Figure 1. Pressure-volume relations of Fastrach #7 and #7.5 endotracheal tubes compared with regular endotracheal tubes of the same sizes. New endotracheal tubes were inflated with air to the indicated volumes through a stopcock using a syringe. The resulting air pressure in the cuff was measured after equilibrium and plotted against the volume of air inflated. 

Figure 1. Pressure-volume relations of Fastrach #7 and #7.5 endotracheal tubes compared with regular endotracheal tubes of the same sizes. New endotracheal tubes were inflated with air to the indicated volumes through a stopcock using a syringe. The resulting air pressure in the cuff was measured after equilibrium and plotted against the volume of air inflated. 