Fig. 2. The harmonic analysis of the frequency response of the liquid-filled catheter-transducer system (CT) allowed determination of the parameters of the transfer function. The maximal amplitude of the transfer function at resonant frequency (Fr), namely the gain (G), was obtained from the amplitude ratio of the CT signal to the reference signal of the catheter-tipped pressure transducer for Fr. The phase of the transfer function was obtained from the phase lag of the CT pressure signal vs.  the reference pressure signal.

Fig. 2. The harmonic analysis of the frequency response of the liquid-filled catheter-transducer system (CT) allowed determination of the parameters of the transfer function. The maximal amplitude of the transfer function at resonant frequency (Fr), namely the gain (G), was obtained from the amplitude ratio of the CT signal to the reference signal of the catheter-tipped pressure transducer for Fr. The phase of the transfer function was obtained from the phase lag of the CT pressure signal vs.  the reference pressure signal.