Fig. 1. ( A ) Rear view, recreated. Temperature cable trapped between knob and base of the automatic pressure limiting valve. ( B ) Front view, recreated. Automatic pressure limiting knob shields view of cable trapped beneath knob. Elevation of knob is subtle and easily missed.