Fig. 3. Relations of (A ) natural frequency and (B ) damping coefficient of the transfer function of the liquid-filled catheter-transducer system (CT) to time. They differ depending on the initial CT fast flush protocol (see Materials and Methods). Each relation is drawn from the connection of means ± SD using five replicate measurements in a catheter at 0, 1, 6, 24, and 48 h.

Fig. 3. Relations of (A ) natural frequency and (B ) damping coefficient of the transfer function of the liquid-filled catheter-transducer system (CT) to time. They differ depending on the initial CT fast flush protocol (see Materials and Methods). Each relation is drawn from the connection of means ± SD using five replicate measurements in a catheter at 0, 1, 6, 24, and 48 h.