To the Editor:-The interesting article by Bratzke et al. [1] uses an unusual measure of ventilatory efficiency: partial pressure of arterial carbon dioxide (PaCO(2))/minute ventilation. This measure is adequate only during very restricted circumstances. As the authors point out, when comparing two modes of ventilation in rapid succession, an increase in the ratio equates with more efficient ventilation. However, the comparison assumes that carbon dioxide output (VCO(2)) and PaCO(2) are "clamped." If cellular carbon dioxide production were to increase, say by 10%, unchanged ventilation would cause PaCO(2) to increase by 10%. The ratio would then also increase by 10%, but this would not signify greater efficiency. In order not to be misleading, this index can only apply to changes in minute ventilation at constant PaCO(2).

For all-around applicability, an index of ventilatory efficiency must always compare expired and blood gas partial pressure of carbon dioxide (PCO(2)), the basis of the dead space concept. The nearest index to the one the authors have used here, i.e., one that increases with increasing efficiency, would be (1 - VD/ VT), which is obtained from mixed expired PCO(2) divided by PaCO(2). It is the same thing as alveolar ventilation divided by total ventilation.

A second point about the article is that the assumption that anatomic dead space was constant may not be correct. Increases in end-inspiratory pause of a fraction of a second can reduce anatomic dead space significantly (L. Nordstrom, personal communication, November, 1998), and, therefore, the long continuous positive airway pressure (CPAP) periods used by Bratzke et al. [1] may markedly reduce this dead space, contributing considerably to the increase in alveolar ventilation noted.

Roger Fletcher, M.D., F.R.C.A.

Consultant Anesthetist; Directorate of Anesthesia; Manchester Royal Infirmary; Manchester M13 9ES, United Kingdom

(Accepted for publication December 17, 1998.)

1.
Bratzke E, Downs JB, Smith RA: Intermittent CPAP: A new mode of ventilation during general anesthesia. Anesthesiology 1998; 89:334-40