Background

In the partial CO(2) rebreathing method, monitored changes in CO(2) elimination and end-tidal CO(2) in response to a brief rebreathing period are used to estimate cardiac output. However, dynamic changes in CO(2) production during ischemia and reperfusion may affect the accuracy of these estimates. This study was designed to compare measurements of cardiac output as produced by the partial CO(2) rebreathing (NICO), bolus (BCO), and continuous thermodilution (CCO) methods of monitoring cardiac output.

Methods

Cardiac output was continuously monitored using both NICO and CCO in 28 patients undergoing aortic reconstruction. BCO measurements were taken at the following intervals when hemodynamic stability was achieved: (1) after anesthetic induction; (2) during aortic cross-clamp; (3) at reperfusion of the iliac artery; and, (4) during peritoneal closure.

Results

The bias and precision (1 SD) derived from all the measurements between NICO and BCO was -0.58 +/- 0.9 l/min, whereas for CCO and BCO it was 0.38 +/- 1.17 l/min. The bias between NICO and BCO was small after anesthetic induction and during cross-clamp, but increased following reperfusion. The bias between CCO and BCO was relatively small until reperfusion but increased significantly at peritoneal closure.

Conclusions

Results indicate that in aortic reconstruction surgery the performance of NICO monitoring is comparable with that of CCO; however, the direction of bias in these continuous measurement devices is the opposite.

MAINTAINING optimal cardiac output results in improved outcomes and therefore is an important goal of intraoperative hemodynamic management. 1–3However, the pulmonary artery catheter, which is used as a reference method to measure cardiac output, is invasive, and several investigations report less favorable outcomes associated with its use. 4–7To address this problem, several less invasive and continuous cardiac output monitoring methods were developed, such as transesophageal Doppler, arterial pulse contour analysis, thoracic bioimpedance, and partial CO2rebreathing. 8Of these methods, the partial CO2rebreathing method offers several advantages as an intraoperative monitor, including noninvasiveness, nonreliance on operator skill, and resistance to electromagnetic interference. However, the accuracy and clinical applicability of the Noninvasive Cardiac Output Monitor (NICO, Novametrix Medical Systems, Wallingford, CT) has not been fully evaluated. This system uses changes in CO2elimination and end-tidal CO2measured in response to a brief rebreathing period to calculate pulmonary blood flow. Because the NICO system relies on stable CO2production and a constant Etco2-Paco2gradient during the measurements, 9,10the dynamic changes in CO2production during ischemia and reperfusion that characterize patients undergoing aortic reconstruction may affect the accuracy of this method. We designed this prospective study to compare the accuracy of the NICO monitor and continuous thermodilution cardiac output versus  standard bolus thermodilution in patients undergoing aortic reconstruction surgery.

After institutional review board approval, 28 patients undergoing elective aortic reconstruction for infrarenal abdominal aortic aneurysm were enrolled in this study when their informed consent was obtained. Anesthetic management was standardized in these patients as follows. After epidural catheterization at the Th10/11 or 11/12 interspace, anesthesia was induced with intravenous fentanyl and propofol and maintained with sevoflurane inhalation with or without nitrous oxide. Patients were paralyzed with vecuronium and mechanically ventilated with either the AS/3 ADU (Datex-Ohmeda, Helsinki, Finland) or KION (Siemens, Solna, Sweden) anesthesia machine. Tidal volume and respiratory rate were initially set at 10 ml/kg and 10 breaths per minute, respectively, and were adjusted to maintain Paco2at 35–45 mmHg. After tracheal intubation, cardiac output was continuously monitored using the NICO monitor (software version 3.1). Cardiac output data obtained with the NICO in average mode were downloaded to a computer for analysis. The NICO sensor was placed distal to the heat and moisture exchanger (Hygrobac S, DAR-Mallinckrodt, Mirandola, Italy). Meticulous attention was paid to maintenance of adequate rebreathing circuit volume during monitoring. An 8-French pulmonary artery catheter (746HF8, Baxter Healthcare, Irvine, CA) was inserted via  the right internal jugular vein. Continuous thermodilution cardiac output (CCO) and mixed venous oxygen saturation were continuously monitored with a Vigilance monitor (Baxter Healthcare, Irvine, CA). The calibration of mixed venous oxygen saturation and subsequent blood gas analysis were performed using a standard blood gas analyzer with Cooximeter (Stat Profile M, Nova Biochemicals, Waltham, MA). These data were downloaded to another computer every 30 s, and data averaged for 3 min before bolus thermodilution cardiac output (BCO) measurement were used for analysis. Fluid administration, blood transfusion, epidural injection of local anesthetics, administration of inotropic and vasodilatory drugs, and other anesthetic management procedures were performed at the discretion of the attending anesthesiologist.

BCO measurements were made at the following predetermined time points: (1) after anesthetic induction (Postinduction); (2) during aortic cross-clamping (XC); (3) at reperfusion of unilateral iliac artery (Declamp); and (4) during peritoneal closure (Endop). The injection of ice-cold saline was repeated four times within 3 min and averaged data were used for analysis. These BCO measurements were performed following the stabilization of intravascular fluid status and cardiac performance. This period was arbitrarily defined as being at least 5 min, during which there was no fluid challenge or change in pharmacologic intervention. Blood gas analysis data were input to the NICO monitor before each BCO measurement to allow precise estimations of shunt fraction and NICO values. At each BCO measurement, the shunt fraction was calculated using the standard formula (see  Appendix).

NICO monitoring was terminated postoperatively, but CCO monitoring was continued during the postoperative period in some patients when clinically indicated. After patients were fully awake and showed stable hemodynamics and acceptable ventilation and oxygenation, they were extubated in the operating room before being transferred to the intensive care unit.

Data are expressed as mean ± SD and were statistically analyzed using the Statview 5.0 software (SAS Institute, Chicago, IL). Correlations between NICO and BCO data and between CCO and BCO data sets were determined by linear regression. A Bland-Altman plot was used to compare the bias (the mean of the differences) and precision (SD of bias) of NICO and CCO against BCO. 11In addition, the relative error (expressed in %) was defined as 100*([either NICO or CCO-BCO]/BCO). 12Intraoperative changes in other parameters were analyzed using a one-way ANOVA with repeated measurements. A value of P < 0.05 was considered significant. Multiple comparisons with Scheffé correction were performed when significant changes were detected.

Demographic and operative data are summarized in table 1. Intraoperative cardiac output as measured by NICO, CCO, and BCO; other variables are summarized in table 2. The interval between the aortic cross-clamp and the second measurement (XC), and the aortic declamp and the third measurement (Declamp) were 17 ± 5 and 20 ± 6 min, respectively. Overall correlations between NICO and BCO and CCO and BCO are expressed in figure 1. These data show that both NICO and CCO measurements correlated significantly with BCO, and that the correlation coefficients were remarkably similar (r = 0.80 and 0.81, respectively). Figure 2shows the Bland-Altman plots of NICO and CCO against BCO. Overall, the bias and precision between NICO and BCO was −0.58 ± 0.90 l/min. The bias and precision between CCO and BCO for all measurements was 0.38 ± 1.17 l/min. These data show that NICO underestimated cardiac output, whereas CCO overestimated cardiac output, when compared with BCO measurements.

Table 1. Demographic and Operative Data

Data are expressed as mean ± SD.

Table 1. Demographic and Operative Data
Table 1. Demographic and Operative Data

Table 2. Intraoperative Data at Four Operative Stages

Data are collected from 28 patients and expressed as mean ± SD. P/F ratio: Pao2/Fio2.

*P < 0.05 vs.  after anesthetic induction. P < 0.05 vs. unilateral iliac reperfusion and peritoneal closure. Qs/Qt was calculated using the following equation: Qs/Qt = (Cco2· Cao2)/(Cco2· Cvo2)*100.

BCO, CCO, and NCO = bolus thermodilution, continuous thermodilution, and noninvasive cardiac output; Cco2, Cao2, and Cvo2= oxygen content of capillary, arterial, and mixed venous blood, respectively; Da-etco2= difference between Paco2and Petco2; declamp = at reperfusion of unilateral iliac artery; endop = during peritoneal closure; postinduction = after anesthetic induction; XC = during aortic cross-clamping.

Table 2. Intraoperative Data at Four Operative Stages
Table 2. Intraoperative Data at Four Operative Stages

Fig. 1. Scatter plot of individual NICO and bolus thermodilution (BCO) values (left ) and continuous thermodilution (CCO) and BCO values (right ). Values from both CCO monitors significantly correlated with BCO (NICO = 0.80 * BCO + 0.35, r = 0.80) and with CCO (CCO = 1.11 * BCO − 0.14, r = 0.81). Filled line = correlation line; dotted line = line of identity.

Fig. 1. Scatter plot of individual NICO and bolus thermodilution (BCO) values (left ) and continuous thermodilution (CCO) and BCO values (right ). Values from both CCO monitors significantly correlated with BCO (NICO = 0.80 * BCO + 0.35, r = 0.80) and with CCO (CCO = 1.11 * BCO − 0.14, r = 0.81). Filled line = correlation line; dotted line = line of identity.

Close modal

Fig. 2. Bland-Altman plot of cardiac output measurements from bolus thermodilution (BCO) and NICO (left ) and from BCO and continuous thermodilution (CCO) (right ). Filled circle = after anesthetic induction (Postinduction); open circle = during aortic cross-clamping (XC); filled box = at reperfusion of unilateral iliac artery (Declamp); open box = during peritoneal closure (Endop). n = 28 measurements in each stage; total = 112 measurements. Results show that NICO has a tendency to underestimate BCO, whereas CCO has a tendency to overestimate BCO.

Fig. 2. Bland-Altman plot of cardiac output measurements from bolus thermodilution (BCO) and NICO (left ) and from BCO and continuous thermodilution (CCO) (right ). Filled circle = after anesthetic induction (Postinduction); open circle = during aortic cross-clamping (XC); filled box = at reperfusion of unilateral iliac artery (Declamp); open box = during peritoneal closure (Endop). n = 28 measurements in each stage; total = 112 measurements. Results show that NICO has a tendency to underestimate BCO, whereas CCO has a tendency to overestimate BCO.

Close modal

A comparison of bias and precision of both NICO and CCO against BCO at four intraoperative stages during aortic reconstruction is shown in table 3. The bias between NICO and BCO was small before cross-clamping, but increased following aortic cross-clamp and peaked after reperfusion of the unilateral iliac artery. However, the precision of the data remained within the same range. The bias and precision of CCO against BCO was comparable with that of NICO after anesthetic induction, and remained low during the aortic cross-clamp and early reperfusion periods. At the end of surgery, however, CCO overestimated cardiac output by 0.72 l/min and the deviation was larger than that observed in the NICO system. The relative error for all measurements as compared with BCO was −11.5 ± 19.4% for NICO and 8.2 ± 23.0% for CCO. Among the total number of measurements (n = 112), 20 NICO measurements and 18 CCO readings exceeded the relative error by 30%. The time-dependent change of the relative error is summarized in table 3. Figure 3shows the relationship between the bias of NICO against BCO and four other parameters with the potential to affect the precision of NICO measurements: Paco2-Pet CO2difference, P/F ratio, Vco2, and shunt fraction (Qs/Qt). The data indicate that deviations in these parameters did not directly account for the differences between NICO and BCO because no specific trend was noted in the relationship between the bias and the value of these parameters.

Table 3. Bias, Precision and Relative Error of NICO and CCO against BCO at Four Operative Stages

Data were collected from 28 patients and expressed as bias ± precision (1 SD of bias). Differences against BCO were statistically analyzed with repeated measure ANOVA. Relative error (%) is defined as [either (NICO or CCO) − BCO]/BCO and expressed in mean ± SD and is not subject to statistical analysis.

*P < 0.05 vs.  after anesthetic induction.

BCO, CCO, and NCO = bolus thermodilution, continuous thermodilution, and noninvasive cardiac output; declamp = at reperfusion of unilateral iliac artery; Endop = during peritoneal closure; Postinduction = after anesthetic induction; XC = during aortic cross-clamping.

Table 3. Bias, Precision and Relative Error of NICO and CCO against BCO at Four Operative Stages
Table 3. Bias, Precision and Relative Error of NICO and CCO against BCO at Four Operative Stages

Fig. 3. Scatter plot between changes of four potentially significant parameters (Paco2-Pet CO2difference, P/F ratio, Vco2, and shunt) and bias of NICO monitoring compared with bolus thermodilution (BCO). No significant relationship was found between these changes in these parameters and the bias between NICO and BCO.

Fig. 3. Scatter plot between changes of four potentially significant parameters (Paco2-Pet CO2difference, P/F ratio, Vco2, and shunt) and bias of NICO monitoring compared with bolus thermodilution (BCO). No significant relationship was found between these changes in these parameters and the bias between NICO and BCO.

Close modal

This study had four main findings. The first was that the NICO monitor provides a reasonably accurate overall estimate of cardiac output. The second finding, however, was that the NICO system experienced decreases in accuracy following aortic cross-clamping and declamping and that it consistently underestimated cardiac output as compared with BCO. The third finding was that, overall, in terms of bias and precision, NICO and CCO methods are comparable relative to BCO. The fourth finding was that analyses of the Paco2-Petco2difference, P/F ratio, Vco2and shunt fraction showed that none of these parameters could account for the increased bias in NICO estimates of cardiac output following reperfusion.

The NICO monitor uses the differential CO2Fick partial rebreathing method 9,10to measure pulmonary capillary blood flow noninvasively and continuously in mechanically ventilated patients. The NICO algorithm is predicated on the following assumptions: (1) a stable CO2dissociation curve to convert Petco2to Caco2; (2) constant mixed venous CO2content; (3) constant dead space; and (4) stable pulmonary capillary blood flow during the measurement cycle. In addition to calculating cardiac output, it is necessary to estimate the shunt fraction correctly using Nunn's isoshunt plots by entering hemoglobin, Pao2, and Paco2values. This device requires minimal operator experience and is not subject to electromagnetic interference during surgery. These characteristics suggest clinical advantages to the use of this device, but only a limited number of studies regarding the accuracy of the NICO monitor have been conducted. 13–17 

In our study, the agreement between NICO and CCO after anesthetic induction (−0.1 ± 0.61 l/min) was comparable with these previous findings, and we suggest that NICO provides an accurate estimate of cardiac output in stable conditions. However, the reports cited above did not specifically address the accuracy of the NICO monitor in the context of intraoperative events. This prompted us to investigate how aortic cross-clamping and declamping affect the accuracy of NICO monitoring and we found that bias increased significantly after aortic cross-clamp release.

Several factors are implicated in the difference between NICO and thermodilution cardiac output monitoring. Tachibana et al.  reported a bias and precision of 0.28 ± 2.04 l/min during large tidal ventilation, but more discrepant results (−1.6 ± 2.24 l/min) during low tidal ventilation. 16They speculated that slower changes in CO2stored in the body and decreased mixed venous CO2content following the decrease in tidal volume result in increased bias. 16It is conceivable that the increased bias immediately following reperfusion may be caused by an abrupt increase in Vco2that cause similar effects to decreasing tidal volume. However, this mechanism may not fully account for our finding because the measurement was performed after Vco2had stabilized. Alternatively, subclinical pulmonary dysfunction might negatively affect estimates of the shunt fraction and, consequently, of cardiac output. Maxwell et al.  used a porcine model of severe chest trauma and hemorrhagic shock in an experimental setting to compare the performance of the NICO monitor with CCO. 15In that study, bias and precision was reported as 0.01 ± 0.69 l/min. In addition, Odenstedt et al.  reported good agreement between NICO and both CCO and BCO (bias ± 1 SD, −0.04 ± 1.72 l/min) in critically ill patients, although they noted significant underestimation of shunt fraction and Vco2. 17Although pulmonary oxygenation remains relatively stable during surgery, subclinical pulmonary dysfunction, especially increased dead space caused by ischemia-reperfusion, may contribute to the increase in bias. However, in our study no factor could be singled out to account for the increase in bias in NICO measurements after reperfusion. We speculate that relatively acute changes in Vco2and mixed venous CO2content persisting more than 15 min after reperfusion may be responsible for the underestimation of cardiac output as measured by the NICO monitor after aortic declamping. Alterations in pulmonary dead space and intrapulmonary shunt not reflected in the blood gas analysis may also contribute to this change. In addition, intraoperative fluctuations in such parameters as pH, temperature, hemoglobin concentration Vco2, dead space, and shunt can also be potential sources of error. Further study is obviously needed to clarify which factors influence the accuracy of NICO after reperfusion.

Comparisons between NICO and other continuous cardiac output monitoring systems provide clinically relevant information about the usefulness of NICO, which, all other considerations aside, is important because continuous and automatic monitoring is preferable in a clinical setting. Previous investigations report the bias and precision between CCO and BCO as 0.49 ± 1.01, 0.12 ± 0.84, and 0.40 ± 1.26 l/min, respectively, in either the Baxter Vigilance or Abbott Opti-Q system (Abbott, Chicago, IL). 18–20Our data show a comparable degree of agreement between CCO and BCO. Although we are not able to specifically address the reason of increased bias of CCO at the end of surgery, increased cardiac output during this period may account for the increased bias of CCO. 21 

In regard to patient outcome, the routine use of pulmonary artery catheters in aortic surgery is associated with conflicting results. Berlauk et al.  report that preoperative hemodynamic optimization through the use of pulmonary artery catheters improved outcome in atherosclerotic aortic occlusive disease patients. 22On the contrary, Valentine et al.  concluded that the routine use of pulmonary artery catheters is not beneficial and may be associated with a higher rate of intraoperative complications. 6 

These conclusions prompted us to investigate whether monitoring cardiac output using less invasive methods is adequate for intraoperative monitoring during elective aortic surgery. Several studies demonstrate that the monitoring of cardiac output and related parameters during aortic cross-clamping provides clinically useful information. For example, it has been shown that changes in cardiac output and CO2production during aortic cross-clamping are a predictor of hypotension at declamping. 23Whalley et al.  reported that the increase in systemic vascular resistance during aortic cross-clamping positively correlated with a base deficit after reperfusion. 24These findings indicate that cardiac output and systemic vascular resistance, which can be monitored by NICO and the central venous catheter, may make it possible to predict declamping shock and take preventive measures. However, NICO cannot provide information regarding intravascular fluid status and may not be useful in the guidance of intraoperative fluid administration for patients with significant cardiac or pulmonary dysfunction. In this regard, it cannot be recommended as a replacement for pulmonary artery catheter monitoring.

In conclusion, it was determined that NICO provides an accurate estimate of cardiac output after anesthetic induction but underestimates cardiac output after aortic cross-clamping and declamping. Changes in Vco2and subsequent total body CO2content change may be responsible for this bias. CCO shows a relatively smaller degree of bias than NICO but overestimates cardiac output at the end of the surgery. The performance of NICO is comparable with CCO in this clinical setting, and we believe that with cautious evaluation of data NICO may prove to be useful during elective aortic reconstruction surgery.

Appendix: Calculation of Parameters

Shunt fraction = (Cco2− Cao2) / (Cco2− Cvo2).

Cco2= ([Hb]* 1.39) + PAo2* 0.031.

Cao2= 1.39 * Sao2[Hb]+ 0.0031 * Pao2.

Cvo2= 1.39 * Svo2*[Hb]+ 0.0031 * Pvo2.

PAo2= Fio2(PB− PH2O) − Paco2/ R.

Where PB= 760, PH2O= 47, R = 0.8.

Cao2= oxygen content of arterial blood; Cco2= oxygen content of pulmonary capillary blood; Cvo2= oxygen content of mixed venous blood; Fio2= fraction of inhaled oxygen concentration; Hb = blood hemoglobin concentration; Paco2= carbon dioxide partial pressure of arterial blood; PAo2= oxygen partial pressure of alveolar gas; Pao2= oxygen partial pressure of arterial blood; PB= barometric pressure; PH2O= partial pressure of water vapor; Pvo2= oxygen partial pressure of mixed venous blood; R = respiratory quotient.

1.
Boyd O, Grounds RM, Bennett ED: A randomized clinical trial of the effect of deliberate perioperative increase of oxygen delivery on mortality in high-risk surgical patients. JAMA 1993; 270: 2699–707
2.
Ueno S, Tanabe G, Yamada H, Kusano C, Yoshidome S, Nuruki K, Yamamoto S, Aikou T: Response of patients with cirrhosis who have undergone partial hepatectomy to treatment aimed at achieving supranormal oxygen delivery and consumption. Surgery 1998; 123: 278–86
3.
Wilson J, Woods I, Fawcett J, Whall R, Dibb W, Morris C, McManus E: Reducing the risk of major elective surgery: Randomised controlled trial of preoperative optimisation of oxygen delivery. BMJ 1999; 318: 1099–103
4.
Connors AF Jr, Speroff T, Dawson NV, Thomas C, Harrell FE Jr, Wagner D, Desbiens N, Goldman L, Wu AW, Califf RM, Fulkerson WJ Jr, Vidaillet H, Broste S, Bellamy P, Lynn J, Knaus WA: The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. JAMA 1996; 276: 889–97
5.
Bender JS, Smith-Meek MA, Jones CE: Routine pulmonary artery catheterization does not reduce morbidity and mortality of elective vascular surgery: Results of a prospective, randomized trial. Ann Surg 1997; 226: 229–37
6.
Valentine RJ, Duke ML, Inman MH, Grayburn PA, Hagino RT, Kakish HB, Clagett GP: Effectiveness of pulmonary artery catheters in aortic surgery: A randomized trial. J Vasc Surg 1998; 27: 203–12
7.
Polanczyk CA, Rohde LE, Goldman L, Cook EF, Thomas EJ, Marcantonio ER, Mangione CM, Lee TH: Right heart catheterization and cardiac complications in patients undergoing noncardiac surgery: An observational study. JAMA 2001; 286: 309–14
8.
Botero M, Lobato EB: Advances in noninvasive cardiac output monitoring: An update. J Cardiothorac Vasc Anesth 2001; 15: 631–40
9.
de Abreu M, Quintel M, Ragaller M, Albrecht D: Partial carbon dioxide rebreathing: A reliable technique for noninvasive measurement of nonshunted pulmonary capillary blood flow. Crit Care Med 1997; 25: 675–83
10.
Jaffe M: Partial CO2rebreathing cardiac output-operating principles of NICO system. J Clin Monit Comput 1999; 15: 387–401
11.
Bland J, Altman D: Statistical methods for assessing agreement between two method of clinical measurement. Lancet 1986; 1: 307–10
12.
Boldt J, Menges T, Wollbruck M, Hammermann H, Hempelmann G: Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient? Crit Care Med 1994; 22: 1913–18
13.
van Heerden PV, Baker S, Lim SI, Weidman C, Bulsara M: Clinical evaluation of the non-invasive cardiac output (NICO) monitor in the intensive care unit. Anaesth Intensive Care 2000; 28: 427–30
14.
Nilsson LB, Eldrup N, Berthelsen PG: Lack of agreement between thermodilution and carbon dioxide-rebreathing cardiac output. Acta Anaesthesiol Scand 2001; 45: 680–85
15.
Maxwell RA, Gibson JB, Slade JB, Fabian TC, Proctor KG: Noninvasive cardiac output by partial CO2rebreathing after severe chest trauma. J Trauma 2001; 51: 849–53
16.
Tachibana K, Imanaka H, Miyano H, Takeuchi M, Kumon K, Nishimura M: Effect of ventilatory settings on accuracy of cardiac output measurement using partial CO2rebreathing. A nesthesiology 2002; 96: 96–102
17.
Odenstedt H, Stenqvist O, Lundin S: Clinical evaluation of a partial CO2rebreathing technique for cardiac output monitoring in critically ill patients. Acta Anaesthesiol Scand 2002; 46: 152–59
18.
Burchell SA, Yu M, Takiguchi SA, Ohta RM, Myers SA: Evaluation of a continuous cardiac output and mixed venous oxygen saturation catheter in critically ill surgical patients. Crit Care Med 1997; 25: 388–91
19.
Mihm FG, Gettinger A, Hanson CW, Gilbert HC, Stover EP, Vender JS, Beerle B, Haddow G: A multicenter evaluation of a new continuous cardiac output pulmonary artery catheter system. Crit Care Med 1998; 26: 1346–50
20.
Zollner C, Polasek J, Kilger E, Pichler B, Jaenicke U, Briegel J, Vetter HO, Haller M: Evaluation of a new continuous thermodilution cardiac output monitor in cardiac surgical patients: A prospective criterion standard study. Crit Care Med 1999; 27: 293–98
21.
Jacquet L, Hanique G, Glorieux D, Matte P, Goenen M: Analysis of the accuracy of continuous thermodilution cardiac output measurement: Comparison with intermittent thermodilution and Fick cardiac output measurement. Intensive Care Med 1996; 22: 1125–29
22.
Berlauk JF, Abrams JH, Gilmour IJ, O'Connor SR, Knighton DR, Cerra FB: Preoperative optimization of cardiovascular hemodynamics improves outcome in peripheral vascular surgery: A prospective, randomized clinical trial. Ann Surg 1991; 214: 289–99
23.
Boccara G, Jaber S, Eliet J, Mann C, Colson P: Monitoring of end-tidal carbon dioxide partial pressure changes during infrarenal aortic cross-clamping: A non-invasive method to predict unclamping hypotension. Acta Anaesthesiol Scand 2001; 45: 188–93
24.
Whalley DG, Salevsky FC, Ryckman JV: Haemodynamic and metabolic consequences of aortic occlusion during abdominal aortic surgery. Br J Anaesth 1993; 70: 96–98