from 0.5 to 0.8 mA (mean +/- SD, 0.71 +/- 0.1 mA). The stimulation intensities used for each elicited motor response were as follows:(1) eversion: 0.63 +/- 0.12 (range, 0.5–0.8 mA);(2) inversion: 0.75 +/- 0.12 (range, 0.5–0.8 mA);(3) plantar flexion: 0.7 +/- 0; and (4) dorsiflexion: 0.77 +/- 0.10 (range, 0.7–0.8 mA).

The ideal stimulus intensity required to stimulate a nerve has not been clearly defined. Pither et al. stated that movement is elicited in the appropriate muscle when the needle tip is likely to be 1–2 cm from the nerve and that 0.5–1 mA is required when the needle is touching or very close to the nerve. [2]Magora et al. showed that 0.5 mA was needed for direct stimulation of the obturator nerve and that their blocks were unsuccessful when 1–3 mA were needed to elicit a motor response. [5]Riegler found that currents ranging from 0.2 to 1.5 mA were sufficient for localization of the brachial plexus whether the interscalene, supraclavicular, or axillary approach was used. [6]A review of the minimum currents used by Riegler showed 0.66 +/- 0.4 mA for the interscalene, 0.71 +/- 0.03 mA for the supraclavicular, and 0.72 +/- 0.03 mA for the axillary approach. These stimulating intensities are the same as the ones we used in our study. Shannon et al. accepted 0.6 mA as their endpoint with their lateral femoral cutaneous block. [7]In the new edition of Cousins and Bridenbaugh, a stimulus intensity of 0.5 mA was recommended. [8] 

Our hesitancy to use currents less than 0.3 mA was precipitated by the occurrence of an intraneural injection with 0.2 mA during one of our trial blocks, before we formally started our study. The subject had severe shooting pain to his foot during the initial injection of 2 ml, and although the needle was withdrawn 1 mm, he had paresthesias for 1 week. Singelyn et al. noted the occurrence of paresthesias with stimulating currents “less than 1 mA.”[3]In our clinical practice, we use stimulating intensities of 0.3–0.8 mA.

Bridenbaugh and Crews stated that the “injection of 1 to 2 ml of local anesthetic will immediately abolish nerve stimulation and muscle contraction if the tip of the needle is at the site of the nerve.”[8]This has been our experience, as well as that of other investigators. [2,9,10]As stated by Dr. Vloka, this rapid response is the result of the nerve being displaced away from the needle tip. This phenomenon has been confirmed in studies wherein air produced the same response as the local anesthetic. [11]If the needle tip is beyond the nerve and if the shaft of the needle is causing the stimulation, then the injection will not change the motor response. In this case, the needle should be withdrawn slightly and the test repeated.

In our initial study sessions, we used initial currents of 2–3 mA because these were the currents recommended by Riegler. [6]We then decreased the current output as we approached the target nerve. After four study sessions, we used initial currents of 1.5–2 mA because 3 mA was painful. Dr. Vloka will probably agree that the initial current used is less important than the actual current when the injection was made.

Dr. Vloka made calculations based on “1.0-mA stimulus intensity used by Benzon” when we stated clearly that we used currents “less than 1 mA.” Perhaps it was our fault and may be we should have been more specific.

The 98% success rate of Vloka and Hadzic is to be expected because they “stimulated the division of the popliteal nerve that predominantly innervated the surgical area.”[12]It was also not surprising that the common peroneal nerve was the nerve that was usually stimulated first in their lateral approach group because the common peroneal nerve is located laterally, in relation to the tibial nerve. These two points emphasize the importance of knowledge of the anatomy involved; simply demonstrating nerve stimulation at low current is not enough to ensure adequate block. [2,13] 

Dr. Vloka stated that they performed sensory evaluations in their study. [12]However, they did not assess, in detail, the sensory blockade of the areas in the foot innervated by the different branches of the sciatic nerve, i.e., the posterior tibial, deep peroneal, superficial peroneal, and sural nerves. Incomplete blockade of some of the areas innervated by the branches of the sciatic nerve may have been masked by adequate sensory anesthesia in the operative area.

Although there may be a common epineural sheath as Dr. Vloka mentioned, there may also be a sheath within the nerves. In two study sessions in our study, [1]we found that partial blockade of the posterior tibial nerve involved the area innervated by the medial calcaneal branch of the tibial nerve, with no blockade of the medial and lateral plantar branches.

Based on Dr. Vloka's publications, it appears that the lateral approach to sciatic nerve blockade in the popliteal fossa is reliable and should be used more frequently. We use the posterior approach simply because of familiarity with the technique. We have used this approach even in patients in the lateral position and found it simple and effective. If we find that inversion or combined inversion/plantar flexion (the two elicited foot movements associated with complete sensory blockade of the foot [1]) is difficult to elicit, then we use the double injection technique. [14]In this technique, we inject two 15-ml injections after identification of the tibial (elicited plantar flexion) and peroneal components (elicited dorsiflexion or eversion) of the sciatic nerve.

Honorio T. Benzon, M.D.

Professor of Anesthesiology; Northwestern University Medical School; Chicago, Illinois

(Accepted for publication January 8, 1998.)

1.
Benzon HT, Kim C, Benzon HP, Silverstein ME, Jericho B, Prillaman K, Buenaventura R: Correlation between evoked motor response of the sciatic nerve and sensory blockade. Anesthesiology 1997; 87:547-52.
2.
Pither CE, Raj PP, Ford DJ: The use of peripheral nerve stimulators for regional anesthesia. A review of experimental characteristics, technique, and clinical applications. Reg Anesth 1985; 10:49-58.
3.
Singelyn FJ, Gouverneur JA, Gribomont BF: Popliteal sciatic nerve block aided by nerve stimulator: A reliable technique for foot and ankle surgery. Reg Anesth 1991; 16:278-81.
4.
Mansour N: Compartment block for foot surgery. A new approach to tibial nerve and common peroneal nerve block. Reg Anesth 1995; 20:95-9.
5.
Magora E, Rozin R, Ben-Menachem Y, Magora A: Obturator nerve block: an evaluation of technique. Br J Anaesth 1969; 41:695-8.
6.
Riegler F: Brachial plexus block with the nerve stimulator: Motor response characteristics at three sites. Reg Anesth 1992; 17:295-9.
7.
Shannon J, Lang SA, Yip RW, Gerard M: Lateral femoral cutaneous nerve block revisited. A nerve stimulator technique. Reg Anesth 1995; 20:100-4.
8.
Bridenbaugh PO, Crews JC: Perioperative management of patients for neural blockade, Neural Blockade in Clinical Anesthesia and Management of Pain, 3rd Edition. Edited by Cousins MJ, Bridenbaugh PO. Philadelphia, Lippincott-Raven Publishers, 1997, pp 179-99.
9.
Sims J: A modification of landmarks for infraclavicular approach to brachial plexus block. Anesth Analg 1977; 56:554-5.
10.
Montgomery SJ, Raj PP, Nettles D, Jenkins MT: The use of the nerve stimulator with standard unsheathed needles in nerve blockade. Anesth Analg 1973; 52:827-31.
11.
Raj PP: Adjuvant techniques in regional anesthesia, Handbook of Regional Anesthesia. Edited by Raj PP. New York, Churchill-Living-stone, 1985, pp 250-8.
12.
Hadzic A, Vloka JD: The lateral approach to popliteal nerve block. A comparison with posterior approach. Anesthesiology 1998; accepted for publication.
13.
Smith BL: Efficacy of a nerve stimulator in regional analgesia: Experience in a resident training programme. Anaesthesia 1976; 31:778-82.
14.
Bailey SL, Parkinson SK, Little WL, Simmerman SR: Sciatic nerve block. A comparison of single versus double injection technique. Reg Anesth 1994; 19:9-13.