RAPID and accurate diagnosis of the misplacement of an endotracheal tube in the esophagus remains one of the most important problems in anesthesia and intensive care. Clinical signs observed through auscultation, inspection of thoracic excursions, refilling of the breathing bag, or water vapor condensing at the endotracheal tube may be misleading in as many as 30% of patients. [1] Although end-tidal carbon dioxide measurement and bronchoscopy remain the gold standards for evaluating the endotracheal tube position, the esophageal detector device (EDD) recently has assumed an important clinical role, with sensitivity and specificity rates for identifying the endotracheal tube position in adults of nearly 100%. [2–8] An EDD may even indicate correct endotracheal tube position in situations in which the end-tidal carbon dioxide concentration fails, such as cardiopulmonary resuscitation or severe bronchospasm. [8] Most importantly, misdiagnosis of tubes placed in the esophagus using an EDD in adults is rare. [6,9] Of more than 1,000 reported applications, only two cases have been described in which air could be aspirated from a tube placed in the esophagus: one in a morbidly obese patient [7] and one in a patient with gastric distension after vigorous attempts to apply mask ventilation. [8] Unreliable results also were found in infants younger than 1 yr [10] or when technically unsuited EDDs were used in adults (i.e., those with a 20-ml bulb).

We report the misdiagnosis of an esophageally placed endotracheal tube with a 60-ml syringe-type EDD in an adult with unexpected laryngospasm after reinsertion of an intubating laryngeal mask airway (ILMA), [11] in whom gastric distension of the stomach had occurred as a result of attempted positive-pressure ventilation.

A 44-yr-old man (weight, 90 kg) was scheduled for reconstruction of the mandible 1 yr after radical resection of a tonsillar carcinoma with suprahyoidal and neck dissection bilaterally and radionecrosis of the mandible thereafter. A percutaneous gastrostomy had been performed because the patient could not swallow. Examination of the airway showed a slightly reduced thyromental distance (5.5 cm), a moderately reduced mouth opening (3.2 cm awake, 4.5 cm during anesthesia), a fixed retrognathia, and a 1.5-cm right-sided displacement of the thyroid cartilage by scar tissue. General anesthesia was started with 4 mg/kg thiopentone; 0.1 mg fentanyl; and enflurane, 2.5 vol%, in 100% oxygen. Ventilation by face mask was achieved readily, and 0.5 mg/kg atracurium was given. Modest insufflation of the stomach was apparent during mask ventilation and also during ventilation with a correctly inserted ILMA at normal airway pressures (15 cm water) well below the leak pressure of the ILMA (40 cm water).

This patient was a participant in a study to compare the success of blinded intubation through an ILMA versus a standard laryngeal mask airway. His trachea was intubated successfully using an ILMA, but a tube could not be placed through a standard laryngeal mask airway. Therefore, we decided to use blinded intubation through the ILMA, again followed by a tube exchange maneuver for definite nasal intubation. [12] We reinserted the ILMA a second time without difficulty 30 min after induction. Laryngospasm occurred but was not detected immediately by the ventilating anesthetist, who was a novice user of the ILMA. Gastric ventilation occurred with the ILMA but was difficult to detect because of the patient's obesity. Thoracic excursions with ventilation could be seen in the xiphoid region; they were slightly more pronounced in the left hemithorax with no movement in the infraclavicular parts. No expired carbon dioxide was detected. Blinded tracheal intubation through the ILMA was attempted using a 7.5-mm ID endotracheal tube without the application of force to an approximate depth of 24 cm. The position of the endotracheal tube was identified using a 60-ml syringe-type EDD. One hundred twenty milliliters of air was aspirated readily, followed by two breaths that did not deliver carbon dioxide. Atracurium (15 mg) was given. The stomach was emptied by manual compression and the EDD was used again. After aspirating another 90 ml air, further air aspiration was no longer possible, indicating the esophageal position of the endotracheal tube by the EDD. The endotracheal tube was removed. Ventilation of the lungs via the ILMA was possible, as apparent by carbon dioxide detected in the expired air. Blinded oral intubation through the ILMA was performed, and subsequently this tube was exchanged for a nasotracheal tube. The operation was uncomplicated and the patient had no further problems. Oxyhemoglobin saturation by pulse oximetry was greater than 92% during the entire procedure.

This case shows that an EDD may not detect esophageal positioning of an endotracheal tube when gastric inflation has occurred inadvertently. This can happen with a face mask, with all types of laryngeal masks, or with an esophageally placed endotracheal tube when positive-pressure ventilation is performed. Repeated gastric inflation also may fill the esophagus with air, which apparently happened in our patient. We initially questioned whether a mispositioned ILMA might have facilitated such distension, but the correct insertion procedure, the position of the ILMA, and correct ventilation after relaxation do not support a malposition of the ILMA after its second insertion. However, users of the ILMA regularly find an esophageal tube position if it is diverted from the laryngeal inlet by laryngospasm.

Studies with an EDD reported a sensitivity rate of 100% for discriminating esophageal from endotracheal intubation in healthy adults classified as American Society of Anesthesiologists grades I and II who were undergoing routine operations, [1,2–4,13] in children between the ages of 2 and 10 yr, [10] or when the EDD was used by paramedics. [14] Even a nasogastric tube or an endotracheal tube placed in the esophagus did not interfere with the reliability of detecting esophageal intubation. [6] One consistent finding in most studies was delayed refill of the bulb-type EDD, which occurred in 6–30%, [3,4,6–8] which may be interpreted falsely as the mispositioning of the endotracheal tube when, in fact, it was correctly placed.

Some studies have directly addressed the problem of EDD failure in the setting of previous air insufflation into the esophagus. In a study that used one pig in which the esophagus was ventilated with a bag via a tube for 1 min before the EDD was tested, 30 investigators did not fail to discriminate endotracheal from esophageal intubation. [1] Salem et al. [5] reported that the delivery of three small breaths (tidal volume of 300–350 ml each) through a tube placed in the esophagus to mimic gastric insufflation did not affect the reliability of the self-inflating bulb, indicating neither instantaneous nor delayed reinflation. This led to the conclusion that false-positive findings with an EDD are "probably nonexistent" [6] or "only theoretical." [9] However, Lang et al. [7] reported a misdiagnosis in 1 of 54 grossly obese patients when the bulb was squeezed through an esophageally placed endotracheal tube before aspiration. A second case of an esophageal intubation undetected by an EDD was reported after vigorous attempts at mask ventilation before testing with a bulb-type EDD. [8] In addition, the EDD was found to be unreliable in children younger than 1 yr [10] and in adults if the bulb used was too small.

In conclusion, we add one more case in which an EDD failed to show correct endotracheal tube placement in a situation of repeated ventilation of the stomach. If there is any doubt about correct endotracheal tube placement, all achievable indicators of correct endotracheal tube placement should be sought: direct visualization of the transglottic tube passage, auscultation of the epigastrium and the chest, measurement of end-tidal carbon dioxide concentration for a series of breaths, and bronchoscopy.

1.
Foutch RG, Magellsen MD, MacGilen JG: The esophageal detector device: A rapid and accurate method for assessing tracheal versus esophageal intubation in a porcine model. Ann Em Med 1992; 21:1073-6
2.
Wee MYK: The oesophageal detector device: Assessment of a new method to distinguish oesophageal from tracheal intubation. Anaesthesia 1988; 43:27-9
3.
Williams KN, Nunn FJ: The oesophageal detector device. Anaesthesia 1989; 44:412-14
4.
Zaleski L, Abello D, Gold MI: The oesophageal detector device. Does it work? Anesthesiology 1993; 79:244-7
5.
Salem MR, Wafai Y, Baraka A, Taimorrazy B, Joseph NJ, Nimmagadda U: Use of the self-inflating bulb for detecting esophageal intubation after 'esophageal ventilation.' Anesth Analg 1993; 77:1227-31
6.
Salem MR, Wafai Y, Joseph NJ, Baraka A, Czinn EA: Efficacy of the self-inflating bulb in detecting esophageal intubation. Anesthesiology 1994; 80:42-8
7.
Lang DJ, Wafai Y, Salem MR, Czinn EA, Halim AA, Baraka A: Efficacy of the self-inflating bulb in confirming tracheal intubation in the morbidly obese. Anesthesiology 1996; 85:246-53
8.
Kasper CL, Deem S: The self-inflating bulb to detect esophageal intubation during emergency airway management. Anesthesiology 1998; 88:898-902
9.
Benumof JL: The ASA difficult airway algorithm. 47th Annual Refresher Course Lectures and Clinical Update Program, edited by The American Society of Anesthesiologists, 1996; 24:1-7
10.
Haynes SR, Morton NS: Use of the oesophageal detector device in children under one year of age. Anaesthesia 1990; 45:1067-9
11.
Brain AIJ, Verghese C, Addy EV, Kapila A: The intubating laryngeal mask. I: Development of a new device for intubation of the trachea. Br J Anaesth 1997; 79:699-703
12.
Langenstein H: Die Kehlkopfmaske bei schwieriger Intubation. Anaesthesist 1995; 44:712-18
13.
Mathias JA: Oesophageal detector device. Anaesthesia 1989; 44:931
14.
Donahue PL: The oesophagus detector device. Anaesthesia 1994; 49:863-5